Agilent 1200 Series HPLC Fraction Collector, G1364C (Refurbished)

Analytical-scale HPLC fraction collector for ≤10 mL/min with low delay volume, fast diverter valve, multiple trigger modes, broad vessel support, GLP features, and full Agilent 1200 Series integration.
Purchase price
6,500
without VAT
In stock
+ 200  delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Installation
Extended warranty
Basic training

Features

The Agilent 1200 Series Fraction Collector G1364C is an analytical-scale fraction collection module designed for precise, automated fraction collection at flow rates below 10 mL/min. It supports a wide range of vial, tube, plate, and funnel configurations and integrates seamlessly with Agilent 1200 Series LC detectors and control modules.

Key Features

  • Analytical-scale fraction collection for flow rates up to 10 mL/min (higher flows possible with modification)
  • Multiple trigger modes: time-based, peak-based, timetable, and manual triggering
  • Discrete and continuous flow collection modes
  • Low internal delay volumes for high chromatographic fidelity
  • Fast-switching low-volume diverter valve (<100 ms switching time)
  • Automatic tray and vessel detection
  • Optional cooling via G1330B Fraction Collector Thermostat
  • GLP support with electronic maintenance and error logs
  • Leak detection and exhaust fan for safe handling of hazardous solvents
  • CAN-based system integration with optional LAN and RS232 connectivity

Technical Specifications (from Agilent brochure)

Flow & Performance
  • Maximum flow rate: 10 mL/min (analytical scale)
  • Maximum pressure at diverter valve: 6 bar
  • Diverter valve internal volume: ~15 µL
  • Switching time: <100 ms
  • Delay volumes (typical):
    • Inlet to diverter valve: ~50 µL
    • Diverter valve to needle: ~10 µL
    • Needle: ~4 µL
  • Delay calibration sensor: Single-wavelength absorbance detector (654 nm LED)
Fraction Collection Modes
  • Trigger modes:
    • Time slices
    • Peak-based (threshold, up/down slope)
    • Timetable (time + peak)
    • Manual trigger (with G1323B control module)
  • Operating modes:
    • Discrete fractions (default)
    • Continuous flow (deep well plates only)
    • Needle-into-location mode
    • Droplet setup mode (bubble-free small fractions)
Fraction Vessels & Capacity
  • Supports:
    • Well plates (MTP, deep well plates)
    • Vials (2 mL)
    • Test tubes (12–30 mm OD)
    • Funnel trays for external containers
  • Maximum tube height:
    • 48 mm (standard needle)
    • 75 mm (short needle assembly G1364-87202)
  • Maximum tube volume:
    • ~20 mL (48 mm tubes)
    • ~30 mL (75 mm tubes)
    • Unlimited with external funnels
  • Automatic tray identification
Detector Compatibility
  • Fully supported:
    • Agilent 1200 Series DAD/MWD (G1315, G1365)
    • Fluorescence detector
    • LC-MSD (G1946C/D, G1956A/B)
  • Other detectors usable but not officially supported for fraction triggering
Interfaces & System Integration
  • CAN (Controller Area Network)
  • Optional:
    • LAN
    • RS232C
    • External contacts interface
    • APG-remote I/O
  • Direct interface to G1330A/B Thermostat
  • CAN-DC-out for Agilent-approved external devices
Safety & Compliance
  • Leak detection with safe leak handling
  • Exhaust fan for solvent vapor removal
  • Early Maintenance Feedback (EMF)
  • Safety Class I instrument
  • Certified to IEC, CSA, UL
  • Installation Category II, Pollution Degree 2
  • Indoor use only
Physical Specifications
  • Dimensions (H × W × D): 200 × 345 × 440 mm
  • Weight: 13.5 kg
  • Power supply: 100–240 VAC, ±10 %
  • Frequency: 50/60 Hz
  • Power consumption: max. 180 W (active), 200 VA (apparent)
  • Operating temperature: 4–55 °C
  • Humidity: <95 % (non-condensing)
  • Operating altitude: up to 2000 m
Seller
LabRulez
Stop looking and start finding. We are your source for news, applications, products, articles, theory, job postings, webinars, e-shop or bazaar in the world of analytical chemistry.
Manufacturer
Agilent Technologies
Agilent Technologies
Analytical scientists and clinical researchers worldwide rely on Agilent to help fulfill their most complex laboratory demands. Our instruments, software, services and consumables address the full range of scientific and laboratory management needs.
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike