Waters Xevo G2-XS QTof - INSTRUMENT SPECIFICATIONS

Brochures and specifications | 2017 | WatersInstrumentation
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
Industries
Manufacturer
Waters

Summary

Significance of the Topic


The advent of high–resolution accurate mass spectrometry has transformed analytical chemistry by enabling detailed qualitative and quantitative analysis of complex samples. Instruments like the Xevo G2-XS QTof deliver superior mass accuracy, dynamic range and sensitivity, supporting applications in proteomics, metabolomics, small-molecule research and quality control.

Objectives and Overview of the Instrument


This summary presents the key specifications and performance characteristics of the Xevo G2-XS QTof mass spectrometer. It highlights system hardware and software features, acquisition modes, mass ranges, resolution capabilities and sensitivity metrics that define the instrument’s analytical power.

Methodology and Instrumentation


The Xevo G2-XS QTof integrates several advanced technologies:
  • StepWave™ ion optics and XS Collision Cell for robust ion transmission and MS/MS fragmentation
  • High-resolution quadrupole (MS1) and oa-Tof analyser (MS2) achieving >40,000 FWHM resolution
  • Versatile ZSpray™ dual-orthogonal API sources supporting ESI, APCI, APPI, nanoFlow™ ESI and other modes
  • Automated vacuum system with differential pumping and digital monitoring
  • IntelliStart™ software for automated system checks, calibration and sample delivery

Main Performance Characteristics and Discussion


Acquisition flexibility includes conventional MS/MS, MS E for simultaneous precursor and fragment data, FastDDA for automated data-dependent MS/MS and Tof-MRM for targeted quantitation. The mass range extends from m/z 20 to 100,000. Mass accuracy better than 1 ppm RMS and dynamic range of four orders of magnitude support trace-level detection. Sensitivity assessments demonstrate >80,000 cps for leucine enkephalin (50 pg/µL) in positive mode and >110,000 cps for raffinose (500 pg/µL) in negative mode. MS/MS sensitivity yields >7,500 cps for y-ions of [Glu1]-Fibrinopeptide B at 100 fmol/µL.

Benefits and Practical Applications


The Xevo G2-XS QTof’s combination of high resolution, mass accuracy and rapid scan rates facilitates confident compound identification, structural elucidation and quantitative analysis across pharmaceutical development, environmental monitoring, food safety and clinical research. Automated workflows ensure reproducibility and ease of use for expert and non-expert operators alike.

Future Trends and Opportunities


Ongoing developments may include enhanced ion mobility integration, expanded dynamic range technologies, real-time data processing and AI-driven acquisition strategies. Miniaturized front-ends and novel ambient ionization sources will further broaden application areas in field analysis and single-cell studies.

Conclusion


The Xevo G2-XS QTof represents a state-of-the-art platform for high-resolution mass spectrometry, combining robust hardware with intelligent software to deliver comprehensive qualitative and quantitative data. Its versatility and performance make it a valuable tool in modern analytical laboratories.

References


  • Waters Corporation. Xevo G2-XS QTof Instrument Specifications. Technical Bulletin, March 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Waters Xevo TQD - INSTRUMENT SPECIFICATIONS
Waters Xevo TQD - INSTRUMENT SPECIFICATIONS
2019|Waters|Brochures and specifications
[ INSTRUMENT SPECIFICATIONS ] Xevo TQD The Waters™ Xevo™ TQD is an advanced benchtop tandem quadrupole mass detector designed for ultra high performance LC-MS/MS applications. With a wide range of ionization options including Atmospheric Pressure Photo Ionization (APPI), Atmospheric Pressure…
Key words
mrm, mrmspecifications, specificationsscan, scantqd, tqdion, ionxevo, xevoautomated, automatedacquisition, acquisitionunispray, unisprayradar, radarinstrument, instrumentproduct, productscheduling, schedulingsir, sirmodes
Waters XEVO G2-XS QTOF
Waters XEVO G2-XS QTOF
2019|Waters|Brochures and specifications
[ X E V O G 2 -X S QT O F ] EXT RAORDINARY POSSIBILITIES FOR YOUR EV ERYDAY ANALYSIS EXTREME SENSITIVITY, EXTRAORDINARY SELECTIVITY, Whether you’re a scientist who wants to truly understand your samples, or a lab manager…
Key words
uplc, uplcfastdda, fastddaatmospheric, atmosphericyour, yourintellistart, intellistartsimplicity, simplicityquantof, quantoftransmission, transmissionarchitecture, architectureion, ionsystem, systemseamlessly, seamlesslyapci, apcitof, tofionization
Waters Xevo TQ-S - INSTRUMENT SPECIFICATIONS
Waters Xevo TQ-S - INSTRUMENT SPECIFICATIONS
2017|Waters|Brochures and specifications
[ INSTRUMENT SPECIFICATIONS ] Xevo TQ-S Xevo® TQ-S is an advanced benchtop tandem quadrupole mass spectrometer, designed to meet the needs of your most demanding quantitative LC-MS/MS applications. Xevo TQ-S features revolutionary off-axis StepWave™ ion transfer optics that dramatically increases…
Key words
mrm, mrmscanwave, scanwaveion, ionspecifications, specificationsscan, scanxevo, xevoacts, actstrigger, triggerscanwaveenhanced, scanwaveenhancedproduct, productautomated, automatedacquisition, acquisitionvacuum, vacuumcollection, collectionautomatic
Waters Xevo TQ Absolute - instrument specifications
Waters Xevo TQ Absolute - instrument specifications
2022|Waters|Brochures and specifications
[ INSTRUMENT CAPABILITIES ] Xevo TQ Absolute Xevo™ TQ Absolute is an ultimate performance benchtop tandem quadrupole mass spectrometer which combines Absolute performance, efficiency, productivity, and confidence to give you the Absolute Power to quantitate your most challenging compounds and…
Key words
absolute, absolutemrm, mrmcapabilities, capabilitiesvacuum, vacuumscan, scanunispray, unisprayradar, radarperformance, performancexevo, xevoneutral, neutralion, ionactively, activelymodes, modessystem, systemautomated
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike