LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Analysis of Amphoteric Surfactants using LC-MS

Applications |  | ShimadzuInstrumentation
LC/MS, LC/SQ
Industries
Energy & Chemicals
Manufacturer
Shimadzu

Summary

Importance of the Topic


Amphoteric surfactants, such as fatty acid amidopropylbetaine, are widely used in detergents and industrial formulations due to their ability to reduce surface tension and adapt charge according to pH. Their persistence in domestic and industrial wastewater raises environmental and ecotoxicological concerns, driving the need for sensitive and selective analytical methods to monitor these compounds and protect ecosystems.

Objectives and Overview of the Study


This work demonstrates a liquid chromatography–mass spectrometry (LC-MS) method for the separation, identification, and quantification of amphoteric surfactant homologues ranging from C8:0 to C18:0 fatty acid amidopropylbetaine. The study presents chromatographic behavior, mass spectral characteristics, and performance parameters essential for environmental monitoring applications.

Methodology


An isocratic-gradient reversed-phase LC method was employed with water/acetic acid and methanol mobile phases. The chromatographic gradient enabled baseline resolution of homologues within a 45-minute run. Detection was carried out in positive APCI mode, scanning m/z 50–800 with 1.5 s per scan, allowing simultaneous qualitative and quantitative analysis of multiple chain lengths.

Used Instrumentation


  • LC Column: Shimadzu Shim-pack VP-ODS (2.0 mm I.D. × 150 mm), 40 °C
  • Mobile Phase A: Water + 0.2% acetic acid; B: Methanol
  • Gradient: 10% B to 100% B over 15 min; hold to 35 min; re-equilibrate to 10% B by 45 min
  • Flow Rate: 0.2 mL/min; Injection Volume: 5 µL
  • APCI Source: Probe +4.5 kV; probe 400 °C; CDL 230 °C; nebulizing gas 2.5 L/min; CDL −40 V; Q-array +37 V
  • MS Scan Range: m/z 50–800 (1.5 s/scan)

Main Results and Discussion


Chromatographic separation achieved distinct retention for each homologue. Retention times increased with carbon number:
  • C8:0 ~10 min
  • C10:0 ~12.5 min
  • C12:0 ~15 min
  • C14:0 ~17.5 min
  • C16:0 ~20 min
  • C18:1 ~21 min; C18:0 ~22.5 min
The APCI-MS spectra exhibited strong [M+H]+ ions and characteristic fragment ions corresponding to headgroup cleavage and fatty acid moieties. For example, C12:0 amidopropylbetaine showed prominent ions at m/z 285, 329, and 240, facilitating unequivocal identification. Signal intensity and peak shape confirmed method suitability for quantitative work.

Benefits and Practical Applications


This LC-APCI-MS approach offers high specificity and sensitivity for amphoteric surfactant analysis in complex matrices. It supports environmental monitoring, wastewater quality control, and industrial process validation. Rapid homologue profiling enables regulators and laboratories to track pollutant loads and comply with discharge regulations.

Future Trends and Opportunities


Advances may include coupling to high-resolution MS for non-target screening of novel surfactant derivatives, automated sample preparation for high throughput, and adoption of greener solvents or shorter columns to reduce analysis time. Integration with data analytics can further enhance pattern recognition and environmental risk assessment.

Conclusion


The presented LC-MS method reliably separates and identifies amphoteric surfactant homologues from C8 to C18 chains using APCI in positive mode. Its robust performance and detailed mass spectral information make it a valuable tool for environmental and industrial surfactant analysis.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Simultaneous Analysis of Irganox Polymer Additive using LC-MS
LC-MS Application Data Sheet No. 048 Simultaneous Analysis of Irganox Polymer Additive using LC-MS acid. To solve this problem, in this example, a GPC column was used to prevent component adsorption and ionization was conducted under neutral mobile phase conditions.…
Key words
cdl, cdlvoltage, voltagearray, arrayirganox, irganoxdeformation, deformationtemperature, temperaturemarketing, marketingflow, flowserious, seriousmobile, mobiledivision, divisioncolumn, columngpc, gpcadditive, additivesheet
Analysis of penicillin analogs using LC-MS
LC-MS Application Data Sheet No. 034 Analysis of penicillin analogs using LC-MS types that are widely used in medicine, including β-lactams (penicillins, cephalosporins, etc.), macrolides, and tetracyclines. This data sheet introduces an example of the analysis of typical synthetic penicillins…
Key words
cooh, coohconc, conccdl, cdlflucroxacillin, flucroxacillinarea, areasheet, sheetticarcillin, ticarcillinkills, killspenicillins, penicillinsmicroorganism, microorganismkill, killdeprotonated, deprotonatedhinder, hinderampicillin, ampicillinamoxicillin
Analysis of Microcystins using LS-MS (No. 2)
LC-MS Application Data Sheet No. 049 Analysis of Microcystins using LS-MS (No. 2) Microcystins (Fig. 1) are liver toxins (carcinogenic promoters) produced by blue-green algae (cyanobacteria) in outbreaks of eutrophication. A 2001 revision of Japanese Drinking Water Test Method in…
Key words
microcystin, microcystinmicrocystins, microcystinstenth, tenthcdl, cdlvoltage, voltagecooh, coohconc, concmethyldehydroalanine, methyldehydroalaninearray, arraydrinking, drinkingpromoters, promoterscyanobacteria, cyanobacteriaarea, areaalgae, algaechromatograms
Analysis of Bisphenol A and Nonylphenol using LC-MS
LC-MS Application Data Sheet No. 032 Analysis of Bisphenol A and Nonylphenol using LC-MS exogenous endocrine disrupter compounds. These phenols are generally derivatized using trimethylsilyl (TMS) reagent before analysis by GC-MS. However, this data sheet introduces the direct analysis of…
Key words
bisphenol, bisphenolnonylphenol, nonylphenolcdl, cdlsheet, sheetconc, concvoltage, voltagearray, arraytemperature, temperatureprobe, probemean, meantrimethylsilyl, trimethylsilylmarketing, marketingmin, mintms, tmsphenols
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike