Technical notes | 2020 | Thermo Fisher ScientificInstrumentation
LC/MS, LC/SQ
Industries
Manufacturer
Thermo Fisher Scientific

Summary

Importance of the Topic


A single quadrupole mass detector enhances conventional liquid chromatography by adding molecular mass information, higher sensitivity and selectivity, and the ability to detect compounds lacking UV chromophores. This capability addresses challenges in complex sample matrices and broadens the range of analytes accessible to routine analysis.

Objectives and Overview


This technical note reviews the fundamental principles, benefits, and practical implementation of single quadrupole mass detection in liquid chromatography. It focuses on Thermo Scientific ISQ EC and ISQ EM instruments, describes ionization modes, examines compatibility considerations, and highlights the impact on analytical workflows.

Methodology and Instrumentation


Liquid chromatography is coupled to a single quadrupole mass detector comprising four main components
  • Ion source with heated electrospray ionization or atmospheric pressure chemical ionization for gas phase ion generation
  • Ion guide system for neutral removal and ion focusing
  • Quadrupole mass filter separating ions by mass to charge ratio
  • Ion detector using a dynode and electron multiplier to amplify ion signals
These detectors integrate seamlessly into the Chromeleon chromatography data system. Automated method setup via AutoSpray simplifies parameter selection based on flow rate, solvent composition and analyte properties. Two detector models cover mass ranges from 10–1250 and 10–2000 daltons, and support flow rates from 50 microliters to 2 milliliters per minute without splitting.

Main Results and Discussion


Key performance highlights
  • Ionization modes HESI and APCI offer soft ionization to preserve molecular ions with minimal fragmentation
  • Mass resolution of approximately one atomic mass unit enables monitoring of specific ions or full scan acquisition for unknown screening
  • Automated tuning and calibrant delivery reduce downtime and maintenance to under one hour
  • Mobile phase compatibility requires volatile additives such as formic acid or ammonium formate to avoid source fouling
  • Ion suppression effects can be managed by sample cleanup, chromatographic separation, choice of ionization mode and use of internal standards
Adduct formation and isotopic patterns provide additional selectivity but may complicate spectra. Collision induced dissociation within the ion guide can reduce adduct interference.

Benefits and Practical Applications


Adding single quadrupole MS detection to an LC system delivers
  • Unambiguous analyte identification based on accurate mass and isotopic distribution
  • Improved quantitation of coeluting compounds by monitoring distinct m/z values
  • Enhanced sensitivity compared to UV detection, lowering detection limits and sample requirements
  • Detection of nonchromophoric and poorly absorbing molecules
  • Reduced sample preparation and faster method development for routine analyses in environmental, pharmaceutical, academic and industrial laboratories

Future Trends and Possibilities for Use


Emerging directions include
  • Integration of advanced data processing and machine learning for automated spectrum interpretation
  • Development of hybrid detectors combining quadrupole with time of flight or orbitrap for higher resolution
  • Expansion of ambient and direct sampling interfaces to reduce or eliminate chromatography for rapid screening
  • Improved ion source designs for challenging matrices and nonvolatile mobile phases
  • Wider adoption of mass detection in quality control and regulated workflows

Conclusion


Single quadrupole mass detection represents a cost-effective and robust upgrade to conventional UV-based LC systems. It delivers molecular specificity, improved sensitivity and simplified workflows. With minimal maintenance and user-friendly software integration, these detectors are well suited for routine analysis across a wide range of applications.

Reference


1. Dan Kalu Appulage and Stephan Meding. Thermo Fisher Scientific Technical Note 73050 Quick and Easy Tuning and Troubleshooting of ISQ EC and ISQ EM Mass Spectrometers
2. Konermann L, Ahadi E, Rodriguez AD, Vahidi S. Unraveling the Mechanism of Electrospray Ionization. Analytical Chemistry 2013, 85, 2–9
3. Dass C. Fundamentals of Contemporary Mass Spectrometry. John Wiley & Sons, 2007
4. Carroll DI, Dzidic I, Horning EC, Stillwell RN. Atmospheric Pressure Ionization Mass Spectrometry. Applied Spectroscopy Reviews 1981, 17, 337–406
5. Annesley TM. Ion Suppression in Mass Spectrometry. Clinical Chemistry 2003, 49, 1041–1044

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Thermo Scientific ISQ EC and ISQ EM single quadrupole mass spectrometers
Thermo Scientific ISQ EC and ISQ EM single quadrupole mass spectrometers
2018|Thermo Fisher Scientific|Brochures and specifications
Join the mass movement towards mass spectrometry Thermo Scientific ISQ EC and ISQ EM single quadrupole mass spectrometers Embrace the power of mass spectrometry Achieving a comprehensive understanding of the samples you analyze is always a challenge. At Thermo Fisher…
Key words
isq, isqmass, masshesi, hesiapci, apcispectrometers, spectrometersthermo, thermoscientific, scientificspectrometer, spectrometersim, simtenofovir, tenofoviryour, yourchromeleon, chromeleonimpurity, impurityroutine, routinesensitivity
A multi-detector platform comprising UV/Vis, charged aerosol, and single quadrupole mass spectrometric detection for comprehensive sample analysis
APPLICATION NOTE 72869 A multi-detector platform comprising UV/Vis, charged aerosol, and single quadrupole mass spectrometric detection for comprehensive sample analysis Authors Stephan Meding, Katherine Lovejoy, Remco Swart, Frank Steiner, and Martin Ruehl Thermo Fisher Scientific, Germering, Germany Keywords Vanquish UHPLC,…
Key words
cad, cadhesi, hesiapci, apciinverse, inverseerucamide, erucamidevanquish, vanquishirg, irgmethanol, methanoldiverter, diverteraerosol, aerosolpiece, pieceisq, isqsplitter, splitterthermo, thermostearic
A multi-detector platform comprising UV/Vis, charged aerosol, and single quadrupole mass spectrometric detection for comprehensive sample analysis
APPLICATION NOTE 72869 A multi-detector platform comprising UV/Vis, charged aerosol, and single quadrupole mass spectrometric detection for comprehensive sample analysis Authors Stephan Meding, Katherine Lovejoy, Remco Swart, Frank Steiner, and Martin Ruehl Thermo Fisher Scientific, Germering, Germany Keywords Vanquish UHPLC,…
Key words
cad, cadhesi, hesiapci, apciinverse, inverseerucamide, erucamidemethanol, methanolvanquish, vanquishirg, irgdiverter, diverteraerosol, aerosolpiece, pieceisq, isqsplitter, splitterthermo, thermostearic
Seeking More Productive Chromatography? See Productivity in a New Light
Seeking More Productive Chromatography? See Productivity in a New Light Joachim Weiss, Technical Director, Thermo Fisher Scientific, Dreieich, Germany Timothy Cross, Sr. Regional Marketing Manager – HPLC, Thermo Fisher Scientific, Hemel Hempstead, UK The world leader in serving science Seminar…
Key words
isq, isqvanquish, vanquishduo, duotenofovir, tenofovirchromeleon, chromeleoninverse, inversedual, dualcds, cdsthermo, thermosystem, systemmass, massgradient, gradientproductivity, productivityscientific, scientificsource
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike