How to recalibrate or add a peak to the existing calibration in Clarity chromatography software

Tu, 4.11.2025
| Original article from: DataApex
Learn how to manually recalibrate specific calibration levels for individual compounds using the Recalibration Mode and Add Peak functions to improve curve accuracy and analytical reliability.
Video placeholder
  • Photo: DataApex: How to recalibrate or add a peak to the existing calibration
  • Video: DataApex: How to recalibrate or add a peak to the existing calibration

Calibration is a crucial step in quantitative analytical chemistry, ensuring that the relationship between instrument response and analyte concentration remains accurate and reliable. However, during instrument operation, individual calibration levels may occasionally produce outliers that distort the overall calibration curve. This instructional guide provides a detailed procedure for manually recalibrating one or more levels for specific compounds when a correct standard has already been measured. 

Objective

To demonstrate how to correct a single calibration level (e.g., Level 3 of Compound B) that appears as an outlier by using a correctly measured standard. The recalibration procedure replaces the incorrect response value, resulting in a more accurate and reliable calibration curve.

Required Tools and Prerequisites

Before performing recalibration, ensure the following conditions are met:

  • Software: DataApex Clarity Chromatography Software
  • Access rights: User must have permission to modify calibration data.
  • Measured standards: A valid chromatogram or data file from the correct standard measurement is available.
  • Known target compound(s): Identification of the compound(s) requiring recalibration (e.g., “Compound B”).

Optional but recommended:

  • Prior knowledge of the software’s Calibration Options settings.
  • Familiarity with Add Peak and Calibration Curve visualization tools.

Step-by-Step Procedure

1. Identify the Outlier in the Calibration Curve

Begin by examining your existing calibration plot. Identify which compound and which level (e.g., Level 3 of Compound B) deviates significantly from the expected regression line.

Outliers often result from manual integration errors, incorrect concentration entries, or instrumental noise during standard measurement.

2. Open the Compound Record
  1. Navigate to the Compounds tab in your software.
  2. Select the correct standard that you plan to use for recalibration.
  3. Confirm that the chromatographic data corresponds to the compound and calibration level you wish to correct.
3. Switch to Recalibration Mode
  1. Ensure that the displayed compound level matches the desired calibration level (e.g., Level 3).
  2. In the toolbar or menu, locate the Mode selection option and switch to Recalibration.

This mode allows the replacement of response data for specific calibration levels while preserving other calibration information.

4. Configure Calibration Options
  1. Open the Calibration Options dialog.
  2. Set the Recalibration Mode to Replace.
    • This ensures that the new measurement replaces the existing data point instead of averaging or appending it.
  3. If needed, refer to your software’s reference video or manual section titled “How to Modify the Calibration Settings” for detailed configuration guidance.
5. Use the “Add Peak” Function
  1. Click Add Peak.
  2. Select the chromatographic peak corresponding to the compound you wish to recalibrate.
    • The Add Peak function inserts a single selected peak into the chosen calibration level.
  3. Confirm your selection.
    • The system will automatically update the response value for the current calibration level.
6. Verify the Calibration Curve

Once the recalibration is complete:

  1. Reopen the calibration curve view for the affected compound (e.g., Compound B).
  2. Check that the curve now displays improved linearity and correlation.
  3. If multiple compounds or levels require recalibration, repeat the process for each one.
7. Add New Compounds (Optional)

This same procedure can also be applied to add new compounds to an existing calibration series:

  1. Enter the compound name and concentration (amount).
  2. Use the Add Peak function to associate the appropriate signal with the correct level.
  3. Repeat for each required calibration level.

Discussion and Best Practices

  • Accuracy and validation: Always verify recalibrated results using a quality control sample or check standard to confirm that the calibration update improved analytical accuracy.
  • Documentation: Record each recalibration event in the laboratory logbook or electronic data management system, noting the reason for adjustment and the file used.
  • Automation caution: Avoid overreliance on automated recalibration features—manual review ensures that only valid data points are replaced.
  • System suitability: Confirm that recalibration does not violate method validation requirements in regulated environments (e.g., GMP, ISO 17025).

Conclusion

Manual recalibration of individual calibration levels is a valuable technique for maintaining data integrity and analytical reliability. By using the Recalibration Mode, Calibration Options, and Add Peak functions, users can efficiently replace erroneous responses with accurate measurements from valid standards.

This targeted recalibration approach improves the precision of the calibration curve, extends instrument usability, and enhances confidence in quantitative results.

For additional training, tutorials, and troubleshooting resources, visit the official YouTube channel or user documentation provided by your software vendor.

DataApex
LinkedIn Logo
 

Related content

Sensitive and selective quantitation of bile acids using targeted MS2/MS3 on the Stellar mass spectrometer

Posters
| 2025 | Thermo Fisher Scientific (MSACL)
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Selective and sensitive measurement of 17 steroids in human serum using a Stellar mass spectrometer

Technical notes
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Automated Multiple Reaction Monitoring (MRM) Method Development for Peptide Drugs Using waters_connect for Quantitation Software

Applications
| 2025 | Waters
Instrumentation
Software, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Waters
Industries
Pharma & Biopharma

Constant Neutral Loss, Precursor Ion Scanning, Product Ion Scanning in Support of DMPK Studies Using waters_connect for Quantitation Software Solution and Xevo TQ Absolute XR Mass Spectrometer

Applications
| 2025 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ, Software
Manufacturer
Waters
Industries
Pharma & Biopharma, Clinical Research, Metabolomics

pH gradient Analysis of Infliximab charge variant

Applications
| 2025 | Shimadzu
Instrumentation
HPLC, Consumables, LC columns
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

How to recalibrate a level in the calibration in Clarity Chromatography Software
Article | Product

How to recalibrate a level in the calibration in Clarity Chromatography Software

Learn how to correct faulty calibration levels in Clarity software. This guide explains when to remove a point, when to recalibrate, and how to keep calibration curves accurate and reliable.
DataApex
tag
share
more
Work Smarter, Review Faster with Chromeleon CDS
Article | Product

Work Smarter, Review Faster with Chromeleon CDS

Discover how Chromeleon CDS View Settings streamline data processing and review, improve SOP compliance, reduce errors, and boost productivity across analytical labs.
Thermo Fisher Scientific
tag
share
more
What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1
Article | Academy

What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1

Learn how to balance purity, yield, and throughput when scaling analytical HPLC methods to preparative chromatography.
Phenomenex
tag
share
more
News from LabRulezLCMS Library - Week 47, 2025
Article | Application

News from LabRulezLCMS Library - Week 47, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Waters Corporation and poster by Thermo Fisher Scientific / HPLC!
LabRulez
tag
share
more
 

Related content

Sensitive and selective quantitation of bile acids using targeted MS2/MS3 on the Stellar mass spectrometer

Posters
| 2025 | Thermo Fisher Scientific (MSACL)
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Selective and sensitive measurement of 17 steroids in human serum using a Stellar mass spectrometer

Technical notes
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Automated Multiple Reaction Monitoring (MRM) Method Development for Peptide Drugs Using waters_connect for Quantitation Software

Applications
| 2025 | Waters
Instrumentation
Software, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Waters
Industries
Pharma & Biopharma

Constant Neutral Loss, Precursor Ion Scanning, Product Ion Scanning in Support of DMPK Studies Using waters_connect for Quantitation Software Solution and Xevo TQ Absolute XR Mass Spectrometer

Applications
| 2025 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ, Software
Manufacturer
Waters
Industries
Pharma & Biopharma, Clinical Research, Metabolomics

pH gradient Analysis of Infliximab charge variant

Applications
| 2025 | Shimadzu
Instrumentation
HPLC, Consumables, LC columns
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

How to recalibrate a level in the calibration in Clarity Chromatography Software
Article | Product

How to recalibrate a level in the calibration in Clarity Chromatography Software

Learn how to correct faulty calibration levels in Clarity software. This guide explains when to remove a point, when to recalibrate, and how to keep calibration curves accurate and reliable.
DataApex
tag
share
more
Work Smarter, Review Faster with Chromeleon CDS
Article | Product

Work Smarter, Review Faster with Chromeleon CDS

Discover how Chromeleon CDS View Settings streamline data processing and review, improve SOP compliance, reduce errors, and boost productivity across analytical labs.
Thermo Fisher Scientific
tag
share
more
What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1
Article | Academy

What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1

Learn how to balance purity, yield, and throughput when scaling analytical HPLC methods to preparative chromatography.
Phenomenex
tag
share
more
News from LabRulezLCMS Library - Week 47, 2025
Article | Application

News from LabRulezLCMS Library - Week 47, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Waters Corporation and poster by Thermo Fisher Scientific / HPLC!
LabRulez
tag
share
more
 

Related content

Sensitive and selective quantitation of bile acids using targeted MS2/MS3 on the Stellar mass spectrometer

Posters
| 2025 | Thermo Fisher Scientific (MSACL)
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Selective and sensitive measurement of 17 steroids in human serum using a Stellar mass spectrometer

Technical notes
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Automated Multiple Reaction Monitoring (MRM) Method Development for Peptide Drugs Using waters_connect for Quantitation Software

Applications
| 2025 | Waters
Instrumentation
Software, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Waters
Industries
Pharma & Biopharma

Constant Neutral Loss, Precursor Ion Scanning, Product Ion Scanning in Support of DMPK Studies Using waters_connect for Quantitation Software Solution and Xevo TQ Absolute XR Mass Spectrometer

Applications
| 2025 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ, Software
Manufacturer
Waters
Industries
Pharma & Biopharma, Clinical Research, Metabolomics

pH gradient Analysis of Infliximab charge variant

Applications
| 2025 | Shimadzu
Instrumentation
HPLC, Consumables, LC columns
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

How to recalibrate a level in the calibration in Clarity Chromatography Software
Article | Product

How to recalibrate a level in the calibration in Clarity Chromatography Software

Learn how to correct faulty calibration levels in Clarity software. This guide explains when to remove a point, when to recalibrate, and how to keep calibration curves accurate and reliable.
DataApex
tag
share
more
Work Smarter, Review Faster with Chromeleon CDS
Article | Product

Work Smarter, Review Faster with Chromeleon CDS

Discover how Chromeleon CDS View Settings streamline data processing and review, improve SOP compliance, reduce errors, and boost productivity across analytical labs.
Thermo Fisher Scientific
tag
share
more
What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1
Article | Academy

What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1

Learn how to balance purity, yield, and throughput when scaling analytical HPLC methods to preparative chromatography.
Phenomenex
tag
share
more
News from LabRulezLCMS Library - Week 47, 2025
Article | Application

News from LabRulezLCMS Library - Week 47, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Waters Corporation and poster by Thermo Fisher Scientific / HPLC!
LabRulez
tag
share
more
 

Related content

Sensitive and selective quantitation of bile acids using targeted MS2/MS3 on the Stellar mass spectrometer

Posters
| 2025 | Thermo Fisher Scientific (MSACL)
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Selective and sensitive measurement of 17 steroids in human serum using a Stellar mass spectrometer

Technical notes
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Clinical Research

Automated Multiple Reaction Monitoring (MRM) Method Development for Peptide Drugs Using waters_connect for Quantitation Software

Applications
| 2025 | Waters
Instrumentation
Software, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Waters
Industries
Pharma & Biopharma

Constant Neutral Loss, Precursor Ion Scanning, Product Ion Scanning in Support of DMPK Studies Using waters_connect for Quantitation Software Solution and Xevo TQ Absolute XR Mass Spectrometer

Applications
| 2025 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ, Software
Manufacturer
Waters
Industries
Pharma & Biopharma, Clinical Research, Metabolomics

pH gradient Analysis of Infliximab charge variant

Applications
| 2025 | Shimadzu
Instrumentation
HPLC, Consumables, LC columns
Manufacturer
Shimadzu
Industries
Pharma & Biopharma
 

Related articles

How to recalibrate a level in the calibration in Clarity Chromatography Software
Article | Product

How to recalibrate a level in the calibration in Clarity Chromatography Software

Learn how to correct faulty calibration levels in Clarity software. This guide explains when to remove a point, when to recalibrate, and how to keep calibration curves accurate and reliable.
DataApex
tag
share
more
Work Smarter, Review Faster with Chromeleon CDS
Article | Product

Work Smarter, Review Faster with Chromeleon CDS

Discover how Chromeleon CDS View Settings streamline data processing and review, improve SOP compliance, reduce errors, and boost productivity across analytical labs.
Thermo Fisher Scientific
tag
share
more
What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1
Article | Academy

What is the most appropriate way to approach scaling up your analytical method when transferring it to preparative chromatography scale? Part 1

Learn how to balance purity, yield, and throughput when scaling analytical HPLC methods to preparative chromatography.
Phenomenex
tag
share
more
News from LabRulezLCMS Library - Week 47, 2025
Article | Application

News from LabRulezLCMS Library - Week 47, 2025

This week we bring you application notes by Agilent Technologies, Shimadzu and Waters Corporation and poster by Thermo Fisher Scientific / HPLC!
LabRulez
tag
share
more
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike