Increasing Consistency Across Biopharmaceutical Labs with the ACQUITY Premier LC Platform

Technical notes | 2021 | WatersInstrumentation
HPLC
Industries
Pharma & Biopharma
Manufacturer
Waters

Summary

Significance of the topic


This study addresses the critical need for reproducible analytical results across biopharmaceutical laboratories when migrating chromatographic assays. Consistent data continuity and regulatory compliance depend on minimizing variability in retention time, peak area, and glycan identification. The ACQUITY Premier LC platform with MaxPeak High Performance Surfaces offers improved handling of metal-sensitive analytes to support reliable cross-lab comparisons.

Objectives and study overview


The work evaluated whether two configurations of the ACQUITY Premier LC System could replicate a legacy HILIC-based glycan profiling method. A released glycan performance standard was analyzed on both a BioAccord system with waters_connect and an Empower 3 system with QDa detection. Key metrics included chromatographic profiles, relative peak areas, retention times, and glucose unit (GU) values.

Methodology


  • Standard: RapiFluor-MS Glycan Performance Test.
  • Column: ACQUITY Premier Glycan BEH Amide, 1.7 µm, 130 Å, 2.1 × 150 mm.
  • Gradient: 35 min at 0.6% B/min under HILIC conditions.
  • Detection: fluorescence and orthogonal mass spectrometry (QDa).
  • Evaluation: relative peak area percentages, retention time shifts, GU comparisons.

Instrumentation used


  • Two ACQUITY Premier LC systems.
  • BioAccord LC-MS with waters_connect informatics.
  • Empower 3 CDS paired with ACQUITY QDa Mass Detector.

Main results and discussion


Chromatograms from both platforms showed highly comparable peak intensities and glycan species. Relative peak area differences were below 0.5%, and retention time offsets averaged 0.08 min. GU values calculated via a dextran ladder aligned within 0.1 units of legacy library entries. Orthogonal QDa data confirmed the identity of the FA2G2 glycan at m/z 1051, corresponding to GU 7.44.

Benefits and practical applications


  • Reliable method transfer between development and supporting labs.
  • Seamless compatibility with existing GU libraries and scientific databases.
  • Reduced analyte–surface interactions for metal-sensitive species.
  • Consistent monitoring of critical quality attributes.

Future trends and potential applications


Future developments may extend MaxPeak HPS technology to a broader range of analytes and instrument platforms. Integration with automated QC workflows and advanced data analytics could further streamline cross-platform standardization. Enhanced informatics connectivity may enable real-time normalization and global comparability of analytical methods.

Conclusion


The ACQUITY Premier LC System delivers robust, cross-instrument reproducibility for HILIC glycan assays while matching legacy performance. Its consistent retention, peak area, and GU results support rigorous biopharmaceutical development and manufacturing workflows.

References


  • Delano et al. Using Hybrid Organic-Inorganic Surface Technology to Mitigate Analyte Interactions with Metal Surfaces in UHPLC. Anal Chem. 2021;93(14):5773–5781.
  • Birdsall et al. Reducing Metal-Ion Mediated Adsorption of Acidic Peptides in RPLC-Based Assays using Hybrid Silica Chromatographic Surfaces. J Chromatogr B. 2021;1179.
  • Liu X, Lauber M. MaxPeak High Performance Surfaces Technology Improves HILIC Profiling of Released N-Glycans. Waters Application Note. 2021.
  • Higel F et al. N-glycans of Complex Glycosylated Biopharmaceuticals and Their Impact on Protein Clearance. Eur J Pharm Biopharm. 2019;139:123–131.
  • Yu Y. Applying RapiFluor-MS and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis. Waters Application Note. 2015.
  • Zhang X et al. Increasing Productivity and Confidence for N-linked Glycan Analysis of Biosimilars Using the BioAccord System. Waters Application Note. 2019.
  • Zhang X, Kellett J, Birdsall R, Yu Y. Improving Released N-Glycan Analysis Using the ACQUITY Premier Solution with MaxPeak HPS Technology. Waters Application Note. 2021.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Improving Released N-glycan Analysis in  Biotherapeutic Development Using the  ACQUITY Premier Solution with MaxPeak  High Performance Surfaces (HPS) Technology
Application Note Improving Released N-glycan Analysis in Biotherapeutic Development Using the ACQUITY Premier Solution with MaxPeak High Performance Surfaces (HPS) Technology Ximo Zhang, Jacob Kellett, Robert E. Birdsall, Ying Qing Yu Waters Corporation Abstract Characterization and monitoring of glycosylation are…
Key words
glycan, glycanglycans, glycansacquity, acquitybioaccord, bioaccordpremier, premierreleased, releasedunifi, unifiamide, amidebeh, behflr, flrhps, hpsmaxpeak, maxpeaklysosomal, lysosomalsystem, systemmetal
Migration of an HILIC Method for Released  Glycan Analysis from the ACQUITY UPLC H- Class PLUS Bio System to the Arc Premier  System
Application Note Migration of an HILIC Method for Released Glycan Analysis from the ACQUITY UPLC HClass PLUS Bio System to the Arc Premier System Robert E. Birdsall, Jacob Kellett, Ying Qing Yu Waters Corporation This is an Application Brief and…
Key words
arc, arcpremier, premiermigration, migrationbio, bioacquity, acquityplus, plussystem, systemuplc, uplchilic, hilicglycan, glycanmigrated, migratedreleased, releasedclass, classmethod, methodfrom
Improving Chromatographic Separations Of Biopharmaceuticals With MaxPeak High Performance Surfaces (HPS) Technology
Improving Chromatographic Separations Of Biopharmaceuticals With MaxPeak High Performance Surfaces (HPS) Technology Analytical Solutions Using MaxPeak High Performance Surfaces (HPS) Technology for Biopharmaceutical Therapeutics Biopharmaceutical therapeutics have consistently been the fastest growing segment within the pharmaceutical space and require advances…
Key words
premier, premieracquity, acquityglycan, glycanmaxpeak, maxpeaknucleotide, nucleotideculture, culturehps, hpspeptide, peptideintact, intactsystem, systembioaccord, bioaccordcolumn, columnread, readprotein, proteinglycans
Evaluating the Waters ACQUITY Premier System as a Flexible LC Platform That Can Be Broadly Deployed in Biopharmaceutical Labs
Application Note Evaluating the Waters ACQUITY Premier System as a Flexible LC Platform That Can Be Broadly Deployed in Biopharmaceutical Labs Jacob Kellett, Brooke M. Koshel, Hua Yang, Robert E. Birdsall, Stephan M. Koza, Ying Qing Yu Waters Corporation This…