Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS
Applications | 2016 | Agilent TechnologiesInstrumentation
Accurate measurement of plasma metanephrines and 3-methoxytyramine plays a critical role in diagnosing catecholamine-secreting tumors such as pheochromocytoma and paraganglioma. These metabolites also serve as biomarkers for sympathetic nervous system activity and are essential for clinical research, quality control, and therapeutic monitoring in endocrinology and oncology.
This study aimed to develop and validate a highly sensitive and specific LC/MS/MS method for simultaneous quantitation of metanephrine, normetanephrine, and 3-methoxytyramine in human plasma. Key goals included achieving:
Sample preparation utilized off-line solid phase extraction with Agilent SampliQ WCX cartridges (30 mg, 1 mL; p/n 5982-3513) to remove matrix interferences. The LC system consisted of an Agilent 1290 Infinity binary pump, thermostat-controlled well-plate sampler, and a temperature-controlled column compartment set at 40 °C. Chromatographic separation was achieved on an Agilent Pursuit PFP column (2×150 mm, 3 µm; p/n A3051150X020) with MetaGuard guard column. Mobile phases were 0.2 % formic acid in water (A) and methanol (B) with a programmed gradient at 0.3–0.5 mL/min.
Mass spectrometric detection employed an Agilent 6460 triple quadrupole with Jet Stream ESI+ ionization. Key MS parameters included 350 °C gas temperature, 5 L/min drying gas, 400 °C sheath gas at 11 L/min, and 3,000 V capillary voltage. Multiple reaction monitoring transitions were optimized for each analyte and corresponding deuterated internal standards. Data processing was performed with Agilent MassHunter Quantitative Analysis (B.07.00) using 1/x weighting.
Chromatographic separation resolved closely eluting analytes and catecholamines, preventing cross-interference. Calibration curves over 15.63–10,000 pg/mL exhibited linearity with R² values above 0.9996. Matrix effects ranged from 36 % to 78 % but were effectively compensated by internal standards. Recovery efficiencies exceeded 87 % for all compounds. Intra- and inter-day precision studies showed CVs below 6 % and accuracy within 85–117 %. ChromSystems control samples validated method robustness with CV <6 %.
This validated protocol offers:
Advances may include automated high-throughput SPE systems, miniaturized nano-LC/MS platforms, expanded metabolite panels for broader endocrine profiling, and integration with high-resolution mass spectrometry. Machine learning algorithms are expected to enhance data analysis, enabling real-time decision support and personalized medicine applications.
The presented LC/MS/MS method combines solid phase extraction with triple quadrupole detection to deliver a sensitive, precise, and linear assay for plasma metanephrines and 3-methoxytyramine. Its robustness and reliability make it well suited for clinical research, diagnostic support, and quality assurance in analytical laboratories.
1. Whiting MJ. Simultaneous measurement of urinary metanephrines and catecholamines by liquid chromatography with tandem mass spectrometric detection. Ann Clin Biochem. 2009;46:129–136.
2. Gabler JA. A sensitive and interference-free liquid chromatography tandem mass spectrometry method for measuring metanephrines in plasma. J Chromatogr Separat Techniq. 2012;S2.
Sample Preparation, Consumables, LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Accurate measurement of plasma metanephrines and 3-methoxytyramine plays a critical role in diagnosing catecholamine-secreting tumors such as pheochromocytoma and paraganglioma. These metabolites also serve as biomarkers for sympathetic nervous system activity and are essential for clinical research, quality control, and therapeutic monitoring in endocrinology and oncology.
Objectives and Study Overview
This study aimed to develop and validate a highly sensitive and specific LC/MS/MS method for simultaneous quantitation of metanephrine, normetanephrine, and 3-methoxytyramine in human plasma. Key goals included achieving:
- Low functional sensitivity to detect trace levels.
- A broad dynamic range for clinical relevance.
- High reproducibility (CV <6%).
- Excellent linearity (R² >0.9996).
Methodology and Instrumentation
Sample preparation utilized off-line solid phase extraction with Agilent SampliQ WCX cartridges (30 mg, 1 mL; p/n 5982-3513) to remove matrix interferences. The LC system consisted of an Agilent 1290 Infinity binary pump, thermostat-controlled well-plate sampler, and a temperature-controlled column compartment set at 40 °C. Chromatographic separation was achieved on an Agilent Pursuit PFP column (2×150 mm, 3 µm; p/n A3051150X020) with MetaGuard guard column. Mobile phases were 0.2 % formic acid in water (A) and methanol (B) with a programmed gradient at 0.3–0.5 mL/min.
Mass spectrometric detection employed an Agilent 6460 triple quadrupole with Jet Stream ESI+ ionization. Key MS parameters included 350 °C gas temperature, 5 L/min drying gas, 400 °C sheath gas at 11 L/min, and 3,000 V capillary voltage. Multiple reaction monitoring transitions were optimized for each analyte and corresponding deuterated internal standards. Data processing was performed with Agilent MassHunter Quantitative Analysis (B.07.00) using 1/x weighting.
Key Results and Discussion
Chromatographic separation resolved closely eluting analytes and catecholamines, preventing cross-interference. Calibration curves over 15.63–10,000 pg/mL exhibited linearity with R² values above 0.9996. Matrix effects ranged from 36 % to 78 % but were effectively compensated by internal standards. Recovery efficiencies exceeded 87 % for all compounds. Intra- and inter-day precision studies showed CVs below 6 % and accuracy within 85–117 %. ChromSystems control samples validated method robustness with CV <6 %.
Benefits and Practical Applications
This validated protocol offers:
- High sensitivity and specificity for trace metabolites.
- Robust performance suitable for clinical and research laboratories.
- Compatibility with routine QA/QC workflows.
- Potential for integration into diagnostic panels.
Future Trends and Opportunities
Advances may include automated high-throughput SPE systems, miniaturized nano-LC/MS platforms, expanded metabolite panels for broader endocrine profiling, and integration with high-resolution mass spectrometry. Machine learning algorithms are expected to enhance data analysis, enabling real-time decision support and personalized medicine applications.
Conclusion
The presented LC/MS/MS method combines solid phase extraction with triple quadrupole detection to deliver a sensitive, precise, and linear assay for plasma metanephrines and 3-methoxytyramine. Its robustness and reliability make it well suited for clinical research, diagnostic support, and quality assurance in analytical laboratories.
References
1. Whiting MJ. Simultaneous measurement of urinary metanephrines and catecholamines by liquid chromatography with tandem mass spectrometric detection. Ann Clin Biochem. 2009;46:129–136.
2. Gabler JA. A sensitive and interference-free liquid chromatography tandem mass spectrometry method for measuring metanephrines in plasma. J Chromatogr Separat Techniq. 2012;S2.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Plasma Catecholamines by LC/MS/MS
2016|Agilent Technologies|Applications
Plasma Catecholamines by LC/MS/MS Using Agilent Captiva NDLipids, BondElut PBA SPE, 1290 Infinity LC, and 6460 Triple Quadrupole LC/MS Application Note Clinical Research Authors Abstract Linda Côté and Christophe Deckers We developed a highly sensitive and specific LC/MS/MS method to…
Key words
epinephrine, epinephrinedopamine, dopaminenorepinephrine, norepinephrineagilent, agilentplasma, plasmaefficiencies, efficienciesndlipids, ndlipidswere, wereresponses, responsesextracted, extractedchromsystems, chromsystemsext, extnitrogen, nitrogencatecholamines, catecholaminesconcentration
Urinary Catecholamines, Metanephrines, and 3-Methoxytyramine in a Single LC/MS/MS Run
2016|Agilent Technologies|Applications
Urinary Catecholamines, Metanephrines, and 3-Methoxytyramine in a Single LC/MS/MS Run Using Agilent Bond Elut Plexa SPE, 1290 Infinity LC, and 6460 Triple Quadrupole LC/MS Application Note Clinical Research Authors Abstract Linda Côté and Christophe Deckers We developed a highly sensitive…
Key words
metanephrine, metanephrineepinephrine, epinephrinenorepinephrine, norepinephrinedopamine, dopaminenormetanephrine, normetanephrinenmol, nmolresponses, responsesintraday, intradayconcentration, concentrationfree, freetotal, totalrelative, relativeboronate, boronateurine, urineagilent
Rapid and Simultaneous Analysis of Plasma Catecholamines and Metanephrines Using Mixed-Mode SPE and Hydrophilic Interaction Chromatography (HILIC) for Clinical Research
2014|Waters|Applications
Rapid and Simultaneous Analysis of Plasma Catecholamines and Metanephrines Using Mixed-Mode SPE and Hydrophilic Interaction Chromatography (HILIC) for Clinical Research Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA A P P L I…
Key words
metanephrines, metanephrinescatecholamines, catecholamineshilic, hilicdopamine, dopaminemetanephrine, metanephrinenormetanephrine, normetanephrineplasma, plasmahydrophilic, hydrophilicmean, meaninteraction, interactionepinephrine, epinephrinenorepinephrine, norepinephrinenmt, nmtmixed, mixedspe
Reversed-phase UPLC-MS/MS Analysis of Plasma Catecholamines and Metanephrines For Clinical Research
2020|Waters|Applications
[ TECHNOLOGY BRIEF ] Reversed-phase UPLC-MS/MS Analysis of Plasma Catecholamines and Metanephrines For Clinical Research Jonathan P. Danaceau,¹ Li Qianqian,² Stephen Balloch,³ and Lisa J. Calton³ ¹Waters Corporation, Milford, MA, USA; ²Waters Solutions Centre, Beijing, China; ³Waters Corporation, Wilmslow, UK…
Key words
catecholamines, catecholaminesuplc, uplchss, hssmetanephrines, metanephrinesplasma, plasmapfp, pfpacquity, acquityreversed, reversedrecoveries, recoveriesmetanephrine, metanephrinebrief, briefphase, phaseclinical, clinicalwcx, wcxbatch