Automated Device Shutdown After Sequence Completion in Clarity Chromatography Software

Tu, 16.12.2025
| Original article from: DataApex
Learn how shutdown methods in Clarity Sequence Options enable automatic post-sequence actions such as cooling devices, switching off lamps, or sending shutdown commands after the last run.
Video placeholder
  • Photo: DataApex: Automated Device Shutdown After Sequence Completion in Clarity
  • Video: DataApex: How to shutdown the devices after the sequence is finished

Introduction

In automated chromatographic workflows, it is often necessary to ensure that controlled devices perform specific actions once a measurement sequence has finished. The video demonstrates how Clarity Chromatography Software  allows users to define such actions by configuring a Shutdown Method within the Sequence Options.

This functionality enables devices to transition into a safe or standby state automatically after the last sequence row is measured, reducing manual intervention and supporting unattended operation.

Purpose of the Shutdown Method

The Shutdown Method is designed to define post-sequence actions for controlled devices once a sequence has finished.

This method can be used to:

  • Gradually lower the flow
  • Switch off detector lamps
  • Cool down devices
  • Bring instruments into standby or shutdown state

These actions are executed automatically based on the user-defined settings in the sequence configuration.

Accessing Sequence Options

To configure post-sequence behavior, it is necessary to:

  1. Open the Sequence Options
  2. Navigate to the section labeled “After sequence is finished”

This section contains all relevant controls for defining how and when shutdown actions are applied.

Configuration in the “After Sequence Is Finished” Section

1. Selecting the Shutdown Method

In the After sequence is finished section, the user must:

  • Select a prepared Shutdown Method
  • Define how the method should be executed

The Shutdown Method must already exist and contain the desired device actions (e.g. cooling, lamp switch-off, flow reduction).

2. Differences Between Devices

The necessary settings differ according to the used devices. This means shutdown behavior is device-dependent, and not all instruments require the same configuration.

Examples:

  • Some devices only require the method to be sent
    • For example, when the goal is to initiate cooling
  • Other devices require the method to be run
    • For example, detectors that include an option such as “switch the lamp off after run”

3. Performing Device Shutdown

In addition to sending or running the shutdown method, the user can enable:

  • Perform Shutdown

When this option is checked:

  • A shutdown command is sent directly to the devices
  • Each device may react differently to this command

Device responses to this command depend on the specific hardware.

Execution Order After Sequence Completion

Once the last row of the sequence is measured, Clarity executes shutdown actions in a defined order:

  1. The Shutdown Method is sent
  2. The Shutdown Method is run (if configured)
  3. The Perform Shutdown command is sent (if enabled)

This ensures that all configured post-sequence actions are applied consistently and automatically.

Practical Implications

The demonstrated configuration allows users to:

  • Automate post-analysis instrument handling
  • Reduce the need for manual shutdown procedures
  • Ensure devices enter a defined and safe state after sequence completion

This is particularly useful for long or unattended sequences where operator presence at the end of the run is not guaranteed.

Conclusion

Clarity provides a flexible mechanism to control device behavior after a sequence has finished through the use of a Shutdown Method and the After sequence is finished settings.

By selecting an appropriate shutdown method and defining whether it should be sent, run, or followed by a shutdown command, users can tailor post-sequence behavior to the requirements of their specific devices. The execution is performed automatically once the final sequence row is measured, ensuring consistent and reliable shutdown behavior.

DataApex
LinkedIn Logo
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

How to set the Sequence Options in Clarity SW
Article | Product

How to set the Sequence Options in Clarity SW

Learn how to configure sequence options in Clarity software, including passive vs. active mode, idle time, file naming, calibration settings, and shutdown methods.
DataApex
tag
share
more
How post run actions work in Clarity  software
Article | Product

How post run actions work in Clarity software

Clarity software offers powerful post-run actions for exporting data, running external programs, managing calibration, and SST inclusion, streamlining chromatographic workflows.
DataApex
tag
share
more
How to print a report from Sequence in Clarity software
Article | Product

How to print a report from Sequence in Clarity software

Learn how to configure Clarity Chromatography Software to automatically generate and print PDF reports after measurements using post-run settings.
DataApex
tag
share
more
Clarity chromatograhy software: Tutorial - Sequence window - Overview
Article | Product

Clarity chromatograhy software: Tutorial - Sequence window - Overview

Learn how to use the Sequence Window in Clarity CDS. Step-by-step guide to sample setup, injection control, calibration options, and sequence management.
DataApex
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

How to set the Sequence Options in Clarity SW
Article | Product

How to set the Sequence Options in Clarity SW

Learn how to configure sequence options in Clarity software, including passive vs. active mode, idle time, file naming, calibration settings, and shutdown methods.
DataApex
tag
share
more
How post run actions work in Clarity  software
Article | Product

How post run actions work in Clarity software

Clarity software offers powerful post-run actions for exporting data, running external programs, managing calibration, and SST inclusion, streamlining chromatographic workflows.
DataApex
tag
share
more
How to print a report from Sequence in Clarity software
Article | Product

How to print a report from Sequence in Clarity software

Learn how to configure Clarity Chromatography Software to automatically generate and print PDF reports after measurements using post-run settings.
DataApex
tag
share
more
Clarity chromatograhy software: Tutorial - Sequence window - Overview
Article | Product

Clarity chromatograhy software: Tutorial - Sequence window - Overview

Learn how to use the Sequence Window in Clarity CDS. Step-by-step guide to sample setup, injection control, calibration options, and sequence management.
DataApex
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

How to set the Sequence Options in Clarity SW
Article | Product

How to set the Sequence Options in Clarity SW

Learn how to configure sequence options in Clarity software, including passive vs. active mode, idle time, file naming, calibration settings, and shutdown methods.
DataApex
tag
share
more
How post run actions work in Clarity  software
Article | Product

How post run actions work in Clarity software

Clarity software offers powerful post-run actions for exporting data, running external programs, managing calibration, and SST inclusion, streamlining chromatographic workflows.
DataApex
tag
share
more
How to print a report from Sequence in Clarity software
Article | Product

How to print a report from Sequence in Clarity software

Learn how to configure Clarity Chromatography Software to automatically generate and print PDF reports after measurements using post-run settings.
DataApex
tag
share
more
Clarity chromatograhy software: Tutorial - Sequence window - Overview
Article | Product

Clarity chromatograhy software: Tutorial - Sequence window - Overview

Learn how to use the Sequence Window in Clarity CDS. Step-by-step guide to sample setup, injection control, calibration options, and sequence management.
DataApex
tag
share
more
 

Related content

Oligonucleotide Analysis Using the Agilent InfinityLab Pro iQ and Altura Oligo HPH-C18 Column

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/SQ, Consumables, LC columns
Manufacturer
Agilent Technologies
Industries
Pharma & Biopharma

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics
 

Related articles

How to set the Sequence Options in Clarity SW
Article | Product

How to set the Sequence Options in Clarity SW

Learn how to configure sequence options in Clarity software, including passive vs. active mode, idle time, file naming, calibration settings, and shutdown methods.
DataApex
tag
share
more
How post run actions work in Clarity  software
Article | Product

How post run actions work in Clarity software

Clarity software offers powerful post-run actions for exporting data, running external programs, managing calibration, and SST inclusion, streamlining chromatographic workflows.
DataApex
tag
share
more
How to print a report from Sequence in Clarity software
Article | Product

How to print a report from Sequence in Clarity software

Learn how to configure Clarity Chromatography Software to automatically generate and print PDF reports after measurements using post-run settings.
DataApex
tag
share
more
Clarity chromatograhy software: Tutorial - Sequence window - Overview
Article | Product

Clarity chromatograhy software: Tutorial - Sequence window - Overview

Learn how to use the Sequence Window in Clarity CDS. Step-by-step guide to sample setup, injection control, calibration options, and sequence management.
DataApex
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike