Utilization of the Arc Premier System and XSelect Premier Columns for Sensitive LC-MS Quantitation of the Steroid Phosphates, Betamethasone, Dexamethasone, and Hydrocortisone
Applications | 2021 | WatersInstrumentation
In pharmaceutical R&D and manufacturing, precise quantitation of steroid phosphates is critical for drug efficacy and safety. These analytes tend to interact with metal surfaces in LC systems, causing peak distortion and poor reproducibility, which undermines method reliability.
This study compares the performance of the Arc Premier System with XSelect Premier HSS T3 Column against a standard LC setup to develop a robust, sensitive LC-MS/MS method for quantifying dexamethasone, betamethasone, and hydrocortisone phosphates.
A targeted multiple reaction monitoring (MRM) method was employed using a 6.5-minute gradient with 0.1% formic acid in water (mobile phase A) and acetonitrile (mobile phase B). Calibration ranged from 1–250 ng/mL (steroidal phosphates) and up to 1000 ng/mL for hydrocortisone phosphate, applying 1/x2 weighting.
The Premier system demonstrated immediate out-of-box recovery and sharp peaks for all steroid phosphates, with consistent MS response from the first injection. In contrast, the standard setup required multiple injections to approach comparable performance and exhibited tailing and low recovery. Linearity exceeded R2 of 0.99, accuracy and precision within ±15%, and LLOQs of 1 ng/mL with S/N ratios up to 278.
Advancements may focus on extending high-performance surface technologies to other metal-sensitive analytes, integrating automated workflows for routine QC, and adapting methods for complex biological matrices in pharmacokinetic studies.
The combination of the Arc Premier System and XSelect Premier HSS T3 Column delivers a rapid, reliable LC-MS/MS assay for steroid phosphates with excellent sensitivity and reproducibility, streamlining bioanalytical workflows in drug development.
Consumables, LC/MS, LC/MS/MS, LC columns, LC/QQQ
IndustriesPharma & Biopharma
ManufacturerWaters
Summary
Significance of the Topic
In pharmaceutical R&D and manufacturing, precise quantitation of steroid phosphates is critical for drug efficacy and safety. These analytes tend to interact with metal surfaces in LC systems, causing peak distortion and poor reproducibility, which undermines method reliability.
Objectives and Overview of the Study
This study compares the performance of the Arc Premier System with XSelect Premier HSS T3 Column against a standard LC setup to develop a robust, sensitive LC-MS/MS method for quantifying dexamethasone, betamethasone, and hydrocortisone phosphates.
Methodology
A targeted multiple reaction monitoring (MRM) method was employed using a 6.5-minute gradient with 0.1% formic acid in water (mobile phase A) and acetonitrile (mobile phase B). Calibration ranged from 1–250 ng/mL (steroidal phosphates) and up to 1000 ng/mL for hydrocortisone phosphate, applying 1/x2 weighting.
Used Instrumentation
- Arc Premier System equipped with MaxPeak High Performance Surfaces technology
- XSelect Premier HSS T3 Column (2.5 µm, 2.1 x 100 mm)
- Xevo TQ-S micro Triple Quadrupole Mass Spectrometer
Main Results and Discussion
The Premier system demonstrated immediate out-of-box recovery and sharp peaks for all steroid phosphates, with consistent MS response from the first injection. In contrast, the standard setup required multiple injections to approach comparable performance and exhibited tailing and low recovery. Linearity exceeded R2 of 0.99, accuracy and precision within ±15%, and LLOQs of 1 ng/mL with S/N ratios up to 278.
Benefits and Practical Applications
- Elimination of lengthy system and column passivation procedures
- Improved detection limits and reproducibility for metal-sensitive steroids
- Fast analysis time (6.5 minutes) supporting high throughput
Future Trends and Applications
Advancements may focus on extending high-performance surface technologies to other metal-sensitive analytes, integrating automated workflows for routine QC, and adapting methods for complex biological matrices in pharmacokinetic studies.
Conclusion
The combination of the Arc Premier System and XSelect Premier HSS T3 Column delivers a rapid, reliable LC-MS/MS assay for steroid phosphates with excellent sensitivity and reproducibility, streamlining bioanalytical workflows in drug development.
References
- Viswanathan CT, Bansal S, Booth B, et al. Quantitative Bioanalytical Methods Validation and Implementation: Best Practices for Chromatographic and Ligand Binding Assays. Pharm Res. 2007;24:1962–1973.
- Bansal S, DeStefano A. Key Elements of Bioanalytical Method Validation for Small Molecules. AAPS J. 2007;9:E109–114.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Batch-to-Batch Robustness of MaxPeak Premier Columns for the Analysis Dexamethasone Phosphate and Related Compounds
2021|Waters|Applications
Application Note Batch-to-Batch Robustness of MaxPeak Premier Columns for the Analysis Dexamethasone Phosphate and Related Compounds Fadi L. Alkhateeb, Paul D. Rainville Waters Corporation Abstract An Ultra High Performance Liquid Chromatography method that uses an Arc Premier System with MaxPeak…
Key words
batch, batchpremier, premiermaxpeak, maxpeakcolumns, columnsdexamethasone, dexamethasonephosphate, phosphatearc, arcbatches, batchesreproducibility, reproducibilitymaterial, materialdifferent, differentcompounds, compoundsqda, qdarelated, relatedconsider
Analytical Quality by Design Based Method Development for the Analysis of Dexamethasone Phosphate and Related Compounds Using Arc Premier MaxPeak High Performance Surfaces (HPS) Technology
2021|Waters|Applications
Application Note Analytical Quality by Design Based Method Development for the Analysis of Dexamethasone Phosphate and Related Compounds Using Arc Premier MaxPeak High Performance Surfaces (HPS) Technology Fadi L. Alkhateeb, Paul D. Rainville Waters Corporation Abstract An Ultra High Performance…
Key words
arc, arcpremier, premierdrylab, drylabmethod, methodmaxpeak, maxpeakdevelopment, developmentaqbd, aqbdperformance, performancesystem, systemchelating, chelatingphosphorylated, phosphorylatedhps, hpsprinciples, principlesmetal, metalexperiments
MaxPeak Premier Columns - Small Molecule Application Notebook
2023|Waters|Guides
MaxPeak Premier Columns Small Molecule Application Notebook CONTENTS Introduction Non-Specific Adsorption: Unwanted Interactions................................................................................... 3 How Do Scientists Solve This Problem Today?......................................................................................... 4 Precision Chemistry for Particles and Surfaces........................................................................................ 5 Fundamentals Waters™ Premier Standards to Investigate the Inertness of Chromatographic Systems....................................................................................................................................7…
Key words
maxpeak, maxpeakpremier, premierfundamentals, fundamentalshps, hpsnotebook, notebookacquity, acquitycolumns, columnssurfaces, surfacesmetabolomics, metabolomicsmetal, metalpharmaceutical, pharmaceuticalapplications, applicationsapplication, applicationread, readtechnology
Recovery of Metal-Sensitive Analytes on the Arc Premier Solution: System-to-System Reproducibility and a Multi-System Comparison to Conventional LC
2021|Waters|Applications
Application Note Recovery of Metal-Sensitive Analytes on the Arc Premier Solution: System-to-System Reproducibility and a Multi-System Comparison to Conventional LC Will Martin, Dhvani Shah, Chris Grzonka, Anthony Dovell, Zhimin Li, Paula Hong, Iva Turyan, Jason Dyke Waters Corporation Abstract Metal-sensitive…
Key words
arc, arcpremier, premiermetal, metalconventional, conventionalsolution, solutionsensitive, sensitiveadp, adpatp, atpsystem, systemhps, hpsmaxpeak, maxpeakcomparison, comparisonnonmetal, nonmetalprecision, precisionanalytes