High-precision analysis of the photodegradable compound Naproxen using the UV cut-off filter of the photodiode array detector SPD-M40
Technical notes | 2021 | ShimadzuInstrumentation
Photodegradable pharmaceuticals can undergo light-induced decomposition during analysis, leading to inaccurate quantification and compromised quality control. Implementing a UV cut-off filter in photodiode array (PDA) detection mitigates photolysis, ensuring reliable results for low-concentration samples and critical regulatory environments.
This study evaluates the performance of the Shimadzu SPD-M40 PDA detector equipped with a UV cut-off filter in high-precision analysis of Naproxen. Key aims include:
Naproxen standard solutions (0.1–40 ppm) were analyzed under reversed-phase HPLC conditions. Analytical parameters:
The core instrument was the Shimadzu SPD-M40 photodiode array detector featuring a removable UV cut-off filter. The filter blocks short-wavelength UV light entering the flow cell, reducing high-energy irradiation of the sample.
Without the UV cut-off filter, calibration curves exhibited large negative intercepts due to photodegradation of Naproxen at low concentrations. Introducing the filter:
These findings highlight that suppression of UV-induced breakdown is most critical for trace-level analyses and low flow conditions.
Integrating a UV cut-off filter in PDA detectors delivers:
Advancements may include:
The SPD-M40 PDA detector’s UV cut-off filter effectively prevents photodegradation of Naproxen, improving calibration accuracy under diverse analytical settings. This innovation supports rigorous quantification of photodegradable compounds in pharmaceutical and environmental laboratories.
HPLC
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Significance of the Topic
Photodegradable pharmaceuticals can undergo light-induced decomposition during analysis, leading to inaccurate quantification and compromised quality control. Implementing a UV cut-off filter in photodiode array (PDA) detection mitigates photolysis, ensuring reliable results for low-concentration samples and critical regulatory environments.
Objectives and Study Overview
This study evaluates the performance of the Shimadzu SPD-M40 PDA detector equipped with a UV cut-off filter in high-precision analysis of Naproxen. Key aims include:
- Comparing calibration behavior with and without the UV cut-off filter.
- Assessing the impact of varying injection volumes and flow rates on photolysis suppression.
Methodology
Naproxen standard solutions (0.1–40 ppm) were analyzed under reversed-phase HPLC conditions. Analytical parameters:
- Column: Shim-pack VP-ODS (150 × 4.6 mm I.D., 5 μm)
- Mobile phase: 50 % methanol with 0.1 % acetic acid
- Flow rate range: 0.25–2.0 mL/min
- Column temperature: 40 °C
- Detection wavelength: 230 nm
Used Instrumentation
The core instrument was the Shimadzu SPD-M40 photodiode array detector featuring a removable UV cut-off filter. The filter blocks short-wavelength UV light entering the flow cell, reducing high-energy irradiation of the sample.
Main Results and Discussion
Without the UV cut-off filter, calibration curves exhibited large negative intercepts due to photodegradation of Naproxen at low concentrations. Introducing the filter:
- Reduced negative intercepts by over 80 % across different injection volumes.
- Maintained linearity (r2 > 0.99998) even at low flow rates, where residence time—and thus photolysis risk—increases.
- Demonstrated consistent slope values, confirming that absorbance measurements reflect intact analyte.
These findings highlight that suppression of UV-induced breakdown is most critical for trace-level analyses and low flow conditions.
Benefits and Practical Applications
Integrating a UV cut-off filter in PDA detectors delivers:
- Enhanced accuracy in quantifying photolabile compounds.
- Reliable detection at low concentrations, crucial for environmental monitoring and pharmaceutical QA/QC.
- Adaptability to variable analytical conditions without compromising data integrity.
Future Trends and Potential Applications
Advancements may include:
- Customizable cut-off wavelengths for broader compound classes.
- Miniaturized flow cells with integrated filters for ultra-trace analysis.
- Coupling with mass spectrometry to combine photolysis suppression with structural elucidation.
Conclusion
The SPD-M40 PDA detector’s UV cut-off filter effectively prevents photodegradation of Naproxen, improving calibration accuracy under diverse analytical settings. This innovation supports rigorous quantification of photodegradable compounds in pharmaceutical and environmental laboratories.
References
- Packer J.L. et al., 2003, Aquatic Sciences, 65, 342–351.
- Jacobs L.E. et al., 2011, Water Research, 45, 4449–4458.
- Hsu Y.H. et al., 2006, Biomedical Chromatography, 20, 787–793.
- Kawabata K. et al., 2018, Chromatographia, 39, 139–146.
- Hayase N. et al., 1994, Journal of Pharmaceutical Sciences, 83, 532–538.
- Ragno G. et al., 2002, Journal of Pharmaceutical and Biomedical Analysis, 27, 19–24.
- Lin A.Y.-C. et al., 2005, Environmental Toxicology and Chemistry, 24, 1303–1309.
- Young R.B. et al., 2013, Environmental Science & Technology, 47, 8416–8424.
- Mori Y. et al., 1969, Journal of the Japanese Society of Food and Nutrition, 22, 12–16.
- Taylor R. et al., 1973, Archives of Biochemistry and Biophysics, 156, 521–533.
- Kawabata K., Mizuta Y., Ishihara K., et al., 2019, Chromatography, 40(3), 157–162.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Improvement of Quantitative Performance for Ibuprofen Using UV Cut-Off Filter on SPD-M40
2019|Shimadzu|Applications
LAAN-A-LC-E312B Application News High Performance Liquid Chromatography Improvement of Quantitative Performance for Ibuprofen Using UV Cut-Off Filter on SPD-M40 No. L539B Ibuprofen is an example of a nonsteroidal anti-inflammatory drug (NSAID) and is widely used as an antipyretic or analgesic…
Key words
cut, cutibuprofen, ibuprofenfilter, filteroff, offenlarged, enlargeddisabled, disabledview, viewcell, celldecomposition, decompositionmau, mauuse, usephoto, photolamp, lamparea, areaiibuprofen
Improvement of quantitative performance for photodegradable compounds using UV cut-off filter on photodiode array detector
2020|Shimadzu|Posters
Improvement of quantitative performance for photodegradable compounds using UV cut-off filter on photodiode array detector Takato Uchikata1; Davide Vecchietti1; Hidetoshi Terada1; Katsuaki Koterasawa1; Keiko Matsumoto1 1 Shimadzu Corporation, Kyoto, Japan 1. Introduction 2-2. UV cut-off filter Analysis using a photodiode…
Key words
cut, cutfilter, filteroff, offibuprofen, ibuprofendisabled, disabledenlarged, enlargedphotodiode, photodiodeview, viewmau, maupda, pdaarray, arrayarea, areaphoto, photolight, lightlamp
Ultra High Performance Liquid Chromatograph Shimadzu Nexera series - Technical Reports / Applications
2020|Shimadzu|Guides
C190-E256 Ultra High Performance Liquid Chromatograph Nexera series Technical Reports / Applications The Nexera Series: Reinvent Your Workflow Just as the internet and smartphones have changed our lifestyles, and robots with AI capabilities continue to make our lives more convenient,…
Key words
acid, acidmau, mauanalysis, analysismobile, mobilemin, mincolumn, columnphase, phaseshim, shimnexera, nexerainjection, injectionpack, packorganic, organicatp, atpacids, acidspantothenate
UHPLC Nexera series - Specifications
2023|Shimadzu|Brochures and specifications
C196-E095B Ultra High Performance Liquid Chromatograph Nexera series Specifications System Configuration UV-VIS Detector Mobile Phase Monitor Photodiode Array (PDA) Detector The monitoring device can be placed in the reservoir tray. The volume of liquid remaining in each mobile phase bottle…
Key words
cell, cellrange, rangetemperature, temperaturepressure, pressuresupply, supplywavelength, wavelengthoperating, operatingunit, unitpower, powergradient, gradientcooler, coolerfcv, fcvkit, kitflow, flowdelivery