Automated Derivatization of Neurotransmitters in Plasma Extracts for Liquid Chromatograph Tandem Mass Spectrometry
Applications | 2020 | ShimadzuInstrumentation
Accurate and efficient quantification of neurotransmitters in plasma provides vital insights into neurological function, disease biomarkers and drug effects. Traditional analysis of highly polar amines and related metabolites is challenged by poor chromatographic retention and complex sample matrices, highlighting the need for robust automated workflows.
This application note describes the development of an automated precolumn derivatization strategy for simultaneous measurement of multiple neurotransmitters in plasma extracts using benzoyl chloride. The main goals were to improve chromatographic separation, enhance sensitivity in tandem mass spectrometry, and streamline sample preparation via autosampler automation.
The protocol combines protein precipitation, benzoylation derivatization under alkaline conditions, and LC-MS/MS analysis:
Chromatographic separation achieved baseline resolution for all benzoylated analytes within a 10 min run. Overlaid MRM chromatograms of standards (10 ng/mL) and plasma samples confirmed specificity and peak integrity. Validation metrics included:
The automated benzoylation workflow reduces manual handling, minimizes variability, and increases sample throughput. Hydrophobic derivatives improve retention on reverse-phase columns, yielding sharper peaks and enhanced MS sensitivity. This method is applicable to pharmacokinetic studies, clinical biomarker research and neurochemical monitoring in drug development.
Potential extensions include multiplexed assays covering additional biogenic amines, adaptation to other biofluids (CSF, urine), and integration with microfluidic sample preparation. Emerging derivatization reagents and green chemistry approaches may further simplify workflows and reduce solvent consumption. High-throughput platforms could enable large-scale epidemiological studies and real-time monitoring in clinical settings.
An automated precolumn benzoylation method on Shimadzu LCMS-8050 provides sensitive, precise and high-throughput quantification of plasma neurotransmitters. The combination of autosampler-driven derivatization and optimized UHPLC-MS/MS conditions addresses chromatographic challenges of polar analytes and supports reliable neurochemical analysis.
1. Rodan L.H., Gibson K.M., Pearl P.L. Clinical Use of CSF Neurotransmitters. Pediatric Neurology, 53, 2015, 277–286.
2. Ghosh S., Das J. Benzoylation of Amines sans Alkali: A Green Protocol in Neat Phase. Organic Chemistry International, 2010, Article ID 743186.
Sample Preparation, LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerShimadzu
Summary
Significance of the Topic
Accurate and efficient quantification of neurotransmitters in plasma provides vital insights into neurological function, disease biomarkers and drug effects. Traditional analysis of highly polar amines and related metabolites is challenged by poor chromatographic retention and complex sample matrices, highlighting the need for robust automated workflows.
Objectives and Study Overview
This application note describes the development of an automated precolumn derivatization strategy for simultaneous measurement of multiple neurotransmitters in plasma extracts using benzoyl chloride. The main goals were to improve chromatographic separation, enhance sensitivity in tandem mass spectrometry, and streamline sample preparation via autosampler automation.
Methodology and Instrumentation
The protocol combines protein precipitation, benzoylation derivatization under alkaline conditions, and LC-MS/MS analysis:
- Sample Preparation: Rat plasma proteins were precipitated with cold acetonitrile, followed by centrifugation. The supernatant underwent automated benzoylation using a SIL-30AC autosampler.
- Derivatization: Benzoyl chloride reacts with primary/secondary amines (Schotten–Baumann reaction) in the autosampler vial under basic conditions to convert polar analytes into hydrophobic derivatives.
- Chromatography: Shimadzu Nexera X2 UHPLC equipped with a Shim-pack GIST C18 column (50×2.1 mm, 2 μm) operated at 30 °C. Mobile phase A: 5 mM ammonium formate, 0.1 % formic acid in water; B: acetonitrile; gradient from 20 % to 70 % B over 7.5 min at 0.3 mL/min.
- Mass Spectrometry: Shimadzu LCMS-8050 with positive-ion ESI. Optimized MRM transitions monitored benzoyl derivatives of 22 neurotransmitters and isotopically labeled internal standards. Ion source settings included a desolvation line at 250 °C, heat block at 400 °C, and gas flows set for nebulizing (3 L/min) and drying (10 L/min).
Key Results and Discussion
Chromatographic separation achieved baseline resolution for all benzoylated analytes within a 10 min run. Overlaid MRM chromatograms of standards (10 ng/mL) and plasma samples confirmed specificity and peak integrity. Validation metrics included:
- Linearity: 1–100 ng/mL, correlation coefficients (r²) >0.99 for all targets.
- Accuracy and Precision: Eight replicates at 10 ng/mL yielded recovery between 96 % and 138 %, with RSDs under 5 % for most analytes.
- Limits of Quantitation: LOQs ranged from 0.03 to 5.16 ng/mL, supporting trace-level detection.
- Internal Standard Repeatability: Four injections showed RSDs below 20 % for all labeled compounds, ensuring reliable quantification.
Benefits and Practical Applications
The automated benzoylation workflow reduces manual handling, minimizes variability, and increases sample throughput. Hydrophobic derivatives improve retention on reverse-phase columns, yielding sharper peaks and enhanced MS sensitivity. This method is applicable to pharmacokinetic studies, clinical biomarker research and neurochemical monitoring in drug development.
Future Trends and Applications
Potential extensions include multiplexed assays covering additional biogenic amines, adaptation to other biofluids (CSF, urine), and integration with microfluidic sample preparation. Emerging derivatization reagents and green chemistry approaches may further simplify workflows and reduce solvent consumption. High-throughput platforms could enable large-scale epidemiological studies and real-time monitoring in clinical settings.
Conclusion
An automated precolumn benzoylation method on Shimadzu LCMS-8050 provides sensitive, precise and high-throughput quantification of plasma neurotransmitters. The combination of autosampler-driven derivatization and optimized UHPLC-MS/MS conditions addresses chromatographic challenges of polar analytes and supports reliable neurochemical analysis.
Reference
1. Rodan L.H., Gibson K.M., Pearl P.L. Clinical Use of CSF Neurotransmitters. Pediatric Neurology, 53, 2015, 277–286.
2. Ghosh S., Das J. Benzoylation of Amines sans Alkali: A Green Protocol in Neat Phase. Organic Chemistry International, 2010, Article ID 743186.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research
2016|Shimadzu|Posters
PO-CON1655E High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research ASMS 2016 MP-079 Atsuhiko TOYAMA1, Kumie SATOU2, Yuki NAKAMURA2, Ichiro HIRANO1 (1) Shimadzu Corporation, Kyoto, Japan. (2) LSI Medience Corporation, Tokyo, Japan. High-sensitivity, high-throughput quantitation of…
Key words
vma, vmahva, hvaclinical, clinicalmetabolites, metabolitesquantitation, quantitationurine, urineresearch, researchurinary, urinaryneurotransmitter, neurotransmitterhplc, hplccalibration, calibrationvanillylmandelic, vanillylmandeliccatecholamine, catecholaminerange, rangeneurotransmitters
A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine
2017|Shimadzu|Posters
PO-CON1737E A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine ASMS 2017 WP 353 Shao Hua Chia1; Zhi Wei Edwin Ting1; Daisuke Kawakami2; Jie Xing1; Zhaoqi Zhan1 1 Shimadzu (Asia…
Key words
illicit, illiciturine, urinearea, areatreatment, treatmentdrugs, drugspre, preratio, ratioprop, propmetd, metdmetq, metqpcp, pcpquantitation, quantitationnovel, novelmdea, mdeanime
Shimadzu Solutions for Application Notebook - Clinical Research - Application Notebook
2018|Shimadzu|Guides
C10G-E055 Solutions for Clinical Research Application Notebook First Edition: March, 2018 © Shimadzu Corporation, 2018 Solutions for Clinical Research Index Application Notebook Therapeutic Drug Monitoring High-Throughput Optimization of Therapeutic Drug Monitoring Using Fully Automated Sample Preparation LC-MS/MS System (CLAM-2000 +…
Key words
blood, bloodpretreatment, pretreatmentautomated, automateddrugs, drugssample, sampleanalysis, analysismethod, methodmrm, mrmidua, iduaarea, areaurine, urinenorfenefrine, norfenefrinefully, fullymts, mtsgaa
Direct Determination of Plasma Free Amino Acids by Combined MRM-SIM Method on LC/MS/MS
2016|Shimadzu|Applications
Application News AD-0119 Clinical Research / LCMS-8040 Direct Determination of Plasma Free Amino Acids by Combined MRM-SIM Method on LC/MS/MS Zhe Sun, Jie Xing, Pei Yee Khoo* and Zhaoqi Zhan Introduction Unbound or free amino acids in human blood…
Key words
gln, glngly, glysim, simmet, metmrm, mrmtyr, tyrasp, aspglu, glutrp, trpthr, thrile, ilearg, argala, alaleu, leuval