Why Is It Necessary to Dry the SPE Sorbent Before Elution?

We, 18.2.2026
| Original article from: Phenomenex
Proper drying of the SPE sorbent is a critical step to achieve high sensitivity, good reproducibility, and maximum analyte recovery, especially for complex sample matrices.
<p><strong>Phenomenex:</strong> Why Is It Necessary to Dry the SPE Sorbent Before Elution?</p>

Phenomenex: Why Is It Necessary to Dry the SPE Sorbent Before Elution?

Solid Phase Extraction (SPE) is one of the most highly selective sample preparation methods, meaning that the targeted analytes and their matrices must be considered when choosing the correct SPE sorbent and format. Drying down the sorbent before elution is an important step.

  • Wash solvents are used to flush out unwanted compounds (i.e., contaminants). If the wash solvent is not completely removed, then residual wash solvent could contaminate the final elution.
  • If you do not dry the SPE sufficiently, it could make the final solution less concentrated and result in a reduction in sensitivity.
  • Drying helps physically break down the interactions between the sorbent and analyte for a more concentrated extract.

After the wash solvent has been passed through the cartridge, continue to apply negative pressure (vacuum) or positive pressure. Use this pressure to dry the cartridge for 2 to 5 minutes (or until dry). Turn off the pressure and wipe the tips of the manifold needles to remove any residual solvent.

Notes
  1. Reproducibility: Drying removes traces of residual sample/matrix and wash solvent from the sorbent. This produces a more concentrated final extract with a consistent volume and composition and yields better reproducibility and higher recoveries.
  2. Immiscibility: Drying is especially critical when using elution solvents that are immiscible with water (e.g. hexane, ethyl acetate, and methylene chloride). Residual water in the sorbent bed may prevent dissolution of the analyte in the elution solvent, resulting in low recoveries and irreproducibility.
  3. Over-drying: Avoid over-drying the sorbent, since in some cases, analyte recovery may decrease with excessive vacuum or drying times above 5 minutes. Dry cartridges look distinctly different from moist ones. Make a note of the cartridge appearance before and after conditioning. Also, you could touch the cartridge while it is drying; wet cartridges feel colder than dry cartridges. 
Phenomenex
LinkedIn Logo
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
The Critical Role of Nitrogen Blowdown in SPE Sample Preparation
Article | Science and research

The Critical Role of Nitrogen Blowdown in SPE Sample Preparation

Explore how nitrogen blowdown enhances sample prep efficiency in SPE workflows using Waters Oasis, Agilent Bond Elut and Phenomenex Strata, improving analyte recovery, sensitivity and method reliability.
Organomation
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019
Article | Environment

Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019

How the EU’s revised Urban Wastewater Treatment Directive is driving demand for ultra-sensitive LC-MS/MS workflows—and why nitrogen evaporation is critical for reliable micropollutant monitoring.
Organomation
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
The Critical Role of Nitrogen Blowdown in SPE Sample Preparation
Article | Science and research

The Critical Role of Nitrogen Blowdown in SPE Sample Preparation

Explore how nitrogen blowdown enhances sample prep efficiency in SPE workflows using Waters Oasis, Agilent Bond Elut and Phenomenex Strata, improving analyte recovery, sensitivity and method reliability.
Organomation
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019
Article | Environment

Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019

How the EU’s revised Urban Wastewater Treatment Directive is driving demand for ultra-sensitive LC-MS/MS workflows—and why nitrogen evaporation is critical for reliable micropollutant monitoring.
Organomation
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
The Critical Role of Nitrogen Blowdown in SPE Sample Preparation
Article | Science and research

The Critical Role of Nitrogen Blowdown in SPE Sample Preparation

Explore how nitrogen blowdown enhances sample prep efficiency in SPE workflows using Waters Oasis, Agilent Bond Elut and Phenomenex Strata, improving analyte recovery, sensitivity and method reliability.
Organomation
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019
Article | Environment

Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019

How the EU’s revised Urban Wastewater Treatment Directive is driving demand for ultra-sensitive LC-MS/MS workflows—and why nitrogen evaporation is critical for reliable micropollutant monitoring.
Organomation
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
The Critical Role of Nitrogen Blowdown in SPE Sample Preparation
Article | Science and research

The Critical Role of Nitrogen Blowdown in SPE Sample Preparation

Explore how nitrogen blowdown enhances sample prep efficiency in SPE workflows using Waters Oasis, Agilent Bond Elut and Phenomenex Strata, improving analyte recovery, sensitivity and method reliability.
Organomation
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019
Article | Environment

Efficient Sample Concentration for Micropollutant Surveillance Under EU Directive 2024/3019

How the EU’s revised Urban Wastewater Treatment Directive is driving demand for ultra-sensitive LC-MS/MS workflows—and why nitrogen evaporation is critical for reliable micropollutant monitoring.
Organomation
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike