Rapid and sensitive UHPLC screening of food dyes in carbonated beverages using UV/Vis wavelength switching

Applications | 2017 | Thermo Fisher ScientificInstrumentation
HPLC
Industries
Food & Agriculture
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


Rapid and sensitive detection of food dyes in carbonated beverages is essential to ensure regulatory compliance and protect consumer health. Many synthetic dyes are restricted or phased out due to potential adverse effects, thus fast and accurate screening methods are required.

Study Objectives and Overview


  • Demonstrate the Hypersil GOLD VANQUISH C18 column and Vanquish Flex Binary UHPLC for rapid separation of ten common dyes.
  • Assess method linearity, reproducibility, and recovery in beverage samples.
  • Illustrate the use of UV/Vis wavelength switching to enhance sensitivity.

Methodology and Instrumentation


Separation employed a 50 mm × 2.1 mm, 1.9 μm Hypersil GOLD VANQUISH C18 column with a binary UHPLC system equipped with a diode array detector and a 2 μL LightPipe flow cell. Mobile phases consisted of 20 mM ammonium phosphate buffer (pH 8.8) and a 50/50 buffer–acetonitrile mixture under a gradient at 0.55 mL/min and 40 °C column temperature. Samples were degassed, diluted, spiked with standards, and filtered through 0.45 μm nylon filters. Calibration standards covered 1–30 μg/mL using primary and mixed working solutions prepared in water. Data was acquired and processed with Chromeleon 7.2 software.

Main Results and Discussion


  • Baseline separation of ten dyes was achieved in approximately two minutes with resolution >1.5.
  • Reproducibility was excellent (retention time, peak area, and width RSDs generally <0.5%).
  • Calibration curves were linear (R2 between 0.9997 and 1.0000).
  • Screening of three commercial beverages detected new coccine and sunset yellow FCF; concentrations ranged from 1.9 to 97.7 μg/mL.
  • Spike recoveries were ≥80% (mostly ≥95%), indicating minimal matrix interference.
  • Wavelength switching based on 3D diode array data improved signal-to-noise ratios by up to nine-fold.

Benefits and Practical Applications


This UHPLC method enables high-throughput, reliable screening of dyes in beverages, offering rapid run times and robust quantification with low detection limits. It supports quality control and regulatory monitoring in food and beverage industries.

Future Trends and Potential Applications


Further developments may include integration with mass spectrometry for confirmatory analysis, extension to other matrices such as juices and wines, automation of sample preparation and data processing, and real-time in-line monitoring in production environments.

Conclusion


The combination of the Hypersil GOLD VANQUISH C18 column and Vanquish Flex Binary UHPLC system delivers rapid, sensitive, and reproducible analysis of food dyes in carbonated beverages. Wavelength switching significantly enhances sensitivity, making this approach well suited for routine food safety and quality control.

References


  • Lamb A. Application Note 21672. Thermo Fisher Scientific; Runcorn, UK; 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Fast HPLC Analysis of Dyes in Foods and Beverages
Fast HPLC Analysis of Dyes in Foods and Beverages
2016|Thermo Fisher Scientific|Applications
Application Note 245 Fast HPLC Analysis of Dyes in Foods and Beverages Introduction Dyes have many applications in the food and beverage industries, such as being used to make food more appealing, hide defects, or to strengthen consumer perception of…
Key words
fcf, fcfsunset, sunsetyellow, yellowtartrazine, tartrazinelin, lindyes, dyesorange, orangeabsorbance, absorbancespiked, spikedcoccine, coccinesample, samplemau, mauminutes, minutesgreen, greendye
Fast, reproducible UHPLC-UV determination of dyes found in beverages
APPLICATION NOTE 21733 Fast, reproducible UHPLC-UV determination of dyes found in beverages Author Application benefits Aaron Lamb, Thermo Fisher Scientific, Runcorn, UK • The Thermo Scientific™ Acclaim™ PolarAdvantage II (PA2) column and Thermo Scientific™ Vanquish™ Flex system proved capable for…
Key words
flex, flexvanquish, vanquishindigo, indigocarmine, carmineeosin, eosinbeng, bengcocc, cocceryth, erythindig, indigphlo, phlormin, rminsuns, sunstartr, tartrgree, greellow
Multiple Wavelength Data Acquisition with the Vanquish Variable Wavelength Detector
Holger Franz and Alexandra Manka Thermo Fisher Scientific, Germering, Germany Te ch ni cal N ote 1 6 9 Multiple Wavelength Data Acquisition with the Vanquish Variable Wavelength Detector Key Words Vanquish, VWD, UV-Vis, Absorption, Wavelength Switching Goal To explain…
Key words
wavelength, wavelengthmultichannel, multichannelmau, mauacquisition, acquisitionvanquish, vanquishchannel, channelcoccine, coccinesingle, singlevwd, vwdcyanosine, cyanosineabsorbance, absorbanceswitching, switchingbengal, bengalgrating, gratingerythrosine
Beverages Applications Notebook - Additives in Beverages
Beverages Applications Notebook Additives in Beverages Table of Contents Index of Analytes......................................................................................................................................................................... 3 Introduction to Beverages........................................................................................................................................................... 4 UltiMate 3000 UHPLC+ Systems............................................................................................................................................... 5 IC and RFIC Systems.................................................................................................................................................................. 6 MS Instruments........................................................................................................................................................................... 7 Chromeleon 7 Chromatography Data System Software............................................................................................................ 8 Process Analytical Systems…
Key words
rebaudioside, rebaudiosideels, elsstevioside, steviosideionpac, ionpacsteviol, steviolaerosol, aerosolcharged, chargedammonium, ammoniumglycosides, glycosidesquaternary, quaternaryfcf, fcfbeverages, beveragesalkanol, alkanolsulfite, sulfitesubstrate
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike