LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Causes and Measures for Addressing Ghost Peaks in Reversed Phase HPLC Analysis

Technical notes | 2011 | ShimadzuInstrumentation
HPLC
Industries
Manufacturer
Shimadzu

Summary

Significance of Ghost Peak Analysis in Reversed Phase HPLC


Ghost peaks are unexpected signals of unclear origin in chromatograms that can interfere with the accurate identification and quantification of analytes. With modern UV detectors capable of detecting traces at low concentration levels, addressing ghost peaks has become critical for reliable data quality and efficient laboratory workflows.

Objectives and Overview


The report examines the causes of ghost peaks in reversed phase HPLC using UV detection and evaluates remedial measures. It categorizes sources of ghost peaks into instrument contamination, sample impurities, and mobile phase impurities, and introduces a sorbent cartridge, Ghost Trap DS, to trap unwanted contaminants in the mobile phase line.

Methodology and Used Instrumentation


Analyses were carried out on a Shimadzu reversed phase HPLC system consisting of an autosampler, solvent delivery pumps for aqueous and organic phases, a gradient mixer, column oven at 45 degrees C, and a UV detector set at 210 nm. Samples and mobile phases were prepared with phosphate buffer and acetonitrile mixtures under gradient elution at 0.65 mL per minute. The Ghost Trap DS cartridges of different dimensions (30 mm × 7.6 mm and 20 mm × 4.0 mm) were tested at various positions in the flow path to assess peak suppression.

Main Results and Discussion


Instrument-related ghost peaks were traced to carryover on the autosampler needle due to adsorption of analytes on its surfaces. Sample-related ghost peaks arose from co-absorbing impurities or dissolved oxygen in solvents, which was mitigated by vial cooling and degassing. Mobile phase contaminants originated from aged solvents, topped-off bottles, or tubing adsorption. Installation of Ghost Trap DS immediately downstream of the gradient mixer maximally reduced ghost peaks, confirming mobile phase as a key source in gradient conditions.

Benefits and Practical Applications


Implementing appropriate sample preparation, solvent degassing, needle rinsing protocols, and inline trapping cartridges enhances the robustness of HPLC assays. Ghost Trap DS cartridges offer a straightforward solution to remove low-level contaminants without altering existing methods, improving reproducibility and data confidence in pharmaceutical and industrial analyses.

Future Trends and Utilization


Future developments may focus on novel sorbent chemistries for broader contaminant profiles, integration of inline degassing systems, and smart autosampler designs with self-cleaning needles. Advances in system diagnostics and real-time monitoring of mobile phase purity will further minimize ghost peak occurrence.

Conclusion


A systematic approach combining instrument maintenance, sample and mobile phase management, and targeted inline trapping effectively addresses ghost peaks in reversed phase HPLC. Adopting these measures is essential for high-sensitivity applications and ensures reliable quantitative results in complex analytical workflows.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Pharmaceuticals - Technologies for Enhanced LC and SFC Workflow
Pharmaceuticals - Technologies for Enhanced LC and SFC Workflow
2018|Shimadzu|Brochures and specifications
C10G-E069 Pharmaceuticals Technologies for Enhanced LC and SFC Workflow Introduction With the rapid growth of global population, the increased incidence and prevalence of diseases, and the continuous aging society in major economies, the global demand for pharmaceuticals can only increase.…
Key words
min, minreport, reportscouting, scoutingnexera, nexerapeak, peakmobile, mobiledata, datapdea, pdeadrec, drecsystem, systemspectrum, spectrumchiralpak, chiralpakmethod, methodtime, timemau
Shim-pack HPLC Column Guidebook - High Performance Liquid Chromatography Columns
C190-E206A High Performance Liquid Chromatography Columns Shim-pack HPLC Column Guidebook Shimadzu Liquid Chromatographs Maximizing the Potential of UHPLC/HPLC Analysis Meet the Nexera X2, the most advanced UHPLC available today. The flexible system design achieves a true fusion between UHPLC and…
Key words
shim, shimpack, packparticle, particlesize, sizecolumns, columnscolumn, columnlength, lengthguidebook, guidebookguard, guardphase, phaseods, odsseries, seriesgis, gismpa, mpahplc
Prominent Features of Shimadzu UHPLC for an LC/MS Assay
C190-E282 Technical Report [White Paper] Prominent Features of Shimadzu UHPLC for an LC/MS Assay Kyoko Watanabe, Ph.D.1, Yuki Ito1, Yusuke Osaka1, Curtis Campbell, Ph.D.2 A b s tra c t: Although mass spectrometry (MS) offers high sensitivity and selectivity, the…
Key words
unit, unitcarryover, carryoverdelivery, deliverysolvent, solventgradient, gradientpressure, pressureport, portnexera, nexeraconsists, consistsautosampler, autosamplerrinse, rinsehigh, highsystem, systemmixer, mixeruhplc
Shim-pack HPLC Column Guidebook
C190-E206D High Performance Liquid Chromatography Columns Shim-pack HPLC Column Guidebook Shimadzu Liquid Chromatographs Nexera X3 model Nexera X3 is Shimadzu’s new flagship UHPLC model. Nexera X3 excels in high throughput environments requiring fast separations. The system easily handles small ID…
Key words
shim, shimpack, packcolumns, columnsparticle, particlesize, sizehplc, hplcuhplc, uhplccolumn, columnguidebook, guidebooklength, lengthseries, seriesyes, yesphase, phaseguard, guardmpa
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike