Low Sample Carryover with Key Performance Indicators on the Alliance HPLC System
Technical notes | 2012 | WatersInstrumentation
The elimination of sample carryover in HPLC analyses is fundamental for achieving accurate quantification, preserving system performance, and avoiding costly downtime for maintenance. Carryover can compromise data integrity, particularly in methods requiring high sensitivity and broad dynamic range, and may even necessitate system cleaning or part replacement if levels become severe.
This work assesses the Alliance HPLC System’s injector design and wash strategies by challenging it with two representative compounds:
The study compares standard wash conditions against an advanced needle-wash feature to quantify carryover reduction.
The investigation employed the following:
The system’s carryover specification of 0.01% was used as the benchmark.
Under the standard wash protocol, carryover for both caffeine and chlorhexidine was at least six times lower than the 0.01% threshold, demonstrating effective removal of both volumetric and surface-adsorbed residues. Increasing the wash time via the advanced wash feature further reduced residual peaks, proving that:
These findings confirm that the flow-through-needle design and flexible wash parameters effectively address multiple origins of carryover.
The Alliance HPLC System’s approach offers several advantages:
Analysts benefit from simplified method set-up and robust performance across a wide dynamic range.
Emerging directions for carryover management include:
Such innovations will enhance throughput, data quality, and method flexibility.
The Alliance HPLC System’s flow-through-needle injector and customizable wash strategies deliver outstanding carryover control for both soluble and surface-active analytes. By combining efficient wash cycles, solvent versatility, and an advanced needle-wash feature, the system ensures reliable quantification, reduces maintenance demands, and supports a wide array of analytical workflows.
HPLC
IndustriesManufacturerWaters
Summary
Importance of the Topic
The elimination of sample carryover in HPLC analyses is fundamental for achieving accurate quantification, preserving system performance, and avoiding costly downtime for maintenance. Carryover can compromise data integrity, particularly in methods requiring high sensitivity and broad dynamic range, and may even necessitate system cleaning or part replacement if levels become severe.
Aims and Overview of the Study
This work assesses the Alliance HPLC System’s injector design and wash strategies by challenging it with two representative compounds:
- Caffeine (a highly soluble, non-adsorbing probe for volumetric carryover)
- Chlorhexidine (a surface-active, ‘sticky’ analyte that tests chemical adsorption removal)
The study compares standard wash conditions against an advanced needle-wash feature to quantify carryover reduction.
Methodology and Instrumentation
The investigation employed the following:
- Alliance HPLC System with flow-through-needle injector architecture
- Mobile-phase gradient that continuously washes the needle interior
- Seal-pack exterior wash using a user-selected solvent compatible with the sample matrix
- Variable wash durations: standard, double (2×), and extended (4×) cycle times
- Test injections: high-concentration standards (caffeine 5 mg/mL; chlorhexidine 2 mg/mL) followed by blank runs
The system’s carryover specification of 0.01% was used as the benchmark.
Main Results and Discussion
Under the standard wash protocol, carryover for both caffeine and chlorhexidine was at least six times lower than the 0.01% threshold, demonstrating effective removal of both volumetric and surface-adsorbed residues. Increasing the wash time via the advanced wash feature further reduced residual peaks, proving that:
- Extended wash cycles yield incremental improvements in carryover suppression
- Solvent selection and wash duration can be tailored to analyte properties
These findings confirm that the flow-through-needle design and flexible wash parameters effectively address multiple origins of carryover.
Benefits and Practical Applications
The Alliance HPLC System’s approach offers several advantages:
- High reproducibility in peak areas by minimizing cross-sample contamination
- Improved uptime and reduced maintenance through efficient onboard cleaning
- Adaptable wash solvent options to accommodate diverse sample chemistries
- Scalability for routine QA/QC labs and complex method development workflows
Analysts benefit from simplified method set-up and robust performance across a wide dynamic range.
Future Trends and Potential Applications
Emerging directions for carryover management include:
- Smart wash protocols that adjust duration and solvent composition based on real-time analyte detection
- Advanced inert coatings and microfluidic injector designs to further reduce adsorption
- Integration of machine-learning tools to predict carryover risks and optimize cleaning cycles
- Expansion into high-throughput and automated platforms for pharmaceutical and environmental analysis
Such innovations will enhance throughput, data quality, and method flexibility.
Conclusion
The Alliance HPLC System’s flow-through-needle injector and customizable wash strategies deliver outstanding carryover control for both soluble and surface-active analytes. By combining efficient wash cycles, solvent versatility, and an advanced needle-wash feature, the system ensures reliable quantification, reduces maintenance demands, and supports a wide array of analytical workflows.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Improved Chlorhexidine Carryover Performance Using the Alliance iS HPLC System
2023|Waters|Applications
Application Note Improved Chlorhexidine Carryover Performance Using the Alliance iS HPLC System Jennifer Simeone, Paula Hong Waters Corporation Abstract Carryover is an all-too-common problem for many users of High Performance Liquid Chromatography (HPLC) systems. There are multiple forms of carryover,…
Key words
chlorhexidine, chlorhexidinecarryover, carryoveralliance, allianceimproved, improvedhplc, hplcperformance, performancesystem, systemusing, usingwash, washunswept, unsweptvolumetric, volumetricadsorptive, adsorptivevolumes, volumesneedle, needlevoid
Alliance Carryover Performance Part 1: Carryover Improvement Achieved Through Instrument Design Changes for the Alliance HPLC System
2018|Waters|Technical notes
[ APPLICATION NOTE ] Alliance Carryover Performance Part 1: Carryover Improvement Achieved Through Instrument Design Changes for the Alliance HPLC System Amanda B. Dlugasch, Jennifer Simeone, Patricia R. McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS ■■ INTRODUCTION Improved carryover…
Key words
carryover, carryoveralliance, alliancecoumarin, coumarincaffeine, caffeinechlorhexidine, chlorhexidinehplc, hplcquetiapine, quetiapinefumarate, fumarateneedle, needleseal, sealdesign, designnormal, normalsolution, solutionsystem, systempack
Exceptional Carryover Performance of the Arc HPLC System for Chlorhexidine Sample
2021|Waters|Technical notes
Application Note Exceptional Carryover Performance of the Arc HPLC System for Chlorhexidine Sample Bheeshmacharyulu S, Dilshad Pullancheri, Tirupateswara Rao Boyidi, Dr. Padmakar Wagh Waters Corporation Abstract Chlorhexidine is a broad-spectrum antimicrobial biguanide used as a topical antiseptic and in dental…
Key words
chlorhexidine, chlorhexidinearc, arccarryover, carryoverexceptional, exceptionalhplc, hplcsystem, systemperformance, performancesample, samplegluconate, gluconatecompetitor, competitormonograph, monographinjector, injectorissue, issuebiguanide, biguanideusp
Alliance iS HPLC System: A New Era of Intuitive Simplicity
2024|Waters|Brochures and specifications
Alliance iS HPLC System: A New Era of Intuitive Simplicity HPLC Analysis Water-Soluble Vitamins Hydrophilic Interaction Troubleshooting Peak Splitting Azithromycin Analysis Atorvastatin Analysis Pharmaceutical Drug Analysis Method Migration Naphazoline Hydrochloride and Pheniramine Maleate Method Modernization Chlorhexidine Carryover USP Ibuprofen Assay…
Key words
alliance, alliancehplc, hplcnaphazoline, naphazolineusp, usppheniramine, pheniramineazithromycin, azithromycinatorvastatin, atorvastatinmigration, migrationmodernization, modernizationchlorhexidine, chlorhexidinemaleate, maleatemethod, methodhydrochloride, hydrochlorideanalysis, analysiscarryover