A Clinical Research Method for the Analysis of Plasma Mycophenolic Acid
Applications | 2015 | WatersInstrumentation
Mycophenolic acid is a key immunosuppressive agent that inhibits inosine monophosphate dehydrogenase, preventing the proliferation of B and T lymphocytes. Accurate monitoring of its plasma concentration is essential in clinical research and therapeutic drug monitoring to optimize dosing, minimize rejection risk in transplant patients, and manage side effects. A robust analytical method that distinguishes the drug from its glucuronide metabolites ensures reliable results.
The primary goal of the study was to develop a rapid, sensitive and specific UPLC-MS/MS assay for quantifying mycophenolic acid in human plasma. Key objectives included:
Sample preparation involved a protein precipitation protocol: 50 µL plasma was mixed with 500 µL of aqueous methanol containing zinc sulfate and internal standard (mycophenolic acid-2H3). After centrifugation, the supernatant was analyzed directly.
Used Instrumentation:
Chromatography employed a 2.5 min gradient (water/methanol with 2 mM ammonium acetate and 0.1% formic acid) at 0.7 mL/min. MS detection used positive-mode ESI with multiple reaction monitoring (MRM) transitions for quantifier, qualifier, and deuterated internal standard.
The calibration curve was linear from 0.1 to 20 µg/mL (r² > 0.994). The lower limit of quantification, defined by a signal-to-noise ratio >10:1 and precision <20% RSD, was 0.075 µg/mL. Intra- and inter-day precision at low (0.5 µg/mL), mid (2.4 µg/mL) and high (5.0 µg/mL) levels showed total RSD ≤5.3%. No significant carryover was observed, and dilution integrity at 40 µg/mL provided 98% accuracy. Matrix effect studies indicated minor ion suppression (matrix factors 0.91–0.94) effectively corrected by the internal standard. Comparison with an independent LC-MS/MS method (n=35) produced a Deming regression of y=0.90x+0.13 and a mean bias of −0.05%.
This method offers:
Advancements may include automation of sample preparation to further increase throughput, exploration of microflow LC-MS platforms to reduce solvent consumption, and integration with digital laboratory information systems for real-time data analysis. Emerging high-resolution mass spectrometry techniques may enable simultaneous profiling of additional immunosuppressant drugs and metabolites in multi-analyte panels.
An analytically robust UPLC-MS/MS assay for mycophenolic acid in plasma has been developed, demonstrating excellent linearity, sensitivity, precision, and metabolite separation. The method is well suited for clinical research settings and supports reliable therapeutic drug monitoring.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerWaters
Summary
Importance of the Topic
Mycophenolic acid is a key immunosuppressive agent that inhibits inosine monophosphate dehydrogenase, preventing the proliferation of B and T lymphocytes. Accurate monitoring of its plasma concentration is essential in clinical research and therapeutic drug monitoring to optimize dosing, minimize rejection risk in transplant patients, and manage side effects. A robust analytical method that distinguishes the drug from its glucuronide metabolites ensures reliable results.
Study Objectives and Overview
The primary goal of the study was to develop a rapid, sensitive and specific UPLC-MS/MS assay for quantifying mycophenolic acid in human plasma. Key objectives included:
- Achieving baseline chromatographic separation between mycophenolic acid and its major metabolites (MPAG and AcMPAG).
- Establishing a simple sample preparation workflow suitable for clinical research laboratories.
- Validating method linearity, precision, accuracy, sensitivity and robustness for routine application.
Methodology and Instrumentation
Sample preparation involved a protein precipitation protocol: 50 µL plasma was mixed with 500 µL of aqueous methanol containing zinc sulfate and internal standard (mycophenolic acid-2H3). After centrifugation, the supernatant was analyzed directly.
Used Instrumentation:
- ACQUITY UPLC I-Class system
- ACQUITY UPLC HSS C18 SB column (2.1 × 30 mm, 1.8 µm)
- Xevo TQD triple quadrupole mass spectrometer
- MassLynx Software with TargetLynx Application Manager
Chromatography employed a 2.5 min gradient (water/methanol with 2 mM ammonium acetate and 0.1% formic acid) at 0.7 mL/min. MS detection used positive-mode ESI with multiple reaction monitoring (MRM) transitions for quantifier, qualifier, and deuterated internal standard.
Main Results and Discussion
The calibration curve was linear from 0.1 to 20 µg/mL (r² > 0.994). The lower limit of quantification, defined by a signal-to-noise ratio >10:1 and precision <20% RSD, was 0.075 µg/mL. Intra- and inter-day precision at low (0.5 µg/mL), mid (2.4 µg/mL) and high (5.0 µg/mL) levels showed total RSD ≤5.3%. No significant carryover was observed, and dilution integrity at 40 µg/mL provided 98% accuracy. Matrix effect studies indicated minor ion suppression (matrix factors 0.91–0.94) effectively corrected by the internal standard. Comparison with an independent LC-MS/MS method (n=35) produced a Deming regression of y=0.90x+0.13 and a mean bias of −0.05%.
Benefits and Practical Applications
This method offers:
- Quick and straightforward sample preparation compatible with high throughput.
- Rapid analysis with a 3-minute run time, improving laboratory efficiency.
- Reliable separation of mycophenolic acid from interfering metabolites, ensuring accurate quantification.
- Sensitivity and precision meeting clinical research requirements for therapeutic monitoring.
Future Trends and Opportunities
Advancements may include automation of sample preparation to further increase throughput, exploration of microflow LC-MS platforms to reduce solvent consumption, and integration with digital laboratory information systems for real-time data analysis. Emerging high-resolution mass spectrometry techniques may enable simultaneous profiling of additional immunosuppressant drugs and metabolites in multi-analyte panels.
Conclusion
An analytically robust UPLC-MS/MS assay for mycophenolic acid in plasma has been developed, demonstrating excellent linearity, sensitivity, precision, and metabolite separation. The method is well suited for clinical research settings and supports reliable therapeutic drug monitoring.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Use of Online SPE for Clinical Research in the Automated Analysis of Mycophenolic Acid by LC-MS
2014|Waters|Applications
Use of Online SPE for Clinical Research in the Automated Analysis of Mycophenolic Acid by LC-MS Martin Eastwood and Lisa Calton, Waters Corporation, Milford, MA, USA Gary Chusney, Hammersmith Hospital, London, UK G OA L T HE SOLUT ION To…
Key words
mpa, mpamycophenolic, mycophenoliconline, onlinespe, spedeming, demingacid, acidautomated, automateduplc, uplcacquity, acquityimmunosuppresant, immunosuppresantnucelotide, nucelotideresearch, researchwaters, watersimmunosuppression, immunosuppressioncalcineurin
Clinical Research Method for the Fast, Efficient Analysis of Mycophenolic Acid Utilizing Online Solid-Phase Extraction in Conjunction with UPLC/MS/MS
2013|Waters|Applications
Clinical Research Method for the Fast, Efficient Analysis of Mycophenolic Acid Utilizing Online Solid-Phase Extraction in Conjunction with UPLC/MS/MS Martin Eastwood,1 Gary Chusney,2 Lisa Calton1 1 Clinical Operations Group, Waters Corporation, Manchester, UK 2 Imperial College Renal and Transplant Centre,…
Key words
mycophenolic, mycophenolicacid, acidonline, onlineanalytical, analyticalsuppression, suppressionsample, samplepreparation, preparationmasstrak, masstrakextraction, extractionminimal, minimalimprecision, imprecisionchromsystems, chromsystemsconditions, conditionsglucuronides, glucuronidesfavorably
Simultaneous Analysis of Vitamins A and E in Serum by UPLC-MS/MS for Clinical Research
2019|Waters|Applications
[ APPLICATION NOTE ] Simultaneous Analysis of Vitamins A and E in Serum by UPLC-MS/MS for Clinical Research Michelle Wills, Gareth Hammond, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS INTRODUCTION ■ Sample requirement of only 100 µL Vitamins…
Key words
vitamin, vitamindeming, deminguplc, uplcvitamins, vitaminstocopherol, tocopherolaltm, altmneqas, neqasserum, serumretinol, retinolclinical, clinicalfit, fittqd, tqdsimultaneous, simultaneousresearch, researchprime
A Reproducible Method for the Quantification of Pioglitazone and Metabolites in Human Plasma Using the ACQUITY UPLC H-Class System and Xevo TQD MS with UNIFI
2012|Waters|Applications
A Reproducible Method for the Quantification of Pioglitazone and Metabolites in Human Plasma Using the ACQUITY UPLC H-Class System and Xevo TQD MS with UNIFI Jennifer Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L…
Key words
pioglitazone, pioglitazonemetabolites, metabolitestqd, tqdreproducible, reproduciblequantification, quantificationdev, devxevo, xevoclass, classunifi, unifilloq, lloqmid, midbias, biasuplc, uplcacquity, acquitymethod