Using Peak Deconvolution of Photodiode Array Data to Attain Faster, Easier Performance of EPA Method 8330B
Applications | 2019 | ShimadzuInstrumentation
Effective detection of low-level explosive residues in environmental matrices is essential for public health protection and contamination control.
This application demonstrates accelerated EPA Method 8330B analysis by incorporating peak deconvolution of photodiode array data, aiming to reduce development time and improve quantitation of coeluting nitroaromatics, nitramines, and nitrate esters.
A Shimadzu Nexera-I MT 3D HPLC system with an LC-2040 PDA detector served as the platform. A primary C-18 column (2.1×200 mm, 5 μm) with a methanol/water mobile phase, including low percentages of acetonitrile, was optimized at 42 °C. LabSolution LC software with i-PDeA II performed multivariate curve resolution alternating least squares deconvolution.
Baseline separations were incomplete, with overlapping peaks for compounds such as Tetryl, NG, TNT isomers, and dinitrotoluenes. Deconvolution resolved these overlaps effectively, achieving spectral similarity indices above 0.999 and calibration linearity (R2 ≥ 0.999) over 2–500 ng on column (2–250 ng for amino isomers).
Peak deconvolution eliminates the need for multiple confirmation columns and extensive method adjustments. It ensures accurate quantitation and spectral verification even for poorly resolved analytes, streamlining QA/QC workflows in environmental analysis.
Combining chemometric deconvolution with UHPLC or high-resolution mass spectrometry may further accelerate analyses and lower detection limits. Automated data processing will support field-deployable screening for rapid environmental monitoring.
Implementing PDA-based deconvolution in EPA Method 8330B significantly reduces method development time while providing reliable quantitation and spectral confirmation of explosive residues.
Shimadzu Application Note No. HPLC-023 High Performance Liquid Chromatography Using Peak Deconvolution of Photodiode Array Data to Attain Faster, Easier Performance of EPA Method 8330B April 2019
HPLC
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Importance of the Topic
Effective detection of low-level explosive residues in environmental matrices is essential for public health protection and contamination control.
Objectives and Study Overview
This application demonstrates accelerated EPA Method 8330B analysis by incorporating peak deconvolution of photodiode array data, aiming to reduce development time and improve quantitation of coeluting nitroaromatics, nitramines, and nitrate esters.
Methodology and Instrumentation
A Shimadzu Nexera-I MT 3D HPLC system with an LC-2040 PDA detector served as the platform. A primary C-18 column (2.1×200 mm, 5 μm) with a methanol/water mobile phase, including low percentages of acetonitrile, was optimized at 42 °C. LabSolution LC software with i-PDeA II performed multivariate curve resolution alternating least squares deconvolution.
- Column: C-18 (ODS)
- Detection: PDA at 254 nm (most analytes) and 210 nm (nitrate esters)
- Software: LabSolution LC with i-PDeA II module
Main Results and Discussion
Baseline separations were incomplete, with overlapping peaks for compounds such as Tetryl, NG, TNT isomers, and dinitrotoluenes. Deconvolution resolved these overlaps effectively, achieving spectral similarity indices above 0.999 and calibration linearity (R2 ≥ 0.999) over 2–500 ng on column (2–250 ng for amino isomers).
Benefits and Practical Applications
Peak deconvolution eliminates the need for multiple confirmation columns and extensive method adjustments. It ensures accurate quantitation and spectral verification even for poorly resolved analytes, streamlining QA/QC workflows in environmental analysis.
Future Trends and Possibilities
Combining chemometric deconvolution with UHPLC or high-resolution mass spectrometry may further accelerate analyses and lower detection limits. Automated data processing will support field-deployable screening for rapid environmental monitoring.
Conclusion
Implementing PDA-based deconvolution in EPA Method 8330B significantly reduces method development time while providing reliable quantitation and spectral confirmation of explosive residues.
Reference
Shimadzu Application Note No. HPLC-023 High Performance Liquid Chromatography Using Peak Deconvolution of Photodiode Array Data to Attain Faster, Easier Performance of EPA Method 8330B April 2019
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of Explosives in Soil Using the Agilent 1260 Infinity III LC System
2025|Agilent Technologies|Applications
Application Note Environmental Determination of Explosives in Soil Using the Agilent 1260 Infinity III LC System Relative response Injector workflows for reliable calibration without manual work Retention time Author Abstract Sonja Schipperges Agilent Technologies, Inc. Both military and certain civil…
Key words
explosives, explosivespetn, petnhmx, hmxhexyl, hexylcalibration, calibrationtetryl, tetrylrdx, rdxretention, retentionmau, mausoil, soilresponse, responsegarden, gardenrelated, relatedinfinitylab, infinitylabmin
Analysis of energetic materials in various water and soil samples using HPLC and LC-MS
|Metrohm|Posters
Analysis of energetic materials in various water and soil samples using HPLC and LC-MS Abstract Diode array detector chromatogram In modern days, a new breed of energetic (explosive) materials is emerging. Traditional aromatic nitrates are still in use, but there…
Key words
tetryl, tetrylenergetic, energetictraditional, traditionalpentaerythritoltetranitrate, pentaerythritoltetranitratepropellant, propellantresiding, residinghplc, hplcundissolved, undissolvednitroglycerin, nitroglycerininterferences, interferencespetn, petnwarfare, warfareinferior, inferiorcontain, containhmx
Dual Confirmation of EPA Method 8330 explosives using DUIS-equipped Single Quadrupole LCMS-2050 and PDA for Tandem Detection
2025|Shimadzu|Posters
ThP 103 Dual Confirmation of EPA Method 8330 explosives using DUIS-equipped Single Quadrupole LCMS-2050 and PDA for Tandem Detection Valeria Zerda-Pinto1; Ethan R Hain1; Lihini Tharanga Mendis1; Sarah Monti1; Samantha A. Olendorff1; Ruth Marfil-Vega1; Landon A Wiest1; Christopher Gilles1 1.…
Key words
min, minexplosives, explosivespetn, petntetryl, tetryltnt, tnthmx, hmxpda, pdardx, rdxexplosive, explosiveloq, loqdna, dnanexera, nexeratetranitrate, tetranitratenitroglycerin, nitroglycerinpentaerythritol
Enhanced Analysis of explosives listed in EPA Method 8330B using APCI-LC-MS/MS
2025|Shimadzu|Posters
ThP 105 Enhanced Analysis of explosives listed in EPA Method 8330B using APCI-LC-MS/MS Valeria Zerda-Pinto1; Sarah A Monti1; Ethan R Hain1; Lihini Mendis1; Samatha A. Olendorff1; Ruth Marfil-Vega1; Landon A Wiest1; Christopher T Gilles1; 1. Shimadzu Scientific Instruments, Columbia, MD…
Key words
cold, coldpetn, petnhmx, hmxrdx, rdxmin, minexplosives, explosiveshot, hottetryl, tetrylriver, rivertnt, tntpatuxent, patuxentdna, dnanexera, nexeraionization, ionizationtetranitrate