UPC2 Method Development for Achiral Impurity Analysis

Applications | 2013 | WatersInstrumentation
LC/MS, SFC, LC/SQ
Industries
Pharma & Biopharma
Manufacturer
Waters

Summary

Importance of the Topic


Understanding impurity profiles in pharmaceuticals is critical for ensuring safety, efficacy, and regulatory compliance. Orthogonal analytical techniques such as UPC2 provide complementary selectivity to reversed-phase LC, aiding in the identification of unknown or late‐eluting impurities while reducing solvent use and analysis time.

Objectives and Study Overview


The study aimed to develop a systematic UPC2 method for achiral impurity analysis of metoclopramide and its related substances. Key goals included optimizing stationary phase selection, mobile phase composition, gradient slope, and additive effects to achieve high resolution and robust peak shapes across diverse functional groups.

Used Instrumentation


  • ACQUITY UPC2 System with PDA (200–410 nm) and SQD MS (150–1200 Da, ESI+/-)
  • ACQUITY UPC2 BEH 2-EP column (3.0 × 100 mm, 1.7 µm)
  • Empower 3 Software for data acquisition and processing
  • ACQUITY SQD Mass Spectrometer for peak tracking and specificity

Methodology


A resolution solution containing metoclopramide and eight related impurities (0.1% w/w) was prepared. Initial screening compared four stationary phases and four modifiers (MeOH, MeOH + 0.5% formic acid, MeOH + 2 g/L ammonium formate, MeOH + 0.5% triethylamine) under a gradient of 5→30% B over 5 min (hold 1 min). Further experiments varied gradient slope (6, 12, 35 min) and mixed MeOH with ACN (50:50 to 100:0) to assess elution strength. Additive concentrations were adjusted, including combined use of ammonium formate and formic acid, to refine peak shape across amine, hydroxyl, ester, and acid functionalities.

Main Results and Discussion


  • Column Screening: BEH 2-EP delivered optimal peak shape and resolution. Other phases showed selectivity issues (e.g., split peaks on CSH Fluoro-Phenyl).
  • Gradient Slope: Extending run time had minimal effect on resolution except for specific peak pairs.
  • Modifier Composition: Blending ACN into MeOH expanded separation space and improved resolution of late-eluting analytes.
  • Additive Effects: Combining 1 g/L ammonium formate with 3% formic acid yielded balanced peak shapes across all impurities.
  • Specificity Assessment: Final method resolved known and unknown impurities in both control and degraded metoclopramide samples.

Benefits and Practical Applications


The optimized UPC2 approach offers rapid, high-resolution separations with reduced organic solvent consumption and orthogonal selectivity. It supports impurity profiling during drug development, stability testing, and quality control, facilitating regulatory compliance and decision‐making.

Future Trends and Opportunities


Further exploration of stationary phase–solute interactions will inform automated method development workflows. Integration with advanced MS detection and AI‐driven optimization algorithms may accelerate impurity identification and streamline regulatory submissions.

Conclusion


A systematic UPC2 method for achiral impurity analysis of metoclopramide has been established by optimizing column type, mobile phase composition, gradient conditions, and additives. The protocol delivers robust resolution, specificity, and peak integrity suitable for pharmaceutical impurity profiling.

References


  1. Jones MD et al. Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC2/MS). Waters Application Note. 2012.
  2. Jones MD et al. Impurity Profiling Using UPC2/MS. Waters Application Note. 2013.
  3. West C, Lesellier E. A unified classification of stationary phases for packed column supercritical fluid chromatography. J Chromatogr A. 2008;1191(1-2):21-39.
  4. West C, Khater S, Lesellier E. Characterization and use of hydrophilic interaction liquid chromatography type stationary phases in supercritical fluid chromatography. J Chromatogr A. 2012;1250:182-95.
  5. Lesellier E. Retention mechanisms in super/subcritical fluid chromatography on packed columns. J Chromatogr A. 2009;1216(10):1881-90.
  6. Zou W, Dorsey JG, Chester TL. Modifier effects on column efficiency in packed-column supercritical fluid chromatography. Anal Chem. 2000;72(15):3620-6.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Impurity Profiling Using UPC2/MS
Impurity Profiling Using UPC2/MS
2013|Waters|Applications
Impurity Profiling Using UPC 2/MS Michael D. Jones and Warren Potts Waters Corporation, Milford, MA, USA A P P L I C AT I O N B E N E F I T S ■■ Solve complex impurity profile challenges…
Key words
impurity, impuritymetoclopramide, metoclopramidesqd, sqdexpired, expiredmethoxylation, methoxylationworking, workingposneg, posnegintensity, intensityacquity, acquitymethylations, methylationsminutes, minutesregarding, regardinginvestigation, investigationquestions, questionspeaks
APPLICATION NOTEBOOK - IMPURITIES - Developing Safe and Effective Drugs
[ APPLICATION NOTEBOOK ] IMPURITIES Developing Safe and Effective Drugs INT RODUCTION NAVIGATION T he pharmaceutical industry is integral in maintaining public health by providing therapeutic and preventative medicines. T he cornerstones of bringing new medicines to market are quality,…
Key words
home, homeuplc, uplcimpurities, impuritiesziprasidone, ziprasidoneusp, uspacquity, acquityimpurity, impuritymethod, methodtioconazole, tioconazoleminutes, minutessimvastatin, simvastatinhplc, hplcusing, usingrelated, relatedanalysis
Getting Started with UltraPerformance Convergence Chromatography
Getting Started with UltraPerformance Convergence Chromatography A Practitioner’s Guide for Utilizing UPC2 in the Chromatographic Laboratory ©2013 Waters Corporation 1 Agenda  What is UPC2?  Getting Started  Important Considerations for UPC2  UPC2 as a Replacement for NPLC…
Key words
resolved, resolvedconvergence, convergencesolvent, solventsolvents, solventsminutes, minutesshape, shaperetention, retentiondensity, densityabpr, abpradditive, additivenormal, normalselectivity, selectivitysolubility, solubilityphase, phasechromatography
Waters ACQUITY UPC2 COLUMNS
Waters ACQUITY UPC2 COLUMNS
2016|Waters|Brochures and specifications
[ ACQUIT Y UPC 2 COLUMNS ] ACQUITY UPC 2 Columns for chiral and achiral separations ACQUITY UPC2 COLUMNS ACQUITY UPC2 Instruments and Columns The ACQUITY UPC 2® System gives scientists the ability to precisely vary mobile phase strength, pressure,…
Key words
chiral, chiralmin, mintorus, torusconvergence, convergenceachiral, achiraldiol, diolcbd, cbdseparations, separationscompounds, compoundssepa, sepadea, deabatch, batchcolumns, columnssfc, sfcphase
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike