LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Setup and transfer of a gradient HILIC method for the impurity analysis of temozolomide

Applications | 2020 | Thermo Fisher ScientificInstrumentation
HPLC
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies, Thermo Fisher Scientific

Summary

Importance of the Topic


This application addresses the rapid and reliable analysis of impurities in temozolomide, a polar oncology drug. Traditional reversed-phase methods often struggle to retain such hydrophilic compounds, leading to lengthy run times, extensive column equilibration, and reliance on ion-pairing reagents. Implementing a gradient HILIC approach streamlines impurity profiling, reducing analysis time and solvent consumption while meeting pharmacopeial requirements.

Objectives and Study Overview


The primary goal was to develop a fast gradient HILIC method for temozolomide impurity analysis on an Agilent 1260 Infinity LC system and demonstrate its straightforward transfer to a Thermo Scientific Vanquish Core HPLC system. Key evaluation metrics included retention time reproducibility, peak resolution, and method robustness across instruments.

Methodology and Instrumentation


A Syncronis HILIC column (4.6 × 100 mm, 3 µm, 175 Å) was employed with mobile phases of 20 mM ammonium acetate at pH 5.1 in water (A) and in 10:90 water:acetonitrile (B). A gradient from 0 to 44% A over 5 min followed by re-equilibration yielded a total run time of 12 min. Column temperature was set to 30 °C, and autosampler temperature at 10 °C. Detection was at 266 nm with response times adjusted for each detector.

Used Instrumentation


  • Agilent 1260 Infinity LC with quaternary pump G1311B, autosampler G1367E, column oven G1316A, VWD G1314F.
  • Thermo Scientific Vanquish Core Quaternary HPLC with Pump C, Split Sampler CT, Column Compartment C, Variable Wavelength Detector C.

Results and Discussion


The HILIC method delivered baseline separation of temozolomide and its four related impurities (A, B, D, E) with only three to four column volumes of equilibration. Forced degradation samples confirmed impurity transformations. Upon transfer, both systems produced equivalent chromatograms: retention time deviations were under 0.1 min, with RSD < 0.08% for retention and < 0.4% for peak area. The Vanquish Core system exhibited slightly narrower peaks and improved resolution due to lower extra-column volume.

Benefits and Practical Applications


  • Shortened analysis time (12 min vs. ~30 min in reversed phase).
  • Minimal column equilibration under HILIC conditions.
  • Smooth transfer between legacy and modern LC platforms without method revalidation.
  • Reliable impurity identification supporting compliance with Ph. Eur. and USP guidelines.

Future Trends and Opportunities


Emerging HILIC research focuses on further reducing equilibration demands through dynamic column preconditioning, integrating high-throughput sample preparation, and exploring novel zwitterionic stationary phases for enhanced selectivity. Additionally, automated volumetric matching tools and cloud-based CDS protocols will simplify method transfers across global laboratories.

Conclusion


A fast gradient HILIC method for temozolomide impurity analysis was successfully established and transferred between two HPLC platforms. Equivalent chromatographic performance, combined with reduced run and equilibration times, illustrates the viability of HILIC for polar drug analysis and highlights the Vanquish Core system’s suitability for robust method implementation.

References


  • European Pharmacopoeia Monograph 2780, Temozolomide.
  • USP43–NF38, Temozolomide Monograph.
  • Paul, C.; Grübner, M. Thermo Scientific WP-72711, 2018.
  • McCalley, D.V. J. Chromatogr. A, 2018, 1554, 61–70.
  • Müllner, T.; Franz, H. Thermo Scientific TN-73371, 2020.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Successful and Stress-Free LC Method Transfers
Successful and Stress-Free LC Method Transfers
2020|Thermo Fisher Scientific|Guides
Table of contents Introduction White Paper 72711 An instrument parameter guide for successful (U)HPLC method transfer Technical Note 73371 Physical adjustment of gradient delay volume as a tool for successful transfer of HPLC methods Application Note 73186 Improving peak results…
Key words
vanquish, vanquishtransfer, transferchlorhexidine, chlorhexidinehplc, hplcimpurity, impuritycore, coresystem, systemmethod, methoduhplc, uhplccompendial, compendialepilogue, epiloguetemozolomide, temozolomidesuccessful, successfulcustom, custommebendazole
Straightforward transfer of an EP method for impurity analysis of chlorhexidine from an Agilent 1260 Infinity LC system to a Vanquish Core HPLC system
APPLICATION NOTE 73309 Straightforward transfer of an EP method for impurity analysis of chlorhexidine from an Agilent 1260 Infinity LC system to a Vanquish Core HPLC system Author: Maria Grübner Thermo Fisher Scientific, Germering, Germany Keywords: HPLC method transfer, Vanquish…
Key words
rit, ritrity, ritycore, corevanquish, vanquishimpurity, impurityhplc, hplcchlorhexidine, chlorhexidinesystem, systemurit, urittransfer, transferdigluconate, digluconategradient, gradientmonograph, monographequivalent, equivalentmau
Straightforward transfer of an EP method for impurity analysis of chlorhexidine from an Agilent 1260 Infinity LC system to a Vanquish Core HPLC system
APPLICATION NOTE 73309 Straightforward transfer of an EP method for impurity analysis of chlorhexidine from an Agilent 1260 Infinity LC system to a Vanquish Core HPLC system Author: Maria Grübner Thermo Fisher Scientific, Germering, Germany Keywords: HPLC method transfer, Vanquish…
Key words
rit, ritrity, ritycore, corevanquish, vanquishimpurity, impurityhplc, hplcchlorhexidine, chlorhexidinesystem, systemurit, urittransfer, transferdigluconate, digluconategradient, gradientmonograph, monographequivalent, equivalentdeviations
An instrument parameter guide for successful (U)HPLC method transfer
WHITE PAPER 72711 An instrument parameter guide for successful (U)HPLC method transfer Authors Carsten Paul, Maria Grübner, Michael Heidorn, Matthias Krajewski, Sabrina Patzelt, Thomas Piecha, and Frank Steiner Thermo Fisher Scientific, Germering, Germany Keywords UHPLC, HPLC, USP <621>, gradient delay…
Key words
pre, pregdv, gdvheater, heaterthermostatting, thermostattingvolume, volumecolumn, columntemperature, temperatureheating, heatingvanquish, vanquishgradient, gradientair, airstill, stillflex, flexheaters, heaterstransfer
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike