Evaluation of the Agilent 1260 LC system for analysis of base oils, as specified by ASTM D7419

Applications | 2015 | Agilent TechnologiesInstrumentation
HPLC
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies

Summary

Significance of the topic


Understanding the aromatic and saturated hydrocarbon content in base oils is essential for classifying lubricant quality, predicting performance, and determining commercial value. The ASTM D7419 method provides a rapid, reliable high-performance liquid chromatography (HPLC) protocol to quantify total aromatics and total saturates, guiding formulation decisions and ensuring product consistency in industrial and petrochemical settings.

Objectives and overview of the study


This study evaluates the performance of the Agilent 1260 Infinity LC system for compliance with ASTM D7419 requirements. Key aims include verifying resolution, capacity factor, signal-to-noise ratio, precision, calibration linearity, aromatic response factor, and low-level detection accuracy for aromatic content in base oils.

Used instrumentation


  • Agilent 1260 Infinity Quaternary Pump (G1311B)
  • Agilent 1260 Infinity Standard Autosampler (G1329B)
  • Agilent 1290 Infinity Column Compartment with valve drive (G1316C)
  • Agilent 1200 Quick Change 2-position/6-port Valve (G4231A)
  • Agilent 1260 Infinity Refractive Index Detector (G1362A)
  • Agilent 1260 Infinity Variable Wavelength Detector with standard flow cell (G1314F, G1314-60083)
  • OpenLab Chromatography Data System for acquisition and control
  • Columns: two ZORBAX SIL 9.4×250 mm, 5 µm (880952-201) and one ZORBAX CN 9.4×250 mm, 5 µm (B12246)
  • Mobile phase: HPLC-grade heptane at 3.0 mL/min

Methodology


Calibration and test solutions were prepared with reagent-grade hexadecane and octadecylbenzene in heptane using Class A glassware and analytical balance (±0.0001 g). The flow path employed series columns with timed backflush to isolate saturates and aromatics. Performance parameters were calculated from retention times and peak shapes according to ASTM D7419 formulas.

Main results and discussion


  • Resolution between hexadecane and octadecylbenzene peaks was 7.2 (requirement ≥5).
  • Capacity factor (kʼ) was 0.6 (requirement ≥0.4).
  • Signal-to-noise ratio at 10 µL injection was 9.4 (requirement ≥8).
  • Short-term precision (five injections) yielded RSDs of 0.07% for hexadecane and 0.41% for octadecylbenzene (requirement <0.6%).
  • Calibration curves were linear with correlation coefficients >0.999 for both analytes.
  • Aromatic response factor was 1.78 (acceptable range 1.67–1.80).
  • 0.1 mass% aromatic standard analysis showed 0.054 mass% total aromatics (requirement <0.15 mass%).

Benefits and practical applications


The validated Agilent 1260 LC system ensures reproducible quantitation of aromatics and saturates in base oils, supporting quality control, product specification compliance, and formulation optimization in lubricant manufacturing and trading environments.

Future trends and opportunities


Advances may include integration of HPLC-MS detectors for compound class identification, use of greener solvent systems, miniaturized or online monitoring systems for real-time process control, and enhanced data analytics for predictive quality assurance.

Conclusion


The Agilent 1260 LC system met or exceeded all ASTM D7419 performance criteria, demonstrating robust capability for total aromatics and saturates analysis in base oils.

References


  1. ASTM D7419-13: "Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography with Refractive Index Detection".

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of Total Aromatic Hydrocarbons and Total Saturates with the Agilent 1260 Infinity Binary LC System with RI Detection According to ASTM D7419
Determination of Total Aromatic Hydrocarbons and Total Saturates with the Agilent 1260 Infinity Binary LC System with RI Detection According to ASTM D7419 Application Note Energy & Chemicals - Petrochemicals Authors Abstract Edgar Naegele This Application Note presents a version…
Key words
hexadecane, hexadecaneoctadecylbenzene, octadecylbenzenesaturates, saturatesaromatic, aromatictotal, totalhydrocarbons, hydrocarbonsnriu, nriumodernized, modernizedaromatics, aromaticssmaller, smallerlube, lubecolumns, columnszorbax, zorbaxcolumn, columninner
Determination of Total Aromatics and Total Saturates in Lube Basestocksby HPLC-RID as per ASTM D7419
High Performance Liquid Chromatograph Nexera™ series Application News Determination of Total Aromatics and Total Saturates in Lube Basestocks by HPLC-RID as per ASTM D7419 Shailesh Damale, Jessin Mathai , Raj Makhamale, Kumar Raju , Atul Narkar Shimadzu Middle East Africa…
Key words
octadecylbenzene, octadecylbenzenesaturates, saturateshexadecane, hexadecanearomatics, aromaticslube, lubebasestocks, basestockspolars, polarstotal, totaldetector, detectorrid, ridbackflush, backflushnexera, nexeraaromatic, aromaticresponse, responsenews
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions Important Note: For reliable column performance, it is essential that Agilent publication 5990-9202EN be accessed and…
Key words
tri, tridah, dahmah, mahpetrodiesel, petrodieselarea, areadev, devaromatics, aromaticsstd, stdcalibrant, calibrantxylene, xylenearomatic, aromaticfluorene, fluorenephenanthrene, phenanthrenecyano, cyanorid
Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent 1260 Infinity Binary HPLC System with RID Detection According to IP391(2000)/ASTM D6591
Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent 1260 Infinity Binary HPLC System with RID Detection According to IP391(2000)/ASTM D6591 Application Note Energy & Chemicals - Petrochemicals Author Abstract Edgar Naegele, This Application Note shows the determination…
Key words
aromatics, aromaticsaromatic, aromatictri, trihydrocarbons, hydrocarbonsmono, monodistillates, distillatesmiddle, middlerid, riddiesel, dieselnriu, nriuarea, areapetroleum, petroleumdetermination, determinationamount, amountinjection
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike