Tips & Tricks for improved quantitative and qualitative LC-MS
Posters | 2020 | MerckInstrumentation
Liquid chromatography–mass spectrometry (LC-MS) is a cornerstone technique in analytical chemistry, enabling trace-level detection and characterization of small molecules, peptides and complex mixtures. Achieving optimal sensitivity and low limits of detection hinges on stringent control of contaminant sources, background noise and adduct formation. The guidelines outlined here address critical factors—from solvent purity to column bleed and labware handling—that directly impact data quality and reproducibility.
This application note compiles practical measures and comparative data to maximize LC-MS performance. Key aims include:
The study employed an LC-ion trap MS system (m/z 50–2000, positive ESI mode) coupled to various reversed-phase and HILIC columns. Solvents (methanol, acetonitrile) and additives (formic/acetic acid, ammonium salts) were sourced as MS-grade or hypergrade. Water was supplied by a Milli-Q® ultrapure system. Labware (glass bottles, plastic vessels) underwent different cleaning and storage protocols. Gradient runs (5–95% organic) and flow injection analyses were performed using reserpine or Glu-Fibrinopeptide B as internal standards.
Implementing these measures enhances assay sensitivity, lowers noise and improves reproducibility. Laboratories engaged in pharmaceutical analysis, proteomics or environmental testing will benefit from clearer spectra, fewer false positives and more reliable quantification at trace concentrations.
Advances in stationary-phase chemistries promise further bleed reduction, while novel volatile additives may optimize ionization. Integration of inline desalting and automated cleaning modules will streamline workflows. Real-time contamination monitoring through software-driven TIC baseline checks could become standard practice.
Careful selection and handling of solvents, additives, columns, water and labware are critical to unlocking the full potential of LC-MS. Adopting MS-grade materials, low-bleed columns and rigorous cleaning protocols ensures high sensitivity, low background and consistent results across laboratories.
LC/MS
IndustriesManufacturerMerck
Summary
Significance of the Topic
Liquid chromatography–mass spectrometry (LC-MS) is a cornerstone technique in analytical chemistry, enabling trace-level detection and characterization of small molecules, peptides and complex mixtures. Achieving optimal sensitivity and low limits of detection hinges on stringent control of contaminant sources, background noise and adduct formation. The guidelines outlined here address critical factors—from solvent purity to column bleed and labware handling—that directly impact data quality and reproducibility.
Objectives and Study Overview
This application note compiles practical measures and comparative data to maximize LC-MS performance. Key aims include:
- Demonstrating the impact of solvent and additive purity on baseline noise and signal suppression.
- Comparing stationary phases with regard to column bleed and background signals.
- Assessing the role of water quality and laboratory equipment in preventing contamination.
Methodology and Instrumentation
The study employed an LC-ion trap MS system (m/z 50–2000, positive ESI mode) coupled to various reversed-phase and HILIC columns. Solvents (methanol, acetonitrile) and additives (formic/acetic acid, ammonium salts) were sourced as MS-grade or hypergrade. Water was supplied by a Milli-Q® ultrapure system. Labware (glass bottles, plastic vessels) underwent different cleaning and storage protocols. Gradient runs (5–95% organic) and flow injection analyses were performed using reserpine or Glu-Fibrinopeptide B as internal standards.
Main Results and Discussion
- Solvent Purity: Hypergrade solvents yielded lower total ion current (TIC) baselines and reduced adduct peaks compared to standard gradient grades.
- Column Bleed: Low-bleed stationary phases (e.g., ZIC®-HILIC) and pre-run flushing cycles minimized background signals across multiple gradient cycles.
- Water Quality: Frequent flushing of the ultrapure water system prevented accumulation of organic contaminants; water stored in treated amber glass remained cleaner than in plastic or dish-washed vessels.
- Labware Effects: Detergent residues and plasticizers leached from glass cleaned in dishwashers or from plastic bottles produced ghost peaks; direct use of surface-treated glassware and solvent rinsing was essential.
- Sodium Content: Increasing Na+ concentrations in sample solutions led to predictable adduct patterns ([M+Na]+, [M+2Na−H]+), emphasizing the need for desalting steps with high-salt samples.
Benefits and Practical Applications
Implementing these measures enhances assay sensitivity, lowers noise and improves reproducibility. Laboratories engaged in pharmaceutical analysis, proteomics or environmental testing will benefit from clearer spectra, fewer false positives and more reliable quantification at trace concentrations.
Future Trends and Applications
Advances in stationary-phase chemistries promise further bleed reduction, while novel volatile additives may optimize ionization. Integration of inline desalting and automated cleaning modules will streamline workflows. Real-time contamination monitoring through software-driven TIC baseline checks could become standard practice.
Conclusion
Careful selection and handling of solvents, additives, columns, water and labware are critical to unlocking the full potential of LC-MS. Adopting MS-grade materials, low-bleed columns and rigorous cleaning protocols ensures high sensitivity, low background and consistent results across laboratories.
References
- Merck KGaA, Darmstadt. Tips & Tricks for improved quantitative and qualitative LC-MS. Lit. No. MK_PS5342EN, 2020.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
LC-MS Grade Solvents and Reagents
2020|Merck|Brochures and specifications
LC-MS Grade Solvents and Reagents for accurate, brilliant results The life science business of Merck operates as MilliporeSigma in the U.S. and Canada. LC-MS: Raising the bar on purity Liquid chromatography - mass spectrometry (LC-MS) is fast becoming a routine…
Key words
lichrosolv, lichrosolvhypergrade, hypergradelichropur, lichropurreserpine, reserpinehydroxy, hydroxysolvents, solventsreagents, reagentsacid, acidamine, aminecarbonyl, carbonyluhplc, uhplcintensity, intensityacetonitrile, acetonitrilepurity, puritydiene
LC-MS Resource Guide
2020|Merck|Guides
LC-MS Resource Guide A Comprehensive Portfolio for Consistent Results in Routine and Advanced LC-MS applications The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada. LC-MS Resource Guide Maximize the Performance of Your…
Key words
chromolith, chromolithzic, ziccustom, customhybridspe, hybridspephase, phasecolumn, columnresource, resourceguard, guardhplc, hplcmobile, mobileproteins, proteinscolumns, columnshilic, hiliclichrosolv, lichrosolvguide
Agilent InfinityLab solvents for HPLC and LC/MS
2024|Agilent Technologies|Brochures and specifications
Maximize Your Efficiency with Precision Solvents Agilent InfinityLab solvents for HPLC and LC/MS Image needed Agilent InfinityLab gradient grade HPLC solvents Precision Solvents for Superior HPLC & UHPLC Choosing the right solvent is pivotal in achieving optimal sample separation, as…
Key words
hazard, hazardinfinitylab, infinitylabstatements, statementsbeilstein, beilsteinagilent, agilentspecification, specificationwgk, wgkreserpine, reserpinewater, waternumber, numberidentity, identityflammables, flammablesgradient, gradientghs, ghsmode
Mastering HILIC-Z Separation for Polar Analytes
2023|Agilent Technologies|Applications
Application Note Omics Mastering HILIC-Z Separation for Polar Analytes Authors Karen E. Yannell, PhD, Jordy Hsiao, PhD, and Dan Cuthbertson, PhD Agilent Technologies, Inc. Abstract This application note presents recommended practices when using an Agilent InfinityLab Poroshell 120 HILIC-Z column…
Key words
hilic, hilicinfinitylab, infinitylabretention, retentionsigma, sigmacolumn, columnagilent, agilentanalytes, analytessolvent, solventpolar, polaracid, acidtime, timezwitterionic, zwitterionicdedicated, dedicatedrecommended, recommendedwater