Monitoring Cell Culture Media for Amino Acids Using an ACQUITY UPLC H-Class PLUS System
Technical notes | 2018 | WatersInstrumentation
Cell culture media composition critically influences cell growth, viability and productivity in biopharmaceutical manufacturing.
Amino acids serve as precursor molecules for proteins and their levels fluctuate during culture due to consumption and release by cells.
Monitoring these fluctuations is essential to maintain optimal culture conditions and ensure product quality.
This study aims to evaluate the performance of the ACQUITY UPLC H-Class PLUS System with UV detection for the accurate analysis of amino acids in cell culture media.
The approach is based on pre-column derivatization using the AccQ•Tag methodology and separation on a UPLC column under optimized conditions.
The analysis follows the Waters UPLC Amino Acid Analysis Application Solution consisting of:
Instrument used:
The method resolved all common cell culture amino acids with clear baseline separation within 8 minutes.
Repeatability was assessed through ten replicate injections of a standard mixture.
Relative standard deviations (%RSD) for retention times and peak areas were comparable between the H-Class and H-Class PLUS systems, demonstrating high reproducibility.
Figure summaries showed consistent peak shape and retention across runs.
The validated method provides:
It enables real-time monitoring of media composition to guide feed strategies and maintain culture performance.
Future developments may include:
The ACQUITY UPLC H-Class PLUS System combined with AccQ•Tag derivatization offers a powerful analytical platform for precise monitoring of amino acids in cell culture media.
Its high throughput, reproducibility and ease of integration support quality assurance in biopharmaceutical workflows.
HPLC
IndustriesPharma & Biopharma
ManufacturerWaters
Summary
Význam tématu
Cell culture media composition critically influences cell growth, viability and productivity in biopharmaceutical manufacturing.
Amino acids serve as precursor molecules for proteins and their levels fluctuate during culture due to consumption and release by cells.
Monitoring these fluctuations is essential to maintain optimal culture conditions and ensure product quality.
Cíle a přehled studie / článku
This study aims to evaluate the performance of the ACQUITY UPLC H-Class PLUS System with UV detection for the accurate analysis of amino acids in cell culture media.
The approach is based on pre-column derivatization using the AccQ•Tag methodology and separation on a UPLC column under optimized conditions.
Použitá metodika a instrumentace
The analysis follows the Waters UPLC Amino Acid Analysis Application Solution consisting of:
- Pre-column derivatization using AccQ•Tag reagents.
- Separation on AccQ•Tag Ultra, 1.7 µm, 2.1×100 mm column (p/n 186003837).
- Mobile phases: AccQ•Tag Ultra Eluent A and B, diluted with water.
- Detection via UV absorbance at 260 nm.
Instrument used:
- ACQUITY UPLC H-Class PLUS System equipped with UV detector.
- Empower software with pre-configured method templates.
Hlavní výsledky a diskuse
The method resolved all common cell culture amino acids with clear baseline separation within 8 minutes.
Repeatability was assessed through ten replicate injections of a standard mixture.
Relative standard deviations (%RSD) for retention times and peak areas were comparable between the H-Class and H-Class PLUS systems, demonstrating high reproducibility.
Figure summaries showed consistent peak shape and retention across runs.
Přínosy a praktické využití metody
The validated method provides:
- Rapid and robust quantification of amino acids in complex matrices.
- High separation efficiency and reproducibility.
- Compatibility with routine quality control workflows in biopharmaceutical production.
It enables real-time monitoring of media composition to guide feed strategies and maintain culture performance.
Budoucí trendy a možnosti využití
Future developments may include:
- Integration with automated sampling and online analysis for in situ monitoring.
- Extension to additional metabolites or biogenic amines.
- Coupling with mass spectrometry for enhanced sensitivity and identification.
- Implementation within process analytical technology (PAT) frameworks for continuous manufacturing.
Závěr
The ACQUITY UPLC H-Class PLUS System combined with AccQ•Tag derivatization offers a powerful analytical platform for precise monitoring of amino acids in cell culture media.
Its high throughput, reproducibility and ease of integration support quality assurance in biopharmaceutical workflows.
Reference
- Cell Culture Media: A Review, Labome Method, https://www.labome.com/method/Cell-Culture-Media-A-Review.html.
- Waters, UPLC Amino Acid Analysis Application Solution, Brochure 720001837EN, January 2008.
- Waters, AccQ•Tag Ultra Derivatization Kit Care and Use Manual, 715001331, July 2014.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
High-Throughput Amino Acid Analysis using Tecan Automated Preparation
2020|Waters|Applications
Application Note High-Throughput Amino Acid Analysis using Tecan Automated Preparation Niamh Stafford, Danielle Cullen, Leanne Davey, Norma Breen, Corey E. Reed, Ning Zhang, Paula Hong, Stephan M. Koza Waters Corporation Abstract The objective of this application note is to demonstrate…
Key words
amino, aminotecan, tecanaccq, accqtag, tagderivatization, derivatizationculture, cultureautomation, automationacid, acidkit, kitmanual, manualpreparation, preparationultra, ultrauplc, uplcautomated, automatednorvaline
Amino Acid Analysis - Application Notebook
2018|Waters|Guides
[ APPLICATION NOTEBOOK ] Amino Acid Analysis Application Notebook [ CONTENTS ] [ INTRODUCTION ] Waters UPLC Amino Acid Analysis Solution................................................................................9 UPLC Amino Acid Analysis Solution............................................................................................... 11 Enhancement of the UPLC Amino Acid Analysis Solution with Flexible Detector Options......................................................................................................... 15…
Key words
amino, aminoacid, acidasp, aspuplc, uplcala, alalys, lysleu, leutyr, tyrgly, glyarg, argser, serthr, thrglu, gluile, ilecys
High-Throughput Amino Acid Analysis Using Hamilton Automated Preparation
2020|Waters|Applications
Application Note High-Throughput Amino Acid Analysis Using Hamilton Automated Preparation Niamh Stafford, Danielle Cullen, Leanne Davey, Norma Breen, Corey E. Reed, Ning Zhang, Paula Hong, Stephan M. Koza Waters Corporation Abstract The objective of this application note is to demonstrate…
Key words
accq, accqhamilton, hamiltonamino, aminotag, tagautomation, automationpreparation, preparationautomated, automatedacids, acidskit, kitderivatization, derivatizationacid, aciduplc, uplcnorvaline, norvalinescripts, scriptsequivalency
Food and Beverage Quality Control and Compositional Testing
2022|Waters|Guides
Food and Beverage Quality Control and Compositional Testing Application Notebook Sugar Substitutes Sugars Vitamins Organic Acids Amino Acids Introduction EFFICIENT ANALYTICAL SOLUTIONS TO SUPPORT A SAFE, NUTRITIOUS, AUTHENTIC AND SUSTAINABLE FOOD SUPPLY CHAIN Welcome to Waters’ food quality control and…
Key words
acids, acidssugar, sugarsubstitutes, substitutesvitamins, vitaminsnotebook, notebookvitamin, vitaminbeverage, beverageamino, aminocompositional, compositionalsugars, sugarsapplication, applicationfood, foodorganic, organicacquity, acquitytesting