Detection of UV-transparent Compounds by Addition of a Mass Detector to an Existing High Performance Liquid Chromatography System with Photodiode Array Detection
Technical notes | 2014 | WatersInstrumentation
The detection of analytes that lack UV chromophores represents a significant challenge in liquid chromatography. Many compounds of pharmaceutical and biological interest elude conventional UV detection, leading to incomplete sample characterization. Integrating mass detection provides an orthogonal approach to identify and quantify UV-transparent species, enhancing the analytical scope of HPLC methods.
Separations were performed on an XSelect CSH C18 column (3.5 µm, 2.1 x 100 mm) at 300 μL/min using an ACQUITY Isocratic Solvent Manager. A splitting restrictor directed flow to both a PDA detector monitored at 220 nm and an ACQUITY QDa mass detector operating in single ion recording mode. Data acquisition and unified reporting were handled by Empower CDS software.
Memantine and its related impurities exhibit negligible UV absorbance at 220 nm but are clearly detected by the mass detector with selected ion recording for m/z 180.08, 208.03, and 221.97. Retention time reproducibility (n = 5) remained below 0.25 %RSD, while peak area %RSD values stayed within 3.00 %RSD for all analytes, confirming robust performance.
The complementary use of mass and UV detection enables reliable identification of non-chromophoric compounds without the need for time-consuming derivatization. The compact ACQUITY QDa detector integrates seamlessly with existing HPLC-PDA setups and Empower CDS, streamlining workflows in pharmaceutical quality control and research laboratories.
Advances in miniaturized mass spectrometry are expected to further enhance sensitivity and accessibility of orthogonal detection methods. Potential developments include automated method optimization, broader ionization techniques for diverse analyte classes, and real-time data processing powered by artificial intelligence.
Adding a mass detector to an HPLC-PDA system significantly improves the detection of UV-transparent compounds, ensuring comprehensive sample analysis. The approach offers reliable reproducibility and seamless data integration, making it a valuable asset in analytical chemistry.
HPLC, LC/MS, LC/SQ
IndustriesManufacturerWaters
Summary
Importance of the Topic
The detection of analytes that lack UV chromophores represents a significant challenge in liquid chromatography. Many compounds of pharmaceutical and biological interest elude conventional UV detection, leading to incomplete sample characterization. Integrating mass detection provides an orthogonal approach to identify and quantify UV-transparent species, enhancing the analytical scope of HPLC methods.
Objectives and Study Overview
- Evaluate the addition of a compact mass detector to an established HPLC system equipped with a photodiode array detector.
- Demonstrate the capability to detect UV-transparent compounds, exemplified by memantine and its impurities.
- Assess reproducibility and sensitivity improvements over UV-only detection.
Methodology and Used Instrumentation
Separations were performed on an XSelect CSH C18 column (3.5 µm, 2.1 x 100 mm) at 300 μL/min using an ACQUITY Isocratic Solvent Manager. A splitting restrictor directed flow to both a PDA detector monitored at 220 nm and an ACQUITY QDa mass detector operating in single ion recording mode. Data acquisition and unified reporting were handled by Empower CDS software.
Main Results and Discussion
Memantine and its related impurities exhibit negligible UV absorbance at 220 nm but are clearly detected by the mass detector with selected ion recording for m/z 180.08, 208.03, and 221.97. Retention time reproducibility (n = 5) remained below 0.25 %RSD, while peak area %RSD values stayed within 3.00 %RSD for all analytes, confirming robust performance.
Contributions and Practical Applications
The complementary use of mass and UV detection enables reliable identification of non-chromophoric compounds without the need for time-consuming derivatization. The compact ACQUITY QDa detector integrates seamlessly with existing HPLC-PDA setups and Empower CDS, streamlining workflows in pharmaceutical quality control and research laboratories.
Future Trends and Opportunities
Advances in miniaturized mass spectrometry are expected to further enhance sensitivity and accessibility of orthogonal detection methods. Potential developments include automated method optimization, broader ionization techniques for diverse analyte classes, and real-time data processing powered by artificial intelligence.
Conclusion
Adding a mass detector to an HPLC-PDA system significantly improves the detection of UV-transparent compounds, ensuring comprehensive sample analysis. The approach offers reliable reproducibility and seamless data integration, making it a valuable asset in analytical chemistry.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Waters ACQUITY QDa Detector - QC APPLICATIONS COMPENDIUM - EDITION 2
2018|Waters|Guides
[ APPLICATION NOTEBOOK ] ACQUITY QDa Detector QC APPLICATIONS COMPENDIUM EDITION 2 Dear Colleague The 2013 introduction of the ACQUITY™ QDa™ Detector was a breakthrough in Chromatography and Mass Spectrometry integration. It was the fulfilment of a vision 20 years…
Key words
acquity, acquityqda, qdauplc, uplcdetector, detectormass, massarc, arcdetection, detectionbound, boundcetrimonium, cetrimoniumusing, usingminutes, minuteswaters, watersintensity, intensityclass, classanalysis
Benefits of Using ACQUITY QDa Mass Detection for Quantitative Analysis of Non-Chromophoric Memantine HCl in Tablet Formulation
2014|Waters|Applications
Benefits of Using ACQUITY QDa Mass Detection for Quantitative Analysis of Non-Chromophoric Memantine HCl in Tablet Formulation Margaret Maziarz, Mark Wrona, and Sean M. McCarthy Waters Corporation, Milford, MA, USA A P P L I C AT I O N…
Key words
memantine, memantinechromophoric, chromophoricqda, qdahcl, hclacquity, acquitydiluent, diluentdetection, detectionuplc, uplcnon, nonquantitative, quantitativeintensity, intensitytablet, tabletbenefits, benefitsusing, usingmass
Repeatability and Quantitation in a Method Using the Alliance HPLC System with the 2998 Photodiode Array Detector and the ACQUITY QDa Detector
2014|Waters|Applications
Repeatability and Quantitation in a Method Using the Alliance HPLC System with the 2998 Photodiode Array Detector and the ACQUITY QDa Detector Aparna Chavali, Thomas E. Wheat, Patricia McConville Waters Corporation, Milford, MA, USA A P P L I C…
Key words
ism, ismalliance, allianceirbesartan, irbesartanqda, qdapda, pdaapi, apiarea, areaacquity, acquityrsd, rsdrca, rcadetector, detectorhplc, hplcsolvent, solventtee, teesplitter
APPLICATION NOTEBOOK - IMPURITIES - Developing Safe and Effective Drugs
2016|Waters|Guides
[ APPLICATION NOTEBOOK ] IMPURITIES Developing Safe and Effective Drugs INT RODUCTION NAVIGATION T he pharmaceutical industry is integral in maintaining public health by providing therapeutic and preventative medicines. T he cornerstones of bringing new medicines to market are quality,…
Key words
home, homeuplc, uplcimpurities, impuritiesziprasidone, ziprasidoneusp, uspacquity, acquityimpurity, impuritymethod, methodtioconazole, tioconazoleminutes, minutessimvastatin, simvastatinhplc, hplcusing, usingrelated, relatedanalysis