Separation and Quantitation of Oxysterol Compounds Using LC-MS/MS
Applications | 2016 | ShimadzuInstrumentation
Oxysterols are oxidized derivatives of cholesterol that play critical roles in cellular signaling, lipid metabolism, and the pathogenesis of neurodegenerative disorders. Quantifying the spectrum of oxysterol species in biological matrices is essential for elucidating their involvement in disease mechanisms and for developing diagnostic or therapeutic strategies. Advanced analytical platforms capable of sensitive, high-throughput oxysterol profiling are therefore in high demand in research and clinical laboratories.
This work aimed to develop and validate a robust liquid chromatography–tandem mass spectrometry (LC-MS/MS) method for the simultaneous separation and quantitation of sixteen oxysterol and cholesterol-related compounds. The method targets key analytes linked to neurodegenerative and metabolic conditions, enabling researchers to monitor relative abundances and ratio changes in complex biological samples.
The chromatographic separation was achieved on a Shimadzu Nexera UHPLC system equipped with a Shim-pack XR-ODS III reversed-phase column (2.0×150 mm). The mobile phase gradient began at 10% organic and ramped to accelerate late-eluting species, giving a total run time of 21 minutes under a maximum pressure of ~11,000 psi. Injection volume was 2 µL with column temperature maintained at 40°C.
MRM transitions were optimized on a Shimadzu LCMS-8060 triple quadrupole mass spectrometer. Three ionization modalities were compared: electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and dual ionization source (DUIS). Key MS parameters included interface temperature (300–350°C), desolvation line (200–250°C), heat block (200–400°C), and gas flows tailored to each source.
Chromatographic resolution successfully separated isobaric and closely eluting pairs, such as 24-hydroxycholesterol vs. 25-hydroxycholesterol and 7α- vs. 7β-hydroxycholesterol. All sixteen analytes were baseline-resolved for accurate quantitation. MRM transitions were selected based on the two most abundant fragment ions per precursor to maximize sensitivity.
This method provides a comprehensive and time-efficient tool for oxysterol profiling in research or clinical laboratories:
Emerging areas of interest include coupling this workflow with high-resolution mass spectrometry for structural elucidation of novel oxysterols, or integrating microflow LC to further reduce solvent consumption and improve sensitivity. Automation of sample preparation and data processing pipelines will facilitate larger-scale clinical studies and longitudinal biomarker monitoring.
A sensitive, selective, and rapid LC-MS/MS method was established for quantifying a broad set of sixteen oxysterol and cholesterol-related compounds. The dual ion DUIS source proved optimal, combining the advantages of ESI and APCI, and enabling all targets to be analyzed in a single chromatographic run with low picogram detection limits. This platform supports advanced lipidomic investigations and has potential for clinical biomarker development.
No external literature references were provided in the source document.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesClinical Research
ManufacturerShimadzu
Summary
Importance of the Topic
Oxysterols are oxidized derivatives of cholesterol that play critical roles in cellular signaling, lipid metabolism, and the pathogenesis of neurodegenerative disorders. Quantifying the spectrum of oxysterol species in biological matrices is essential for elucidating their involvement in disease mechanisms and for developing diagnostic or therapeutic strategies. Advanced analytical platforms capable of sensitive, high-throughput oxysterol profiling are therefore in high demand in research and clinical laboratories.
Objectives and Study Overview
This work aimed to develop and validate a robust liquid chromatography–tandem mass spectrometry (LC-MS/MS) method for the simultaneous separation and quantitation of sixteen oxysterol and cholesterol-related compounds. The method targets key analytes linked to neurodegenerative and metabolic conditions, enabling researchers to monitor relative abundances and ratio changes in complex biological samples.
Methodology and Instrumentation
The chromatographic separation was achieved on a Shimadzu Nexera UHPLC system equipped with a Shim-pack XR-ODS III reversed-phase column (2.0×150 mm). The mobile phase gradient began at 10% organic and ramped to accelerate late-eluting species, giving a total run time of 21 minutes under a maximum pressure of ~11,000 psi. Injection volume was 2 µL with column temperature maintained at 40°C.
MRM transitions were optimized on a Shimadzu LCMS-8060 triple quadrupole mass spectrometer. Three ionization modalities were compared: electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and dual ionization source (DUIS). Key MS parameters included interface temperature (300–350°C), desolvation line (200–250°C), heat block (200–400°C), and gas flows tailored to each source.
Main Results and Discussion
Chromatographic resolution successfully separated isobaric and closely eluting pairs, such as 24-hydroxycholesterol vs. 25-hydroxycholesterol and 7α- vs. 7β-hydroxycholesterol. All sixteen analytes were baseline-resolved for accurate quantitation. MRM transitions were selected based on the two most abundant fragment ions per precursor to maximize sensitivity.
- Limits of detection (LOD) and quantitation (LOQ) across the panel ranged from sub-picogram to low-picogram on-column, depending on the ion source.
- ESI performed well for most oxysterols but struggled with zymosterol and desmosterol.
- APCI improved detection for those sterols but at the expense of sensitivity for other analytes.
- DUIS combined the strengths of ESI and APCI, producing uniformly low LOD/LOQ values and allowing a single-run analysis.
Benefits and Practical Applications
This method provides a comprehensive and time-efficient tool for oxysterol profiling in research or clinical laboratories:
- Single-run analysis reduces sample handling and instrument downtime.
- DUIS source ensures consistent sensitivity across structurally diverse sterols.
- High chromatographic resolution supports accurate quantitation of isomers and low-abundance species.
- Low picogram detection empowers studies on limited or precious biological samples.
Future Trends and Potential Applications
Emerging areas of interest include coupling this workflow with high-resolution mass spectrometry for structural elucidation of novel oxysterols, or integrating microflow LC to further reduce solvent consumption and improve sensitivity. Automation of sample preparation and data processing pipelines will facilitate larger-scale clinical studies and longitudinal biomarker monitoring.
Conclusion
A sensitive, selective, and rapid LC-MS/MS method was established for quantifying a broad set of sixteen oxysterol and cholesterol-related compounds. The dual ion DUIS source proved optimal, combining the advantages of ESI and APCI, and enabling all targets to be analyzed in a single chromatographic run with low picogram detection limits. This platform supports advanced lipidomic investigations and has potential for clinical biomarker development.
Instrumentation
- Shimadzu Nexera UHPLC system
- Shim-pack XR-ODS III reversed-phase column (2.0×150 mm)
- Shimadzu LCMS-8060 triple quadrupole mass spectrometer
- Ionization sources: ESI, APCI, DUIS
Reference
No external literature references were provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
LC-MS/MS detection of fentanyl and related synthetic opioids in biological matrices
2018|Shimadzu|Posters
PO-CON1816E LC-MS/MS detection of fentanyl and related synthetic opioids in biological matrices ASMS 2018 Robert Lockwood1; Katie Pryor2; Michael Parks2; Christopher Gilles2 1 Connecticut Division of Scientific Services, Middletown, Connecticut/ US; 2 Shimadzu Scientific Instruments Inc., Columbia LC-MS/MS detection of…
Key words
fentanyl, fentanylopioids, opioidscarfentanyl, carfentanylnorfentanyl, norfentanylnorcarfentanyl, norcarfentanylsynthetic, synthetictemp, tempbiological, biologicalacetyl, acetylmatrices, matriceslod, lodrelated, relatedocfentanyl, ocfentanylmethylfentanyl, methylfentanylvaleryl
Simultaneous Quantitation of N-nitrosamines and NDSRI in API and Formulation by Using a DUISTM Ionization Source in LC-MS/MS
2025|Shimadzu|Applications
Liquid Chromatograph Mass Spectrometer LCMS-8060NX Application News Simultaneous Quantitation of N-nitrosamines and NDSRI in API and Formulation by Using a DUISTM Ionization Source in LC-MS/MS Nitish Suryawanshi, Shalu Nair, Samruddha Chavan, Nitin Shukla, Devika Tupe, Siddhesh Ghadi, Ramesh Manigiri, Jitendra…
Key words
loq, loqbelow, belowafa, afandea, ndeandma, ndmaafatinib, afatinibnmea, nmeanpyr, npyrnpip, npipneipa, neipandipa, ndipandpa, ndpandba, ndbanitrosamines, nitrosaminesapi
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8060
2019|Shimadzu|Brochures and specifications
C146-E286F Liquid Chromatograph Mass Spectrometer LCMS-8060 CHANGES EVERYTHING Sensitivity Speed Solutions LCMS-8060 The LCMS-8060 pushes the limits of what we can see. Creating the future of LC/MS/MS analysis can mean making something work with higher sensitivity, go faster or cost…
Key words
mrm, mrmlabsolutions, labsolutionsmin, minmediators, mediatorsscan, scanchromatograph, chromatographsurvey, surveymass, massspectrometer, spectrometerblank, blankspiked, spikedpause, pauseinsight, insightivit, ivitpeptide
Shimadzu Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030
2020|Shimadzu|Brochures and specifications
C146-E365B Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 Effortless Performance The LCMS-9030 quadrupole time-of-flight (Q-TOF) mass spectrometer integrates the world’ s fastest and most sensitive quadrupole technology with TOF architecture. A product of Shimadzu's engineering DNA, speed and effortless performance…
Key words
mma, mmaspectrum, spectrumflight, flightetizolam, etizolamquadrupole, quadrupolemass, massflighttube, flighttubeprobe, probemikros, mikroschromatograph, chromatographshimadzu, shimadzuspectrometer, spectrometerspectra, spectravalinomycin, valinomycinmitomycin