LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Tips and Tricks of HPLC System Troubleshooting

Presentations | 2005 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


High-performance liquid chromatography (HPLC) is a cornerstone of modern analytical science, used for separating, identifying, and quantifying compounds in pharmaceuticals, environmental samples, food safety, and more. Troubleshooting HPLC systems ensures consistent data quality, minimizes downtime, and protects costly columns and instrument components.

Objectives and Study Overview


This guide offers a structured workflow for diagnosing and resolving common HPLC issues. It emphasizes systematic checks of system suitability, method compliance, instrument settings, and identifies when and how to probe each component of the chromatographic system.

Methodology and Used Instrumentation


The troubleshooting approach is organized into steps:
  • Confirm whether system suitability tests or actual sample analyses are failing.
  • Verify that the written method is followed and instrument parameters are correct.
  • Document recent instrument changes and performance history.
  • Review all parameters simultaneously, since multiple alterations may contribute to a single problem.
Primary components examined include:
  • Pump
  • Injector/autosampler
  • Column and guard column
  • In-line filters and tubing
  • Detector (UV/DAD/RID) and flow cell
  • Data system/integrator

Main Results and Discussion


Pressure anomalies often result from blocked frits, guard column contamination, or leaks. Back-flushing, solvent washes with strong organic solvents, and replacing or preventing frit clogging via in-line filters and sample filtration restore nominal back pressure.
Peak shape defects—such as splitting, fronting, tailing, or broadening—originate from disrupted flow paths, sample solvent strength mismatches, secondary surface interactions (e.g., silanol activity), column aging or damage, overloading, and extra-column dispersion. Remedies include matching injection solvent strength to the mobile phase, adjusting mobile phase pH or adding modifiers (e.g. triethylamine), performing acid washes to remove metal contamination, optimizing detector response time, and minimizing dwell and extracolumn volumes by using short, small-bore matched fittings and low-volume flow cells.
Retention shifts and selectivity changes can be chemical (buffer pH, ionic strength, mobile phase composition, lot-to-lot column variability) or physical (insufficient equilibration, gradient delay volume, column degradation). Consistent use of well-buffered mobile phases within the pH tolerance of the column, verifying dwell volume compensation, and conditioning new columns and mobile phases are key to reproducible retention.
Detector issues such as lamp aging, improper response time, baseline noise from air bubbles or contamination, and drift during gradients are addressed through regular lamp replacement, detector maintenance, optimized sampling rates, mobile phase degassing, and stable temperature control.

Benefits and Practical Applications of the Method


Implementing this systematic troubleshooting framework reduces analytical downtime, extends column lifetime, improves method robustness, and enhances confidence in quantitative and qualitative HPLC results across quality control, research, and industrial laboratories.

Future Trends and Opportunities


Advances in real-time instrument diagnostics, AI-driven troubleshooting assistants, improved high-pH-stable stationary phases, and novel in-line micro-filter materials will streamline HPLC maintenance. Predictive maintenance algorithms will anticipate system failures before they impact analyses.

Conclusion


A comprehensive, stepwise approach to HPLC troubleshooting—covering pressure, peak shape, retention, and detection—enables analysts to pinpoint root causes effectively. Preventative practices and informed adjustments to solvents, buffers, and hardware settings ensure reliable, high-quality chromatographic performance.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Tips and Tricks of HPLC Separations and Troubleshooting
Tips and Tricks of HPLC Separations and Troubleshooting
2010|Agilent Technologies|Presentations
Advanced Topic: Tips and Tricks of HPLC Separations and Troubleshooting Rita Steed LC Advanced Topics Seminar Series August 25, 2010 Page 1 Troubleshooting Steps You Have Recognized There is a Problem! How Do You Fix It? • 1st Did System…
Key words
column, columnmin, mintime, timefrit, frittailing, tailingpeak, peakmobile, mobileshape, shapecontamination, contaminationphase, phasebuffers, bufferspressure, pressureproblems, problemstitle, titlepresentation
HPLC Column Troubleshooting: Is It Really The Column?
HPLC Column Troubleshooting: Is It Really The Column?
2010|Agilent Technologies|Presentations
HPLC Column Troubleshooting: Is It Really The Column? Agilent Technologies Technologies, Inc. Inc Rita Steed Application Engineer J January 22 22, 2010 Page 1 Agilent Restricted mA U Troubleshooting in HPLC 20 00 15 00 10 5 00 0 0…
Key words
restricted, restrictedagilent, agilentcolumn, columnmin, minmobile, mobiletailing, tailingtime, timephase, phasechange, changeretention, retentionplugged, pluggedplates, plateslot, lotchanges, changespeak
LC Column Troubleshooting - Isolating the Source of the Problem
LC Column Troubleshooting Isolating the Source of the Problem Rita Steed December 12, 2013 Page 1 What Do We Troubleshoot The typical LC troubleshooting approach asks the questions: – What’s wrong with the column? – What’s wrong with the instrument?…
Key words
column, columntailing, tailingmin, minmau, mauvolume, volumemobile, mobilefittings, fittingspeak, peakphase, phaseplugged, pluggedpeaks, peakstime, timefrit, fritred, redextra
The Chromatography Detective: Troubleshooting Tips  & Tools for LC & LCMS
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS Rita Steed LC Columns Application Engineer January 31, 2018 What do you do when..... ➢ Your chromatography changes • Peak shape • Retention ➢ You can’t reproduce a method…
Key words
beginner, beginnerpitfalls, pitfallsavoiding, avoidingcolumn, columndwell, dwellvolume, volumemobile, mobilephase, phasepeak, peakecv, ecvinstrument, instrumentshape, shapetailing, tailingretention, retentionconnections
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike