Analysis of vitamin D in cheese by APCI LC-MS/MS using the Bruker EVOQ™ LC-TQ Elite mass spectrometry system
Applications | 2018 | BrukerInstrumentation
Vitamin D is critical for calcium homeostasis and bone health, and its deficiency or excess poses significant health risks. Cheese, a widely consumed dairy product, can be enriched with vitamin D2 and D3 to help meet dietary recommendations. Accurate determination of vitamin D content in enriched cheese is essential for regulatory compliance and consumer safety.
The primary goal of this study was to develop and validate a rapid, simple, and robust analytical method for quantifying vitamin D2 and D3 in cheese. The method employs liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) with an atmospheric pressure chemical ionization (APCI) interface on the Bruker EVOQ™ LC-TQ Elite system. Key performance criteria included sensitivity, linearity, precision, and recovery.
Sample Preparation:
Instrumentation:
Linearity and Calibration:
The method exhibited excellent linearity (R² > 0.999) over 0.5–1000 µg/kg, with RSD < 25%.
Sensitivity:
Limit of quantification (LOQ) was 0.5 µg/kg and limit of detection (LOD) < 0.1 µg/kg (S/N > 100).
Precision:
Repeatability studies at 0.5, 1, 10, 50, and 100 ppb yielded RSDs < 3.5%.
Accuracy:
Recovery experiments on quark and cottage cheese spiked at 250 ppb and 100 ppb showed recoveries between 90% and 95%, confirming method accuracy.
The developed protocol combines high sensitivity and throughput with minimal sample preparation, making it suitable for routine quality control in food laboratories. Its robustness allows reliable monitoring of vitamin D enrichment in cheese, ensuring compliance with EFSA dietary intake and upper limit guidelines.
Advances may include:
A validated APCI-LC-MS/MS method on the Bruker EVOQ™ LC-TQ Elite enables rapid, sensitive, and accurate determination of vitamin D2 and D3 in cheese. The approach meets stringent analytical criteria and supports regulatory compliance and consumer safety in food quality control laboratories.
[1] MPI Technical Paper No. 2014/03, Determination of vitamin D in foods: Current knowledge and data gaps, New Zealand Government
[2] von Hurst PR, Stonehouse W, Coad J. Vitamin D supplementation reduces insulin resistance in South Asian women living in New Zealand. Br J Nutr. 2010;103(4):549-555.
[3] Verkaik-Kloosterman J, Seves SM, Ocké MC. Vitamin D concentrations in fortified foods and dietary supplements intended for infants. Food Chem. 2017;221:629-635.
[4] EFSA Panel on Dietetic Products, Nutrition and Allergies (NDA). Dietary reference values for vitamin D. EFSA J. 2016;14(10):4547.
[5] Martin R, van’t Slot G. Online Analysis of 25-OH-Vitamin D2/D3 in Plasma and Serum with the EVOQ LC-Triple Quad. Bruker App Note LCMS-103; 2015.
[6] EFSA Panel on Dietetic Products, Nutrition and Allergies (NDA). Tolerable Upper Intake Level of vitamin D. EFSA J. 2012;10(7):2813.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesFood & Agriculture
ManufacturerBruker
Summary
Significance of the Topic
Vitamin D is critical for calcium homeostasis and bone health, and its deficiency or excess poses significant health risks. Cheese, a widely consumed dairy product, can be enriched with vitamin D2 and D3 to help meet dietary recommendations. Accurate determination of vitamin D content in enriched cheese is essential for regulatory compliance and consumer safety.
Objectives and Study Overview
The primary goal of this study was to develop and validate a rapid, simple, and robust analytical method for quantifying vitamin D2 and D3 in cheese. The method employs liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) with an atmospheric pressure chemical ionization (APCI) interface on the Bruker EVOQ™ LC-TQ Elite system. Key performance criteria included sensitivity, linearity, precision, and recovery.
Methodology and Used Instrumentation
Sample Preparation:
- Homogenize 10 g of cheese and add 40 mL pyrogallol solution (2%) and 20 mL 50% KOH.
- Introduce nitrogen and agitate overnight for saponification.
- Extract with 30 mL hexane containing 12.5 mg/L BHT.
- Evaporate 10 mL extract, reconstitute in 1 mL acetonitrile/water (70/30), and filter through 0.45 µm PTFE.
Instrumentation:
- Bruker EVOQ™ LC-TQ Elite MS/MS with APCI source (spray current ±100 µA, probe temp 250 °C).
- Bruker Elute™ UHPLC, Intensity Solo C18 column (100 × 2.1 mm).
- Mobile phases: water +0.1% formic acid and methanol +0.1% formic acid, gradient elution over 9 min.
- MRM transitions: vitamin D3 385.5 > 259.1 (quantifier), vitamin D2 397.6 > 107.0, with confirmation ions.
Main Results and Discussion
Linearity and Calibration:
The method exhibited excellent linearity (R² > 0.999) over 0.5–1000 µg/kg, with RSD < 25%.
Sensitivity:
Limit of quantification (LOQ) was 0.5 µg/kg and limit of detection (LOD) < 0.1 µg/kg (S/N > 100).
Precision:
Repeatability studies at 0.5, 1, 10, 50, and 100 ppb yielded RSDs < 3.5%.
Accuracy:
Recovery experiments on quark and cottage cheese spiked at 250 ppb and 100 ppb showed recoveries between 90% and 95%, confirming method accuracy.
Benefits and Practical Applications
The developed protocol combines high sensitivity and throughput with minimal sample preparation, making it suitable for routine quality control in food laboratories. Its robustness allows reliable monitoring of vitamin D enrichment in cheese, ensuring compliance with EFSA dietary intake and upper limit guidelines.
Future Trends and Applications
Advances may include:
- Integration with automated sample preparation platforms for higher throughput.
- Expansion to simultaneous quantification of other fat-soluble vitamins (A, E, K).
- Application to a wider range of dairy matrices and fortified foods.
- Use of high-resolution mass spectrometry for enhanced identification of vitamin D metabolites.
Conclusion
A validated APCI-LC-MS/MS method on the Bruker EVOQ™ LC-TQ Elite enables rapid, sensitive, and accurate determination of vitamin D2 and D3 in cheese. The approach meets stringent analytical criteria and supports regulatory compliance and consumer safety in food quality control laboratories.
Reference
[1] MPI Technical Paper No. 2014/03, Determination of vitamin D in foods: Current knowledge and data gaps, New Zealand Government
[2] von Hurst PR, Stonehouse W, Coad J. Vitamin D supplementation reduces insulin resistance in South Asian women living in New Zealand. Br J Nutr. 2010;103(4):549-555.
[3] Verkaik-Kloosterman J, Seves SM, Ocké MC. Vitamin D concentrations in fortified foods and dietary supplements intended for infants. Food Chem. 2017;221:629-635.
[4] EFSA Panel on Dietetic Products, Nutrition and Allergies (NDA). Dietary reference values for vitamin D. EFSA J. 2016;14(10):4547.
[5] Martin R, van’t Slot G. Online Analysis of 25-OH-Vitamin D2/D3 in Plasma and Serum with the EVOQ LC-Triple Quad. Bruker App Note LCMS-103; 2015.
[6] EFSA Panel on Dietetic Products, Nutrition and Allergies (NDA). Tolerable Upper Intake Level of vitamin D. EFSA J. 2012;10(7):2813.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Rapid and simple determination of low level fipronil and amitraz in egg samples using the Bruker EVOQ™ LC-TQ Elite MS/MS system
2018|Bruker|Applications
Rapid and simple determination of low level fipronil and amitraz in egg samples using the Bruker EVOQ™ LC-TQ Elite MS/MS system Abstract We have developed a fast, sensitive, and reliable method to determine the presence of fipronil and amitraz in…
Key words
amitraz, amitrazfipronil, fipronilbruker, brukerinsensity, insensitysulfone, sulfoneevoq, evoqmetabolite, metaboliteelite, eliteeggs, eggsegg, eggdmpf, dmpfcompounds, compoundsspiked, spikedinsecticide, insecticidetpp
Estimation of Fat Soluble Vitamins, Ergocalciferol, and Cholecalciferol in Edible Oil
2017|Agilent Technologies|Applications
Estimation of Fat Soluble Vitamins, Ergocalciferol, and Cholecalciferol in Edible Oil Using the Agilent 6470 Triple Quadrupole LC/MS System Coupled to an Agilent 1290 Infinity II LC System Application Note Food Testing and Agriculture Authors Abstract Dr. Priti Amritkar Ergocalciferol…
Key words
ergocalciferol, ergocalciferolcholecalciferol, cholecalciferolsample, samplecounts, countsacquisition, acquisitionapparatus, apparatusvitamin, vitaminconcentration, concentrationadopted, adoptedreflux, refluxmin, minedible, edibleoil, oilvitamins, vitaminsapci
Ultrasensitive Detection and Quantitation of Chloramphenicol using the new Elute UHPLC and EVOQ LC-TQ System
2017|Bruker|Applications
Ultrasensitive Detection and Quantitation of Chloramphenicol using the new Elute UHPLC and EVOQ LC-TQ System Quantitative Analysis of Chloramphenicol in Eggs using the EVOQ Elite TQ coupled to the Elute UHPLC. A liquid chromatography tandem mass spectrometr y (LC-MS/MS) method…
Key words
kcps, kcpschloramphenicol, chloramphenicolbruker, brukerevoq, evoqminutes, minutesegg, eggelute, elutehumans, humansadverse, adverseuhplc, uhplccentrifuge, centrifugechloroamphenicol, chloroamphenicolultrasonicated, ultrasonicatedert, ertlinked
Bruker Product Overview - Life Science Mass Spectrometry
2020|Bruker|Brochures and specifications
Product Overview Life Science Mass Spectrometry Innovation with Integrity Mass Spectrometry Empowering Science with Innovation and Integrity As one of the world’s leading analytical instrumentation companies, Bruker offers a broad spectrum of advanced solutions in all fields of research and…
Key words
maldi, malditof, tofbruker, brukerrapiflex, rapiflexpasef, pasefevoq, evoqspectrometry, spectrometryscimax, scimaxmass, masstimstof, timstofmrms, mrmsproteomics, proteomicsmetaboscape, metaboscapelrf, lrfseries