Others | 2013 | WatersInstrumentation
HPLC
Industries
Manufacturer
Waters

Summary

Significance of the Topic


Refractive index (RI) detection remains the only universal detection mode for analytes lacking UV chromophores. The integration of low-dispersion RI detection into UltraPerformance LC (UPLC) and Advanced Polymer Chromatography (APC) workflows addresses long-standing challenges related to peak broadening, temperature instability, and baseline drift. By minimizing system dispersion and enhancing thermal control, high-resolution, rapid analyses of non-UV absorbing compounds become feasible, improving data quality and laboratory throughput.

Objectives and Overview


This study presents the design features and performance evaluation of the ACQUITY RI Detector when coupled to UPLC and APC systems. Key goals include reducing flow cell volume, optimizing thermal management, integrating fluidic controls, and validating analytical performance for food sugar profiling, polymer molecular weight analysis, and method transfer from HPLC to UPLC.

Methodology


The detector architecture focuses on:
  • Flow cell miniaturization (1.3 µL volume) to preserve narrow UPLC peak shapes and reduce dispersion.
  • Thermal stabilization via a countercurrent heat exchanger that equilibrates incoming solvent temperature and isolates optics from environmental fluctuations.
  • Automated fluidic management, including a solvent recycle valve and reference cell purging for baseline stability.
  • Enhanced data acquisition with sampling rates up to 80 points per second and independent filter time-constant adjustments.
  • High-energy, long-lifetime LED light source to maximize sensitivity.

Instrumentation


The ACQUITY RI Detector was evaluated on:
  • ACQUITY UPLC H-Class System for isocratic and gradient separations.
  • ACQUITY APC System equipped with rigid small-particle columns (125 Å/45 Å pore sizes) for polymer analysis.

Main Results and Discussion


Food sugar profiling: Five common sugars (fructose, glucose, sucrose, maltose, lactose) were resolved in under four minutes across matrices (mixed standards, orange juice, corn syrup, low-fat milk) with excellent peak shape and quantification precision.

Polymer molecular weight analysis: Comparison between a standard HPLC RI detector and the low-dispersion ACQUITY RI Detector on APC columns revealed dramatic improvements in resolving low-molecular-weight oligomers. The reduced detector volume preserved sharp polymer peaks, enabling accurate polydispersity and molecular weight determinations.

HPLC to UPLC method transfer: The official USP method for Topiramate was successfully adapted to UPLC/RI, achieving an 80 % reduction in runtime while maintaining equivalent chromatographic resolution and quantitation performance.

Benefits and Practical Applications


The ACQUITY RI Detector facilitates:
  • Universal detection for compounds without UV absorbance across food, pharmaceutical, and industrial QC workflows.
  • Faster analyses with up to 80 % reduced runtimes, boosting sample throughput.
  • Lower solvent consumption through integrated solvent recycling, reducing operational costs.
  • High-resolution polymer characterization critical for advanced materials research.

Future Trends and Applications


As UPLC and APC technologies evolve, low-dispersion RI detection will enable:
  • Integration with two-dimensional LC systems for comprehensive profiling of complex matrices.
  • Online coupling with fraction collectors and mass spectrometry for multi-detector workflows.
  • Automated method development tools leveraging AI to optimize RI-based separations.
  • Expanded applications in biopolymers, metabolomics, and reaction monitoring.

Conclusion


The ACQUITY RI Detector represents a significant advancement in refractive index detection, delivering low dispersion, robust thermal control, and enhanced sensitivity. When combined with UPLC and APC platforms, it extends universal detection to high-throughput, high-resolution analyses of non-UV absorbing compounds, supporting diverse laboratories in quality control and research.

References


Not applicable

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
ACQUITY Advanced Polymer Chromatography (APC)
ACQUITY Advanced Polymer Chromatography (APC) Applications Notebook [ APPLICATION highlight ] TABLE OF CONTENTS LIST OF POLYMERS ANALY ZED USING ACQUIT Y APC Polymer Part Number Page Poly(styrene) in THF Calibration Curves WA64641 3 Poly(vinylacetate) in THF WA64642 4…
Key words
apc, apcthf, thfpoly, polyacquity, acquityµriu, µriuhighlight, highlightpolymer, polymermethacrylate, methacrylateadvanced, advancedminutes, minutesvinylacetate, vinylacetatecolumns, columnsformaldehyde, formaldehydechromatography, chromatographymethyl
Rapid Method Screening for High Resolution Aqueous SEC Measurements
Rapid Method Screening for High Resolution Aqueous SEC Measurements Mike O’Leary y p International GPC Symposium Philadelphia 2015 ©2015 Waters Corporation 1 Target Attributes: New Platform for SEC Polymer y Characterization  Column – Low energy surface – minimal adsorption…
Key words
apc, apcpoly, polyminutes, minutesbeh, behacquity, acquitysolvent, solventdiol, diolsurface, surfacedispersion, dispersionpore, porehybrid, hybridbridged, bridgedcoverage, coveragegood, goodparticle
Parameters Affecting the Accuracy of Molar Mass Distribution Determined by SEC with a MALS Detector
Rapid Method Screening for High Resolution Aqueous SEC Measurements Mike O’Leary y p International GPC Symposium Philadelphia 2015 ©2015 Waters Corporation 1 Target Attributes: New Platform for SEC Polymer y Characterization  Column – Low energy surface – minimal adsorption…
Key words
apc, apcpoly, polyminutes, minutesbeh, behacquity, acquitysolvent, solventdiol, diolsurface, surfacedispersion, dispersionpore, porehybrid, hybridbridged, bridgedcoverage, coveragegood, goodparticle
High-Speed, High-Resolution Analysis of Low Molecular Weight Polymers Using the Advanced Polymer Chromatography (APC) System
High-Speed, High-Resolution Analysis of Low Molecular Weight Polymers Using the Advanced Polymer Chromatography (APC) System Mia Summers and Michael O’Leary Waters Corporation, Milford, MA, U.S.A. A P P L I C AT I O N B E N E F…
Key words
apc, apcgpc, gpcacquity, acquitypolymer, polymerµriu, µriumolecular, molecularthf, thfweight, weightoligomers, oligomerspolymers, polymersconventional, conventionalsystem, systemminutes, minuteslow, lowadvanced
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike