Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Applications | 2015 | Agilent TechnologiesInstrumentation
HPLC
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Fixed-dose combination drugs often combine high- and low-dose active ingredients to achieve targeted therapeutic effects. Reliable quantification of both major and trace-level compounds in a single run is critical for quality control and regulatory compliance. The Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution addresses challenges associated with limited linear dynamic range in conventional diode array detection, enabling efficient impurity testing in complex pharmaceutical formulations.

Objectives and Study Overview


This study evaluates the performance of the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution for analyzing a fixed-dose combination drug containing vitamin C, paracetamol, caffeine, and chlorphenamine. The goal was to determine whether a single injection could reliably quantify high-dose and low-dose components and trace impurities, compared with the conventional approach requiring two separate injections.

Methodology and Instrumentation


  • Chromatographic system: Agilent 1290 Infinity II LC including high-speed pump, multisampler, multicolumn thermostat, and two diode array detectors with 60-mm, 10-mm, and 3.7-mm Max-Light cartridge cells.
  • Mobile phase: Water with 0.1% TFA (A) and acetonitrile with 0.09% TFA (B); flow rate 1.0 mL/min; gradient from 5% to 95% B over 6.5 minutes; column temperature 40 °C.
  • Sample preparation: Dissolution of capsule contents, ultrasonic extraction, filtration, serial dilution; injection volumes of 1 and 5 μL for conventional DAD and 5 μL for HDR-DAD.
  • Software: Agilent OpenLAB CDS ChemStation.


Key Results and Discussion


The conventional DAD method required two injections to keep all peaks within the linear range; high-dose compounds saturated the detector in the larger injection. In contrast, the HDR-DAD solution combined signals from two detectors with different path lengths in a single 5 μL injection, maintaining all peaks within the expanded linear range (approximately 6.7 AU). Chlorphenamine and trace impurities were reliably quantified alongside high-dose analytes. Limits of detection improved by 29–49% for key compounds using HDR-DAD compared with conventional DAD.

Benefits and Practical Applications


  • Single-injection workflow reduces sample consumption and analysis time.
  • Enhanced linear dynamic range allows simultaneous quantification of high- and low-concentration analytes and trace impurities.
  • Lower detection limits improve sensitivity for regulatory impurity profiling in fixed-dose combination drugs.


Future Trends and Potential Applications


Expanding HDR-DAD technology to other complex pharmaceutical matrices and workflow automation could further streamline impurity testing. Integration with mass spectrometry or high-resolution detectors may offer complementary selectivity and structure elucidation. Ongoing developments in microfluidics and novel detector designs hold promise for even broader dynamic range and throughput enhancements.

Conclusion


The Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution enables accurate, single-injection quantification of high-dose, low-dose, and trace-level analytes in fixed-dose combination drugs. It delivers a wider linear dynamic range, improved sensitivity, and operational efficiency compared with conventional DAD methods.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Comparison of Sensitivity and Linearity of the Agilent 1200 Infinity Series High Dynamic Range Diode Array Detector Solution and the Conventional Agilent 1290 Infinity DAD
Comparison of Sensitivity and Linearity of the Agilent 1200 Infinity Series High Dynamic Range Diode Array Detector Solution and the Conventional Agilent 1290 Infinity DAD Technical Overview Author Abstract A.G.Huesgen The Agilent 1200 Infinity Series High Dynamic Range Diode Array…
Key words
chlorphenamine, chlorphenaminehdr, hdrparacetamol, paracetamolcaffeine, caffeinedad, dadinjected, injectedvitamin, vitaminmau, maulod, lodamt, amtdiode, diodedetectors, detectorsmain, mainamount, amountfactor
Analysis of a Fixed–Dose Combination Drug Using the Agilent 1290 Infinity II Evaporative Light Scattering Detector
Analysis of a Fixed–Dose Combination Drug Using the Agilent 1290 Infinity II Evaporative Light Scattering Detector Application Note Small Molecule Pharmaceuticals and Generics Author Abstract Bettina Schuhn Evaporative light scattering (ELS) is a unique and highly sensitive technique Agilent Technologies,…
Key words
els, elsdose, dosefixed, fixedchlorphenamine, chlorphenaminedynamic, dynamicconventional, conventionaldetector, detectorparacetamol, paracetamolrange, rangevitamin, vitamincaffeine, caffeinepmt, pmthigh, highscattering, scatteringresponse
Determination of Log P for Compounds of Different Polarity Using the Agilent 1200 Infinity Series HDR‑DAD Impurity Analyzer System
Determination of Log P for Compounds of Different Polarity Using the Agilent 1200 Infinity Series HDR‑DAD Impurity Analyzer System Application Note Small Molecule Pharmaceuticals & Generics Author Abstract A.G. Huesgen The determination of log P, the distribution of a compound…
Key words
testosterone, testosteroneibuprofen, ibuprofenpindolol, pindololparacetamol, paracetamoloctanol, octanollog, logmau, mauunionized, unionizedwater, waterpartition, partitionconcentrations, concentrationslipophilicity, lipophilicitywere, werelinearity, linearityphase
Monitoring of Trace Impurities Using the Agilent 1260 Infinity II Prime Online LC System with High Dynamic Range-DAD
Application Note Small Molecule Pharma Monitoring of Trace Impurities Using the Agilent 1260 Infinity II Prime Online LC System with High Dynamic Range-DAD ×103 4.0 3.6 3.2 2.8 2.4 2.0 1.6 1.2 0.8 0.4 0 0.360 0.260 0.837 0.873 0.1…
Key words
anisylidene, anisylideneanisaldehyde, anisaldehydebyproduct, byproducteduct, eductdad, dadhdr, hdronline, onlinedynamic, dynamicreaction, reactionacetone, acetonemolecule, moleculeconcentrated, concentrateddiode, diodehigh, highsmall
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike