HOW LOW CAN WE GO? ANALYSIS OF ALDOSTERONE USING A HIGHLY ANALYTICALLY SENSITIVE TANDEM QUADRUPOLE MASS SPECTROMETER FOR CLINICAL RESEARCH
Posters | 2016 | WatersInstrumentation
Aldosterone is a key mineralocorticoid hormone responsible for maintaining sodium/potassium balance and blood pressure. Accurate measurement at low concentrations is critical in clinical research on aldosterone synthase (CYP11B2) inhibitors. Traditional immunoassays can suffer cross-reactivity with structurally related steroids, whereas LC-MS/MS offers greater selectivity and sensitivity.
This study compared the performance of a conventional Xevo TQ-S tandem quadrupole mass spectrometer and the new Xevo TQ-XS system, both coupled to UPLC, for quantifying aldosterone in human plasma down to 2 pg/mL using 200 µL sample volume. It also evaluated the novel UniSpray ionization against electrospray ionization (ESI) for enhancing analytical sensitivity.
Sample Preparation:
Analytical Sensitivity and Linearity:
The Waters Xevo TQ-XS coupled to UPLC demonstrates superior sensitivity, precision and linearity for plasma aldosterone analysis compared to the Xevo TQ-S. UniSpray ionization shows promising gains in signal and robustness, offering a powerful platform for clinical research requiring low-level steroid quantification.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerWaters
Summary
Importance of the Topic
Aldosterone is a key mineralocorticoid hormone responsible for maintaining sodium/potassium balance and blood pressure. Accurate measurement at low concentrations is critical in clinical research on aldosterone synthase (CYP11B2) inhibitors. Traditional immunoassays can suffer cross-reactivity with structurally related steroids, whereas LC-MS/MS offers greater selectivity and sensitivity.
Study Objectives and Overview
This study compared the performance of a conventional Xevo TQ-S tandem quadrupole mass spectrometer and the new Xevo TQ-XS system, both coupled to UPLC, for quantifying aldosterone in human plasma down to 2 pg/mL using 200 µL sample volume. It also evaluated the novel UniSpray ionization against electrospray ionization (ESI) for enhancing analytical sensitivity.
Methodology
Sample Preparation:
- Automated on a Tecan Freedom EVO 100 liquid handler.
- Protein precipitation with zinc sulfate in methanol and 0.05 % phosphoric acid.
- SPE cleanup using Waters Oasis MAX µElution plates.
- Waters ACQUITY UPLC I-Class system.
- CORTECS UPLC C18 column (2.1×100 mm).
- Water/methanol gradient at 0.3 mL/min.
Used Instrumentation
- Waters Xevo TQ-S and Xevo TQ-XS tandem quadrupole MS.
- UniSpray source with grounded probe and Coanda-effect gas flow.
- Waters ACQUITY UPLC I-Class.
- Tecan Freedom EVO 100.
Main Results and Discussion
Analytical Sensitivity and Linearity:
- LLOQ at 1 pg/mL (2.8 pmol/L) with S/N ≥10 on Xevo TQ-XS (350 fg on column).
- Over 5-fold S/N improvement and over 6-fold peak area increase versus Xevo TQ-S.
- Calibration range 15–1500 pg/mL (42–4161 pmol/L), r² > 0.9999.
- QC levels at 21, 42, 174 and 669 pg/mL showed total precision ≤7.9 % RSD on Xevo TQ-XS versus ≤8.2 % on Xevo TQ-S.
- Deming regression between systems: y = 0.98x – 5.90, no significant bias (p > 0.05).
- Altman-Bland mean bias of –4.2 %.
- Linear correlation r = 0.999.
- Provided >3× increase in S/N at 2 pg/mL versus ESI.
- Peak area reproducibility <20 % RSD at 60 fg on column.
Benefits and Practical Applications
- Enables reliable low-level aldosterone quantification for pharmacokinetic and pharmacodynamic studies.
- Reduces sample volume requirements and increases confidence in data at trace concentrations.
- UniSpray may lower maintenance downtime and extend analyte coverage.
Future Trends and Applications
- Broader validation of UniSpray for diverse steroid panels.
- Integration of high-throughput workflows for clinical bioanalysis.
- Application to additional low-abundance hormones and metabolites.
- Advancements in automation and miniaturized sample prep.
Conclusion
The Waters Xevo TQ-XS coupled to UPLC demonstrates superior sensitivity, precision and linearity for plasma aldosterone analysis compared to the Xevo TQ-S. UniSpray ionization shows promising gains in signal and robustness, offering a powerful platform for clinical research requiring low-level steroid quantification.
References
- Häggström M, Richfield D. Diagram of the pathways of human steroidogenesis. Wikiversity Journal of Medicine. 2014;1(1).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Highly Analytically-sensitive Analysis of Aldosterone in Plasma on the Xevo TQ-XS for Clinical Research
2016|Waters|Applications
[ TECHNOLOGY BRIEF ] Highly Analytically-sensitive Analysis of Aldosterone in Plasma on the Xevo TQ-XS for Clinical Research Dominic Foley Waters Corporation, Milford, MA, USA The Xevo TQ-XS provides significantly improved analytical sensitivity for aldosterone in plasma compared to the…
Key words
aldosterone, aldosteronexevo, xevoimprecision, imprecisionsensitivity, sensitivityimprovement, improvementanalytical, analyticalbrief, briefplasma, plasmaassessed, assessedsignificantly, significantlyzona, zonaserum, serummineralocorticoid, mineralocorticoidfoley, foleyclinical
Ultra-Low Level Analysis of Aldosterone in Plasma Using the Xevo TQ-XS for Clinical Research
2019|Waters|Applications
[ APPLICATION NOTE ] Ultra-Low Level Analysis of Aldosterone in Plasma Using the Xevo TQ-XS for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS ■■ ■■ Analytical selectivity of the Aldosterone is a mineralocorticoid steroid…
Key words
aldosterone, aldosteroneeqa, eqaxevo, xevomean, meanstripped, strippedclinical, clinicalplasma, plasmavanguard, vanguardlow, lowresearch, researchagreement, agreementserum, serumlevel, levelultra, ultramethod
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
2015|Waters|Applications
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK A P P L I C AT I O N B E N E F I T S ■■ Analytical selectivity improves…
Key words
aldosterone, aldosteroneuplc, uplcclinical, clinicalresearch, researchpmol, pmolplasma, plasmaanalytical, analyticalanalytically, analyticallyphysiological, physiologicalinterferences, interferenceslow, lowoasis, oasisacquity, acquitytracking, trackingsensitivity
Analysis of Corticosteroids and Androgens in Serum for Clinical Research
2017|Waters|Applications
[ APPLICATION NOTE ] Analysis of Corticosteroids and Androgens in Serum for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS ■■ ■■ Analytical selectivity of the Steroid hormones encompass a large class of small molecules…
Key words
mean, meancortisol, cortisoleqa, eqadheas, dheasandrostenedione, androstenedionetestosterone, testosteroneandrogens, androgensnmol, nmolcorticosteroids, corticosteroidsclinical, clinicaldifference, differenceagreement, agreementserum, serumresearch, researchsteroid