Development of a Comprehensive Detection Method of Eicosanoids and Platelet Activating Factor Using Ultra-High Performance Liquid Chromatography/Mass Spectrometry
Applications | 2013 | ShimadzuInstrumentation
Comprehensive profiling of eicosanoids and platelet-activating factor (PAF) is crucial for understanding inflammatory processes, cardiovascular regulation, and immune responses. Ultra-high-performance liquid chromatography coupled with triple quadrupole mass spectrometry (UHPLC–MS/MS) enables sensitive and selective detection of multiple lipid mediators in a single run, supporting biomarker discovery and therapeutic monitoring.
This study aimed to develop and validate a simultaneous detection method for 50 eicosanoids and PAF using a Shimadzu UHPLC–MS/MS system (LCMS-8040). The method employed 54 optimized multiple reaction monitoring (MRM) transitions and a single ion monitoring (SIM) channel to achieve sub-picogram quantitation limits while maintaining high throughput.
Sample Preparation and Standards:
Chromatographic Conditions:
Mass Spectrometry Parameters:
"Nexera" UHPLC system with a maximum pressure of 130 MPa and Shimadzu LCMS-8040 triple quadrupole mass spectrometer.
Columns evaluated:
Retention and Separation:
Sensitivity and Linearity:
Polarity Switching and Throughput:
The developed method enables high-throughput, multiplexed analysis of lipid mediators in biological and clinical samples. Sub-picogram sensitivity supports trace-level quantitation in complex matrices. This approach facilitates:
Expansion of the assay to include additional fatty acid metabolites, specialized pro-resolving mediators, and sphingolipids could broaden its utility. Integration with high-resolution mass spectrometry and ion mobility separation may further enhance isomer discrimination. Automation of sample preparation and data processing will streamline large-scale lipidomics studies.
A robust UHPLC–MS/MS method was established for simultaneous detection of 50 eicosanoids and PAF with sub-picogram quantitation limits. Optimized chromatography and polarity switching maximize sensitivity and throughput, making the method suitable for diverse research and applied laboratories.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerShimadzu
Summary
Significance of the Topic
Comprehensive profiling of eicosanoids and platelet-activating factor (PAF) is crucial for understanding inflammatory processes, cardiovascular regulation, and immune responses. Ultra-high-performance liquid chromatography coupled with triple quadrupole mass spectrometry (UHPLC–MS/MS) enables sensitive and selective detection of multiple lipid mediators in a single run, supporting biomarker discovery and therapeutic monitoring.
Objectives and Study Overview
This study aimed to develop and validate a simultaneous detection method for 50 eicosanoids and PAF using a Shimadzu UHPLC–MS/MS system (LCMS-8040). The method employed 54 optimized multiple reaction monitoring (MRM) transitions and a single ion monitoring (SIM) channel to achieve sub-picogram quantitation limits while maintaining high throughput.
Methodology and Instrumentation
Sample Preparation and Standards:
- Fifty eicosanoids, PAF, and an internal standard (PAF-d4) dissolved in methanol.
Chromatographic Conditions:
- Mobile phases: 0.1% formic acid in water (A) and acetonitrile (B).
- Gradient elution optimized for three columns: Shim-pack XR-ODS III, CAPCELL PAK C18 IF-II, and Phenomenex Kinetex C8.
- Flow rates and gradient programs tailored to achieve baseline separation of isomeric species.
Mass Spectrometry Parameters:
- LCMS-8040 with modified ion optics, polarity switching (15 ms), and loop time of 0.57 s.
- Positive and negative mode MRM for each analyte; 54 transitions scheduled per cycle.
- Single ion monitoring at m/z 319 for general detection of hydroxyeicosatetraenoic acids.
Used Instrumentation
"Nexera" UHPLC system with a maximum pressure of 130 MPa and Shimadzu LCMS-8040 triple quadrupole mass spectrometer.
Columns evaluated:
- Shim-pack XR-ODS III (2.2 µm, 2.0 × 150 mm)
- CAPCELL PAK C18 IF-II (2 µm, 2.1 × 100 mm)
- Kinetex C8 (2.6 µm, 2.1 × 150 mm)
Main Results and Discussion
Retention and Separation:
- All 50 analytes were detected, including isomer pairs like PGF2α/ent-PGF2α and 8-iso-PGE2/ent-8-iso-PGE2.
- CAPCELL PAK C18 improved peak shapes for late-eluting mediators such as LTC4 and PAF-d4.
- Kinetex C8 provided optimal resolution for a subset of ten target compounds used in quantitation studies.
Sensitivity and Linearity:
- Limits of quantitation ranged from 0.5 pg to 5 pg depending on the analyte.
- Calibration curves exhibited excellent linearity (R2 > 0.999) over 0.5–1000 pg.
- Positive mode enhanced detection of lactone species, while negative mode was preferred for most fatty acid mediators.
Polarity Switching and Throughput:
- Efficient polarity switching allowed concurrent acquisition of positive and negative MRMs without sensitivity loss.
- Cycle time of 0.57 s ensured at least five data points across narrow UHPLC peaks.
Benefits and Practical Applications
The developed method enables high-throughput, multiplexed analysis of lipid mediators in biological and clinical samples. Sub-picogram sensitivity supports trace-level quantitation in complex matrices. This approach facilitates:
- Biomarker discovery in inflammatory and metabolic diseases.
- Pharmacokinetic and pharmacodynamic studies of anti-inflammatory drugs.
- Quality control and process monitoring in pharmaceutical and biotechnology laboratories.
Future Trends and Opportunities
Expansion of the assay to include additional fatty acid metabolites, specialized pro-resolving mediators, and sphingolipids could broaden its utility. Integration with high-resolution mass spectrometry and ion mobility separation may further enhance isomer discrimination. Automation of sample preparation and data processing will streamline large-scale lipidomics studies.
Conclusion
A robust UHPLC–MS/MS method was established for simultaneous detection of 50 eicosanoids and PAF with sub-picogram quantitation limits. Optimized chromatography and polarity switching maximize sensitivity and throughput, making the method suitable for diverse research and applied laboratories.
References
- Y. Kita, T. Takahashi, N. Uozumi, T. Shimizu, Anal. Biochem. 342 (2005) 134–143.
- M. Masoodi, M. Eiden, A. Koulman, D. Spinaer, D. A. Volmer, Anal. Chem. 82 (2010) 8176–8185.
- S. A. Brose, B. T. Thuen, M. Y. Golovko, J. Lipid Res. 52 (2011) 850–859.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Comprehensive Monitoring Method for Analyzing 158 Lipid Mediator Species Using the Ultra-Fast LCMS-8050
2014|Shimadzu|Applications
LAAN-A-LM-E073 Application News C102 No. Liquid Chromatography Mass Spectrometry Comprehensive Monitoring Method for Analyzing 158 Lipid Mediator Species Using the Ultra-Fast LCMS-8050 Lipid mediator is a generic term for bioactive lipids, which are produced in the body and play a…
Key words
dha, dhaepa, epadgla, dglatetranor, tetranorpaf, pafbrain, braincategory, categoryflowrate, flowrateazelaoyl, azelaoylethanolamide, ethanolamidemediator, mediatoraea, aeametabolites, metaboliteseda, edamediators
Shimadzu Selection Guide Metabolite Analysis - Metabolomics Product Portfolio
2019|Shimadzu|Brochures and specifications
C146-E280D Selection Guide Metabolite Analysis Metabolomics Product Portfolio Expanding Metabolomics Metabolomics refers to an array of techniques used to comprehensively detect and analyze various metabolites formed in vivo during biological activity. The qualitative and quantitative changes in metabolites reflect the…
Key words
acid, acidsba, sbametabolites, metabolitesdatabase, databasepackage, packageacids, acidsmetabolomics, metabolomicsanalysis, analysisphospholipid, phospholipidamino, aminomediators, mediatorsmonophosphate, monophosphatecooh, coohmrm, mrmlcms
Selection Guide Metabolite Analysis - Metabolomics Product Portfolio
2019|Shimadzu|Brochures and specifications
C146-E280D Selection Guide Metabolite Analysis Metabolomics Product Portfolio Expanding Metabolomics Metabolomics refers to an array of techniques used to comprehensively detect and analyze various metabolites formed in vivo during biological activity. The qualitative and quantitative changes in metabolites reflect the…
Key words
acid, acidsba, sbametabolites, metabolitesdatabase, databasepackage, packageacids, acidsmetabolomics, metabolomicsphospholipid, phospholipidanalysis, analysisamino, aminomediators, mediatorsmonophosphate, monophosphatecooh, coohmrm, mrmlcms
Development of a Fast Eicosanoid Profiling Method Using Plasma Microsamples
2021|Shimadzu|Applications
LCMS™-8060 High Performance Liquid Chromatograph Mass Spectrometer Application News Development of a Fast Eicosanoid Profiling Method Using Plasma Microsamples Masaki Yamada, Naoko Nagano, Harumi Kubo, Takanari Hattori User Benefits Eicosanoids and related fatty acids in plasma can be quantified…
Key words
dha, dhamicrosampling, microsamplingdgla, dglaplasma, plasmaprofiling, profilingsnapping, snappingwingtm, wingtmtargets, targetsinternal, internaleicosanoid, eicosanoidpaf, pafepa, epapathophysiology, pathophysiologyfast, fastmicrosamples