Tune Your Mixing Volume for Gradient Generation to Achieve Better Detection Limits
Technical notes | 2016 | Thermo Fisher ScientificInstrumentation
Gradient elution is a cornerstone technique in high-performance liquid chromatography (HPLC), enabling the separation of complex mixtures by varying solvent composition over time. Precise mixing of solvents is essential to achieve stable baselines and optimal detection limits, particularly in UV absorption measurements where minor fluctuations appear as baseline ripples and can compromise the limit of detection (LOD) and limit of quantification (LOQ). Understanding how mixing strategies influence these fluctuations is critical for laboratories focusing on trace analysis, quality control and advanced research applications.
This technical note compares two gradient formation technologies—high-pressure gradient (HPG) proportioning and low-pressure gradient (LPG) proportioning—and investigates the mechanisms that cause solvent composition fluctuations. It further provides practical guidance on selecting an appropriate mixer volume (SpinFlow mixer) to minimize baseline ripple and improve detection performance in HPLC systems.
The study analyzes two solvent delivery configurations:
Theoretical flow diagrams illustrate how pressure pulsations in HPG and discrete solvent plugs in LPG generate compositional fluctuations. UV detection traces reveal baseline ripples when solvents with differing extinction coefficients are not fully homogenized. The investigation evaluates SpinFlow mixers of varying internal volumes (35 µL to 1550 µL) integrated into Thermo Scientific™ Dionex™ UltiMate™ 3000 RS/SD systems to assess ripple reduction.
Both HPG and LPG configurations exhibit solvent composition instabilities. In HPG, slight pump pulsations induce sinusoidal fluctuations; in LPG, sequential plugs require effective longitudinal mixing. Incomplete mixing leads to baseline ripples visible in UV chromatograms, degrading LOD. Larger mixing volumes increase radial and longitudinal homogenization, smoothing solvent composition transitions and eliminating ripples.
Optimizing mixer volume yields several advantages:
Advances in microfluidic mixing technology and real-time feedback control promise even finer gradient precision. Integration of smart pump algorithms and mixer designs could adapt mixing volumes dynamically to method requirements. Emerging applications include ultrafast proteomics, single-cell analysis and on-line coupling with advanced detectors such as ambient ionization mass spectrometry.
Selecting the appropriate mixing volume in gradient-forming HPLC systems is essential to minimize baseline ripple and maximize detection performance. The choice between HPG and LPG proportioning, combined with an optimized SpinFlow mixer size, enables analysts to balance speed, sensitivity and robustness for a broad range of applications.
Schmidt C., Jendreizik V., Niedner W., McLeod F. Impact of Solvent Mixing on Liquid Chromatography Performance. Thermo Fisher Scientific Poster Note LPN 2880-01, Germering, Germany, 2011.
HPLC
IndustriesManufacturerThermo Fisher Scientific
Summary
Significance of the topic
Gradient elution is a cornerstone technique in high-performance liquid chromatography (HPLC), enabling the separation of complex mixtures by varying solvent composition over time. Precise mixing of solvents is essential to achieve stable baselines and optimal detection limits, particularly in UV absorption measurements where minor fluctuations appear as baseline ripples and can compromise the limit of detection (LOD) and limit of quantification (LOQ). Understanding how mixing strategies influence these fluctuations is critical for laboratories focusing on trace analysis, quality control and advanced research applications.
Objectives and article overview
This technical note compares two gradient formation technologies—high-pressure gradient (HPG) proportioning and low-pressure gradient (LPG) proportioning—and investigates the mechanisms that cause solvent composition fluctuations. It further provides practical guidance on selecting an appropriate mixer volume (SpinFlow mixer) to minimize baseline ripple and improve detection performance in HPLC systems.
Methodology and used instrumentation
The study analyzes two solvent delivery configurations:
- HPG proportioning: independent high-pressure pumps deliver solvents that merge post-pump via a tee-piece, followed by mixing.
- LPG proportioning: solenoid valves on the low-pressure side channel solvent plugs into the pump inlet, with mixing occurring downstream.
Theoretical flow diagrams illustrate how pressure pulsations in HPG and discrete solvent plugs in LPG generate compositional fluctuations. UV detection traces reveal baseline ripples when solvents with differing extinction coefficients are not fully homogenized. The investigation evaluates SpinFlow mixers of varying internal volumes (35 µL to 1550 µL) integrated into Thermo Scientific™ Dionex™ UltiMate™ 3000 RS/SD systems to assess ripple reduction.
Main results and discussion
Both HPG and LPG configurations exhibit solvent composition instabilities. In HPG, slight pump pulsations induce sinusoidal fluctuations; in LPG, sequential plugs require effective longitudinal mixing. Incomplete mixing leads to baseline ripples visible in UV chromatograms, degrading LOD. Larger mixing volumes increase radial and longitudinal homogenization, smoothing solvent composition transitions and eliminating ripples.
Benefits and practical applications
Optimizing mixer volume yields several advantages:
- Lower baseline noise and ripple, enhancing UV detection sensitivity.
- Improved LOD and LOQ for trace analyte quantification.
- Tailored solutions for fast separations (smaller mixers) and high-sensitivity methods (larger mixers).
- Compatibility with mass spectrometry and charged aerosol detection when mixing artifacts are less critical.
Future trends and potential applications
Advances in microfluidic mixing technology and real-time feedback control promise even finer gradient precision. Integration of smart pump algorithms and mixer designs could adapt mixing volumes dynamically to method requirements. Emerging applications include ultrafast proteomics, single-cell analysis and on-line coupling with advanced detectors such as ambient ionization mass spectrometry.
Conclusion
Selecting the appropriate mixing volume in gradient-forming HPLC systems is essential to minimize baseline ripple and maximize detection performance. The choice between HPG and LPG proportioning, combined with an optimized SpinFlow mixer size, enables analysts to balance speed, sensitivity and robustness for a broad range of applications.
Reference
Schmidt C., Jendreizik V., Niedner W., McLeod F. Impact of Solvent Mixing on Liquid Chromatography Performance. Thermo Fisher Scientific Poster Note LPN 2880-01, Germering, Germany, 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Thermo Scientific Dionex UltiMate 3000 Pumps - Product Specifications
2016|Thermo Fisher Scientific|Brochures and specifications
C H R O M ATO G R A P H Y The Benchmark in LC Solvent Delivery Produ c t Spe ci f i cat i ons Thermo Scientific Dionex UltiMate 3000 Pumps Thermo Scientific™ Dionex™ UltiMate™ 3000 products…
Key words
proportioning, proportioningxrs, xrspumps, pumpspump, pumpbiocompatible, biocompatiblecylinder, cylinderbio, biopurge, purgepressure, pressurevalve, valveflow, flowunrivaled, unrivaledmixer, mixeroptional, optionalhpg
Thermo Scientific Dionex UltiMate 3000 Series - SD, RS, BM, and BX Pumps
2013|Thermo Fisher Scientific|Manuals
Thermo Scientific Dionex UltiMate 3000 Series SD, RS, BM, and BX Pumps Operating Instructions (Original Operating Instructions) Revision: 1.7 Date: September 2013 © 2013 Thermo Fisher Scientific Doc. No. 4820.4001 UltiMate 3000 Series: SD, RS, BM, and BX Pumps Operating…
Key words
instructions, instructionsoperating, operatingpump, pumpseal, sealwash, washpiston, pistonhead, headcylinder, cylinderincluded, includedpressure, pressurepurge, purgerear, rearchromeleon, chromeleonvalve, valvepumps
Tips and Tricks for HPLC and UHPLC
2016|Thermo Fisher Scientific|Presentations
Tips and Tricks for HPLC and UHPLC Jan Pettersson Nordics Sales Support Specialist Chromatography Thermo Fisher Scientific, Hägersten/Sweden The world leader in serving science Overview Thermo Scientific™ Vanquish™ UHPLC system 2 Thermo Scientific™ UltiMate™ 3000 UHPLC system Common Recommendations Mobile…
Key words
pump, pumpballistic, ballisticeluent, eluentcell, cellslit, slitbandwidth, bandwidthflow, flowautosamplers, autosamplersrecommendations, recommendationspeak, peakeluents, eluentsdiode, diodebuffers, buffersgradients, gradientspre
Intelligent UHPLC pump stroke control - Ensures sensitive peptide analysis by ripple-free baseline in TFA applications
2020|Thermo Fisher Scientific|Technical notes
TECHNICAL NOTE 73582 Intelligent UHPLC pump stroke control Ensures sensitive peptide analysis by ripple-free baseline in TFA applications Authors: Giorgia Greco, Dennis Koehler, Christoph Strobl, Markus M. Martin Thermo Fisher Scientific, Germering, Germany Keywords: Baseline ripples, stroke volume, UHPLC, SmartStroke,…
Key words
smartstroke, smartstroketfa, tfavanquish, vanquishhorizon, horizonstroke, strokeripples, ripplesmixer, mixerbaseline, baselineacn, acnvolume, volumebinary, binaryripple, ripplemobile, mobilepump, pumpmau