The SIL-20AHT autosampler used with the Prominence UFLC system permit ultra-fast analysis while maintaining the injection accuracy and durability requirements demanded of an HPLC system. The retention time reproducibility and injection volume reproducibility are within 0.3 % after endurance testing over 100,000 cycles, results similar to those achieved with standard HPLC.
Injection method: Total-volume sample injection, variable injection volume
Maximum operating pressure: 35 MPa
Injection-volume setting range: 0.1 to 100 μL (standard), 0.1 to 2,000 μL (option)
Number of processed samples: 175 (1 mL vials), 105 (1.5 mL vials), 50 (4 mL vials) 192 (two 96-well MTP/DWP), 768 (two 384-well MTP/DWP) Also, ten 1.5 mL vials in addition to each of the above.
Injection-volume accuracy: 1% max (specified conditions)
Injection-volume precision: RSD: 0.3% max. (specified conditions, typically 0.2% RSD max)
Sample Carryover: 0.005% max. (specified conditions, typically 0.0025% max)
Number of repeated injections: 30 max. per sample
Needle rinsing: Set freely before and after sample injection
Operating pH range: pH 1 to pH 14
Operating temperature range: 4 to 35°C
Dimensions, weight: W260 × D500 × H415 mm, 27 kg
Power requirements: AC 110 V, 230 V, 100 VA, 50/60 Hz