LC-MS/MS Method for the Determination of Testosterone using an Accucore C8 HPLC Column
Applications | 2012 | Thermo Fisher ScientificInstrumentation
Monitoring testosterone levels plays a crucial role in assessing hormone replacement therapy efficacy and detecting illicit anabolic steroid use in sports. Rapid and reliable quantification supports clinical decision-making and ensures compliance with anti-doping regulations.
This study demonstrates a fast, sensitive LC-MS/MS method for testosterone determination using a Thermo Scientific Accucore C8 column. The goal was to achieve sub-2-minute cycle times while preserving peak shape and quantitative precision.
Chromatographic separation was performed on a 2.6 µm, 50 × 2.1 mm Accucore C8 column with a water/0.1% formic acid to acetonitrile/0.1% formic acid gradient ramping from 5% to 95% organic in 0.8 minutes at 1.5 mL/min and 60 °C. A 5 µL injection volume and dual wash solvents minimized carryover.
Testosterone eluted at 0.73 minutes with excellent symmetry and minimal tailing. The method delivered a total run time of 1.5 minutes. Six replicate injections of a 20 ng/mL standard yielded retention time RSD of 0.22% and peak area RSD of 3.01%, demonstrating high precision.
Key advantages include high throughput, low backpressure compared to sub-2 µm phases, robust peak shape, and reproducible quantitation. The method suits clinical laboratories monitoring therapeutic levels and anti-doping analyses requiring rapid turnaround.
Advances may involve integrating high-resolution mass spectrometry for improved selectivity, automating sample preparation workflows, and extending the approach to multiplex steroid panels for comprehensive endocrine profiling.
The optimized LC-MS/MS protocol using an Accucore C8 column provides a fast, reliable tool for testosterone analysis, combining speed with analytical performance suitable for diverse regulated environments.
Consumables, LC/MS, LC/MS/MS, LC columns, LC/QQQ
IndustriesClinical Research
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
Monitoring testosterone levels plays a crucial role in assessing hormone replacement therapy efficacy and detecting illicit anabolic steroid use in sports. Rapid and reliable quantification supports clinical decision-making and ensures compliance with anti-doping regulations.
Aims and Study Overview
This study demonstrates a fast, sensitive LC-MS/MS method for testosterone determination using a Thermo Scientific Accucore C8 column. The goal was to achieve sub-2-minute cycle times while preserving peak shape and quantitative precision.
Methodology
Chromatographic separation was performed on a 2.6 µm, 50 × 2.1 mm Accucore C8 column with a water/0.1% formic acid to acetonitrile/0.1% formic acid gradient ramping from 5% to 95% organic in 0.8 minutes at 1.5 mL/min and 60 °C. A 5 µL injection volume and dual wash solvents minimized carryover.
Used Instrumentation
- HPLC System: Thermo Scientific Accela 600
- MS Detector: Thermo Scientific TSQ Vantage with HESI source
- Core Enhanced Technology column: Accucore C8, 2.6 µm
- Data Processing: Thermo Scientific LC QUAN software
Main Results and Discussion
Testosterone eluted at 0.73 minutes with excellent symmetry and minimal tailing. The method delivered a total run time of 1.5 minutes. Six replicate injections of a 20 ng/mL standard yielded retention time RSD of 0.22% and peak area RSD of 3.01%, demonstrating high precision.
Benefits and Practical Applications
Key advantages include high throughput, low backpressure compared to sub-2 µm phases, robust peak shape, and reproducible quantitation. The method suits clinical laboratories monitoring therapeutic levels and anti-doping analyses requiring rapid turnaround.
Future Trends and Opportunities
Advances may involve integrating high-resolution mass spectrometry for improved selectivity, automating sample preparation workflows, and extending the approach to multiplex steroid panels for comprehensive endocrine profiling.
Conclusion
The optimized LC-MS/MS protocol using an Accucore C8 column provides a fast, reliable tool for testosterone analysis, combining speed with analytical performance suitable for diverse regulated environments.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Separation of Metformin and Voglibose Using a Thermo Scientific Accucore HILIC HPLC Column
2012|Thermo Fisher Scientific|Applications
Krishna Rao Dara, Prakash Chander, Dr. Tushar N Mehta, Centre of Excellence for Asia Pacific Laboratory, Thermo Fisher Scientific, Ahmedabad, India Appli c at i on N ot e 2 0 5 2 5 Separation of Metformin and Voglibose Using…
Key words
voglibose, voglibosemetformin, metforminaccucore, accucorehilic, hilicakums, akumsarb, arbseparation, separationoptima, optimaoral, oralscientific, scientificpressure, pressurehyperglycemic, hyperglycemicpharma, pharmaanti, antisrm
SOLAμ for pre-analysis sample concentration
2014|Thermo Fisher Scientific|Technical notes
Jon Bardsley, Ken Meadows, Thermo Fisher Scientific, Runcorn, UK Key Words SOLAµ, micro elution, reproducibility, matrix effects, SPE, no dry down, niflumic acid, sensitivity, sample concentration Abstract This application note demonstrates the use of Thermo Scientific™ SOLAμ™ SPE product to…
Key words
scientific, scientificelution, elutionthermo, thermoniflumic, niflumicvolumes, volumeslengthy, lengthysample, sampleaccucore, accucorelow, lowsolaμ, solaμsolaµ, solaµvantage, vantagesensitivity, sensitivityarb, arbscale
Zafirlukast in Human Plasma Using Retain AX and Accucore RP-MS Column
2012|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETRETZAF Zafirlukast in Human Plasma Using Retain AX and Accucore RP-MS Column Jamil Ali, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Asthma drug • Zafirlukast • Accolate • Retain AX • Bioanalysis • Glybenclamide A…
Key words
zafirlukast, zafirlukastglybenclamide, glybenclamidefisher, fisherscientific, scientificretain, retainaccucore, accucoreaccolate, accolatethermo, thermohypersep, hypersepgrade, gradeleukotriene, leukotrieneplasma, plasmapart, partnorth, northhplc
LC-MS/MS Method for the Determination of Diclofenac in Human Plasma
2012|Thermo Fisher Scientific|Applications
J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Appli c at i on N ot e 2 0 5 6 9 LC-MS/MS Method for the Determination of Diclofenac in Human Plasma Key Words SPE, SOLA, Accucore RP-MS, diclofenac, Core Enhanced…
Key words
diclofenac, diclofenacscientific, scientificfisher, fisherplasma, plasmahuman, humanaccucore, accucoreqcl, qclgrade, gradelcms, lcmssola, solaaverage, averagediff, diffthermo, thermopart, partoxygenase