A Triple Quadrupole LC/MS/MS Method for Quantitative Analysis of Methylenedioxypyrovalerone (MDPV) and Mephedrone, Common Components of “Bath Salts” in Urine
Applications | 2012 | Agilent TechnologiesInstrumentation
In recent years, synthetic cathinones such as MDPV and mephedrone have emerged as widely abused stimulants, often sold under the name 'bath salts'. Their structural similarity to amphetamines and MDMA, combined with the lack of reliable immunoassay screens and conventional GC/MS confirmation methods, poses significant analytical challenges in forensic and clinical toxicology. Accurate detection and quantification of these compounds in biological matrices like urine are critical for public health, law enforcement and clinical diagnosis.
This application note presents a robust liquid chromatography–tandem mass spectrometry (LC/MS/MS) method for the sensitive and selective quantitation of MDPV and mephedrone in urine. Developed by Sterling Reference Laboratories and Agilent Technologies, the method aims to achieve:
Urine samples spiked with analytes and deuterated internal standards undergo cation-exchange solid-phase extraction (SPE) to remove matrix components and reduce ion suppression. After cartridge conditioning and sequential washes, analytes are eluted with an organic solvent, evaporated, and reconstituted in aqueous mobile phase. Separation is performed on a C18 column using a rapid gradient (4.2-minute run time). Detection employs an Agilent 6460 Triple Quadrupole mass spectrometer with electrospray ionization in positive mode, monitoring two MRM transitions per compound for confirmation and quantitation.
The method demonstrated linear calibration curves (1–5000 ng/mL) with R² > 0.999 for both analytes. Average precision across nine calibration levels was better than 5 % RSD, and accuracy at the LOQ (5 ng/mL) exceeded 80 %. The LOD was established at 1 ng/mL based on signal-to-noise ratios above 19. Endogenous and structurally related compounds (e.g., ephedrine, amphetamine) spiked at high concentrations showed no interference. Carryover after 10,000 ng/mL injections was below 1.5 ng/mL, well under the 25 ng/mL cutoff.
This LC/MS/MS protocol offers forensic and clinical laboratories a high-throughput, sensitive solution for bath salt analysis. The SPE cleanup ensures robust instrument performance and minimizes maintenance. Rapid chromatographic separation and dual MRM monitoring provide confident identification and quantification, supporting large-scale screening and confirmation workflows.
As new synthetic cathinones continue to appear, the described method can be readily expanded by adding MRM transitions for emerging analogs. Advances in high-resolution mass spectrometry and automation of SPE workflows may further enhance throughput and broaden screening panels. Integration with data-processing software will support real-time detection and reporting in clinical and forensic settings.
The validated LC/MS/MS approach reliably quantifies MDPV and mephedrone in urine with exceptional sensitivity, precision, and specificity. Its adaptability and efficiency make it an essential tool for laboratories addressing the evolving landscape of designer stimulant abuse.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesForensics
ManufacturerAgilent Technologies
Summary
Significance of the Topic
In recent years, synthetic cathinones such as MDPV and mephedrone have emerged as widely abused stimulants, often sold under the name 'bath salts'. Their structural similarity to amphetamines and MDMA, combined with the lack of reliable immunoassay screens and conventional GC/MS confirmation methods, poses significant analytical challenges in forensic and clinical toxicology. Accurate detection and quantification of these compounds in biological matrices like urine are critical for public health, law enforcement and clinical diagnosis.
Objectives and Overview
This application note presents a robust liquid chromatography–tandem mass spectrometry (LC/MS/MS) method for the sensitive and selective quantitation of MDPV and mephedrone in urine. Developed by Sterling Reference Laboratories and Agilent Technologies, the method aims to achieve:
- Excellent linearity across a broad calibration range
- Low limits of detection (LOD) and quantitation (LOQ)
- High precision and accuracy
- Minimal interferences and negligible carryover
Methodology and Instrumentation
Urine samples spiked with analytes and deuterated internal standards undergo cation-exchange solid-phase extraction (SPE) to remove matrix components and reduce ion suppression. After cartridge conditioning and sequential washes, analytes are eluted with an organic solvent, evaporated, and reconstituted in aqueous mobile phase. Separation is performed on a C18 column using a rapid gradient (4.2-minute run time). Detection employs an Agilent 6460 Triple Quadrupole mass spectrometer with electrospray ionization in positive mode, monitoring two MRM transitions per compound for confirmation and quantitation.
Used Instrumentation
- Agilent 1200 Series HPLC (binary pump, autosampler, column compartment)
- Agilent 6460 Triple Quadrupole LC/MS/MS system with ESI source
- Biochemical Diagnostics SPE workstation with cation-exchange cartridges
- Polaris C18 column, 50×2.0 mm, 5 µm
Key Results and Discussion
The method demonstrated linear calibration curves (1–5000 ng/mL) with R² > 0.999 for both analytes. Average precision across nine calibration levels was better than 5 % RSD, and accuracy at the LOQ (5 ng/mL) exceeded 80 %. The LOD was established at 1 ng/mL based on signal-to-noise ratios above 19. Endogenous and structurally related compounds (e.g., ephedrine, amphetamine) spiked at high concentrations showed no interference. Carryover after 10,000 ng/mL injections was below 1.5 ng/mL, well under the 25 ng/mL cutoff.
Benefits and Practical Applications
This LC/MS/MS protocol offers forensic and clinical laboratories a high-throughput, sensitive solution for bath salt analysis. The SPE cleanup ensures robust instrument performance and minimizes maintenance. Rapid chromatographic separation and dual MRM monitoring provide confident identification and quantification, supporting large-scale screening and confirmation workflows.
Future Trends and Applications
As new synthetic cathinones continue to appear, the described method can be readily expanded by adding MRM transitions for emerging analogs. Advances in high-resolution mass spectrometry and automation of SPE workflows may further enhance throughput and broaden screening panels. Integration with data-processing software will support real-time detection and reporting in clinical and forensic settings.
Conclusion
The validated LC/MS/MS approach reliably quantifies MDPV and mephedrone in urine with exceptional sensitivity, precision, and specificity. Its adaptability and efficiency make it an essential tool for laboratories addressing the evolving landscape of designer stimulant abuse.
References
- American Association of Poison Control Centers. http://www.aapcc.org
- DEA Drug Fact Sheet, Bath Salts or Designer Cathinones. http://www.justice.gov/dea/pubs/abuse/drug_data_sheets/Bath_Salts.pdf
- Federal Register, Vol. 76, No. 174, September 8, 2011
- Spiller HA, et al. Clinical experience with and analytical confirmation of “bath salts” and “legal highs” (synthetic cathinones) in the United States. Clin Toxicol. 2011;49(6):499–505.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Quantitative determination of seven synthetic cathinones (stimulants) from stabilized human urine by UHPLC-MS/MS for forensic toxicology
2019|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 21965 Quantitative determination of seven synthetic cathinones (stimulants) from stabilized human urine by UHPLC-MS/MS for forensic toxicology Authors Kean Woodmansey and Matthew Franklin Thermo Fisher Scientific, Runcorn, UK Keywords Syncronis, cathinones, “bath salts”, khat, plant food, legal highs,…
Key words
cathinones, cathinonesmethedrone, methedronebuphedrone, buphedronebutylone, butyloneethylone, ethylonemdpv, mdpvmethylone, methylonemephedrone, mephedronepositive, positivegrade, gradeassay, assaymatrix, matrixuhplc, uhplccounts, countsassessed
Analysis of “Bath Salt” Compounds from Urine for Forensic Toxicology Using μElution Mixed-mode SPE Combined with UPLC/MS/MS Detection
2013|Waters|Applications
Analysis of “Bath Salt” Compounds from Urine for Forensic Toxicology Using µElution Mixed-mode SPE Combined with UPLC/MS/MS Detection Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA, USA A P P L I C AT…
Key words
µelution, µelutionbath, bathtoxicology, toxicologyuplc, uplcforensic, forensicurine, urinespe, spesalt, saltcompounds, compoundscathinones, cathinonesmixed, mixedalphapyrrolidinopentiophenone, alphapyrrolidinopentiophenonecombined, combinedmdppp, mdpppmcx
Ultrafast Screen for Bath Salts in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System
2014|Agilent Technologies|Applications
Ultrafast Screen for Bath Salts in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System Application Note Forensic Toxicology Authors Abstract Nikunj R. Parikh, Michelle Romm, and Vaughn P. Miller Agilent Technologies, Inc. Wakefield, MA USA The need for greater…
Key words
interday, interdayintraday, intradayprod, prodauc, aucrapidfire, rapidfiremethoxymethcathinone, methoxymethcathinonemethylone, methyloneurine, urinemethcathinone, methcathinonesalts, saltsmdpv, mdpvmephedrone, mephedronebath, bathfluoromethcathinone, fluoromethcathinoneratio
Automated Hydrolysis, Extraction and Analysis of Synthetic Cathinones in Urine using a Robotic Autosampler and LC-MS/MS Platform
2020|Agilent Technologies|Applications
GERSTEL Application Note No. 215, 2020 Automated Hydrolysis, Extraction and Analysis of Synthetic Cathinones in Urine using a Robotic Autosampler and LC-MS/MS Platform Fred D. Foster1 and John R. Stuff1 1. GERSTEL, Inc., 701 Digital Drive, Suite J, Linthicum, MD…
Key words
hydrolysis, hydrolysisgerstel, gersteldpx, dpxurine, urinecathinones, cathinonessynthetic, syntheticenzymatic, enzymaticmdpv, mdpvautomated, automatedret, retusing, usinghydrolyzed, hydrolyzedmps, mpsmorphine, morphinerobotic