Method transfer of a USP derived acetaminophen assay from an Agilent 1260 Infinity system to an UltiMate 3000 SD system and a Vanquish Flex UHPLC system
Applications | 2018 | Thermo Fisher ScientificInstrumentation
The reliable transfer of liquid chromatographic methods between instruments is critical in pharmaceutical analysis, quality control, and regulated environments. Maintaining identical retention times, resolution, and sensitivity avoids costly revalidation and ensures consistent results across laboratories. Modern HPLC platforms with flexible gradient delay volumes and adjustable thermal controls facilitate smooth method transfer, reducing downtime and enhancing productivity.
This study demonstrates the transfer of a USP-derived assay for acetaminophen and its impurities from an Agilent 1260 Infinity quaternary system to two Thermo Fisher Scientific platforms: the UltiMate 3000 Standard-D quaternary system and the Vanquish Flex UHPLC quaternary system. The goal is to achieve equivalent chromatographic performance without altering the original method parameters by leveraging hardware adaptations.
Chromatographic conditions were kept constant: Hypersil GOLD C8 column (4.6×100 mm, 3 µm), phosphate buffer/methanol gradient, 1 mL/min flow, 35 °C column temperature, 230 nm detection, 1 µL injection. Sample contains 1 mg/mL acetaminophen and 10 µg/mL of each impurity.
Instrumentation summary
On the UltiMate 3000 SD, installing a 7 µL pre-heater and increasing static mixer volume from 350 µL to 750 µL reduced gradient delay volume (GDV) differences. A gradient pre-start offset of –0.27 min aligned retention times within 1 % of the original. For the Vanquish Flex, replacing the default 25 µL loop with 100 µL and raising idle volume to 68 µL achieved comparable retention (≤1.7 % deviation) without a gradient pre-start. Column thermostat mode (still vs forced air) had minimal impact under these conditions. Peak resolution, asymmetry, and precision (RSD < 1 %) were maintained across all systems. Signal-to-noise ratios improved on both Thermo systems, and using the LightPipe DAD HL with a 60 mm flow cell delivered the highest sensitivity.
The development of adaptive system architectures with real-time GDV calibration may further streamline transfers. Integration of machine-learning algorithms for automated parameter optimization and the increasing use of high-sensitivity detectors will expand capabilities for challenging trace analyses and ultra-high-throughput environments.
Through strategic use of eluent pre-heaters, variable static mixers, sample loops, and adjustable idle volumes, the acetaminophen assay was successfully transferred from an Agilent 1260 Infinity to UltiMate 3000 SD and Vanquish Flex systems. Equivalent chromatographic performance and enhanced sensitivity were achieved without altering the validated method, underscoring the value of flexible hardware solutions in method migration.
HPLC
IndustriesPharma & Biopharma
ManufacturerAgilent Technologies, Thermo Fisher Scientific
Summary
Importance of the topic
The reliable transfer of liquid chromatographic methods between instruments is critical in pharmaceutical analysis, quality control, and regulated environments. Maintaining identical retention times, resolution, and sensitivity avoids costly revalidation and ensures consistent results across laboratories. Modern HPLC platforms with flexible gradient delay volumes and adjustable thermal controls facilitate smooth method transfer, reducing downtime and enhancing productivity.
Objectives and study overview
This study demonstrates the transfer of a USP-derived assay for acetaminophen and its impurities from an Agilent 1260 Infinity quaternary system to two Thermo Fisher Scientific platforms: the UltiMate 3000 Standard-D quaternary system and the Vanquish Flex UHPLC quaternary system. The goal is to achieve equivalent chromatographic performance without altering the original method parameters by leveraging hardware adaptations.
Methodology and instrumentation
Chromatographic conditions were kept constant: Hypersil GOLD C8 column (4.6×100 mm, 3 µm), phosphate buffer/methanol gradient, 1 mL/min flow, 35 °C column temperature, 230 nm detection, 1 µL injection. Sample contains 1 mg/mL acetaminophen and 10 µg/mL of each impurity.
Instrumentation summary
- Origin system: Agilent 1260 Infinity quaternary pump, thermostatted autosampler, diode array detector.
- Target systems: UltiMate 3000 SD quaternary pump with optional 7 µL eluent pre-heater and interchangeable static mixers; Vanquish Flex quaternary pump with active pre-heater, exchangeable sample loops, adjustable autosampler idle volume, and LightPipe™ DAD options.
Main results and discussion
On the UltiMate 3000 SD, installing a 7 µL pre-heater and increasing static mixer volume from 350 µL to 750 µL reduced gradient delay volume (GDV) differences. A gradient pre-start offset of –0.27 min aligned retention times within 1 % of the original. For the Vanquish Flex, replacing the default 25 µL loop with 100 µL and raising idle volume to 68 µL achieved comparable retention (≤1.7 % deviation) without a gradient pre-start. Column thermostat mode (still vs forced air) had minimal impact under these conditions. Peak resolution, asymmetry, and precision (RSD < 1 %) were maintained across all systems. Signal-to-noise ratios improved on both Thermo systems, and using the LightPipe DAD HL with a 60 mm flow cell delivered the highest sensitivity.
Benefits and practical applications
- Hardware adjustments (pre-heaters, mixers, sample loops, idle volume) enable transfer without method revalidation.
- Enhanced flexibility supports migration from legacy to UHPLC systems.
- Improved sensitivity via LightPipe technology benefits trace impurity analysis.
Future trends and potential applications
The development of adaptive system architectures with real-time GDV calibration may further streamline transfers. Integration of machine-learning algorithms for automated parameter optimization and the increasing use of high-sensitivity detectors will expand capabilities for challenging trace analyses and ultra-high-throughput environments.
Conclusion
Through strategic use of eluent pre-heaters, variable static mixers, sample loops, and adjustable idle volumes, the acetaminophen assay was successfully transferred from an Agilent 1260 Infinity to UltiMate 3000 SD and Vanquish Flex systems. Equivalent chromatographic performance and enhanced sensitivity were achieved without altering the validated method, underscoring the value of flexible hardware solutions in method migration.
Reference
- Swartz ME, Krull I. Analytical Method Transfer; LCGC North America 2006, 24(11):1204–1214.
- Ermer J et al. The transfer of analytical procedures; J Pharm Biomed Anal 2013, 85:262–276.
- Paul C, Grübner M et al. Thermo Scientific White Paper 72711: An instrument parameter guide for successful (U)HPLC method transfer, 2018.
- United States Pharmacopeial Convention. USP40-NF35 S2, Acetaminophen method, 2017.
- USP40-NF35 S2, General chapter <621> Chromatography, 2017.
- Manka A, Franz H. Thermo Scientific Technical Note 165: Boosting Trace Detection Performance with the Vanquish DAD and LightPipe, 2016.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Strategies for the Transfer of Liquid Chromatographic Methods Between Different Instruments
2019|Thermo Fisher Scientific|Posters
Strategies for the Transfer of Liquid Chromatographic Methods Between Different Instruments Theresa Riley, Maria Grübner, Carsten Paul, Mauro De Pra, Frank Steiner, Thermo Fisher Scientific, Dornierstrasse 4, 82110 Germering, Germany UltiMate 3000 SD Quaternary Vanquish Flex Quaternary Quaternary pump F…
Key words
gdv, gdvstrategies, strategiesquaternary, quaternarycolumn, columnvanquish, vanquishdiode, diodemetering, meteringthermostatting, thermostattingtransfer, transferarray, arraycompartment, compartmentvolume, volumestrong, strongsolvent, solventvolumes
Method transfer of a USP-derived acetaminophen assay from an UltiMate 3000 SD system to a Vanquish Flex UHPLC system
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 72716 Method transfer of a USP-derived acetaminophen assay from an UltiMate 3000 SD system to a Vanquish Flex UHPLC system Authors Maria Grübner, Carsten Paul, Frank Steiner Thermo Fisher Scientific, Germering, Germany Application benefits • Flexible system volume…
Key words
vanquish, vanquishflex, flexpre, preacetaminophen, acetaminophentransfer, transferheating, heatingheater, heatermetering, meteringthermostatting, thermostattingimpurity, impurityscientific, scientificswitchable, switchablemau, mauminutes, minutesautosampler
Successful and Stress-Free LC Method Transfers
2020|Thermo Fisher Scientific|Guides
Table of contents Introduction White Paper 72711 An instrument parameter guide for successful (U)HPLC method transfer Technical Note 73371 Physical adjustment of gradient delay volume as a tool for successful transfer of HPLC methods Application Note 73186 Improving peak results…
Key words
vanquish, vanquishtransfer, transferchlorhexidine, chlorhexidinehplc, hplcimpurity, impuritycore, coresystem, systemmethod, methoduhplc, uhplccompendial, compendialepilogue, epiloguetemozolomide, temozolomidesuccessful, successfulcustom, custommebendazole
An instrument parameter guide for successful (U)HPLC method transfer
2018|Thermo Fisher Scientific|Guides
WHITE PAPER 72711 An instrument parameter guide for successful (U)HPLC method transfer Authors Carsten Paul, Maria Grübner, Michael Heidorn, Matthias Krajewski, Sabrina Patzelt, Thomas Piecha, and Frank Steiner Thermo Fisher Scientific, Germering, Germany Keywords UHPLC, HPLC, USP <621>, gradient delay…
Key words
pre, pregdv, gdvheater, heaterthermostatting, thermostattingvolume, volumecolumn, columntemperature, temperatureheating, heatingvanquish, vanquishair, airgradient, gradientstill, stillflex, flexheaters, heaterstransfer