Simultaneous Analysis of DNPH-Derivatized Aldehydes
Applications | 2021 | ShimadzuInstrumentation
Importance of topic
Aldehydes such as acetaldehyde, propionaldehyde and valeraldehyde contribute to air pollution, unpleasant odors and potential health risks. Accurate quantification of these volatile carbonyls at low concentrations is vital for environmental monitoring, indoor air quality assessment and industrial hygiene. Derivatization with 2,4-dinitrophenylhydrazine (DNPH) transforms reactive aldehydes into stable hydrazones, enabling sensitive reversed-phase liquid chromatography with UV detection.
This application note describes the development of a robust HPLC method for simultaneous separation and quantification of six DNPH-derivatized aldehydes using Shimadzu’s Shim-pack Scepter PFPP-120 column. The goals were to achieve baseline resolution, high sensitivity at the μg/mL level, and reproducible retention times under a gradient elution program compatible with routine environmental analysis.
The derivatized standards (1.0 μg/mL each) were analyzed on a Shim-pack Scepter PFPP-120 column (150 mm × 4.6 mm, 3 μm) installed on a Prominence-I Plus system. The mobile phase consisted of:
A linear gradient from 20% B at 0 min to 55% at 5 min, ramped to 60% at 25 min, held to 35 min, and returned to 20% by 40 min at 1.0 mL/min. Column temperature was maintained at 35 °C, injection volume 20 μL, and detection at 360 nm using a deuterium lamp.
The PFPP stationary phase exhibited strong π-π interactions and shape selectivity, enabling clear separation of isomeric aldehydes (iso-butyraldehyde vs. n-butyraldehyde, iso-valeraldehyde vs. n-valeraldehyde). All six hydrazones eluted within 35 min with retention time RSDs under 1%. UV detection at 360 nm provided limits of detection in the low μg/mL range. The gradient profile minimized baseline drift and ensured reproducible peak shapes even with complex sample matrices.
Advancements may include coupling PFPP columns with tandem mass spectrometry to lower detection limits to ng/mL levels and to broaden the analyte scope to ketones and other carbonyl species. Miniaturized UHPLC formats could further reduce analysis time and solvent use, while on-line sampling integrations may enable real-time air monitoring in industrial and urban settings.
The Shim-pack Scepter PFPP-120 column combined with DNPH derivatization and UV detection offers a robust, selective and reproducible method for quantifying six common aldehydes. This approach meets the demands of routine environmental analysis, providing clear resolution of structural isomers and reliable sensitivity for quality control and regulatory compliance.
Application News L533 (JP, ENG), Shimadzu Corporation, First Edition: Dec. 2021, ERAS-1000-0129
Consumables, HPLC, LC columns
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Simultaneous Analysis of DNPH-Derivatized Aldehydes Using Shim-pack Scepter PFPP-120
Importance of topic
Aldehydes such as acetaldehyde, propionaldehyde and valeraldehyde contribute to air pollution, unpleasant odors and potential health risks. Accurate quantification of these volatile carbonyls at low concentrations is vital for environmental monitoring, indoor air quality assessment and industrial hygiene. Derivatization with 2,4-dinitrophenylhydrazine (DNPH) transforms reactive aldehydes into stable hydrazones, enabling sensitive reversed-phase liquid chromatography with UV detection.
Objectives and Study Overview
This application note describes the development of a robust HPLC method for simultaneous separation and quantification of six DNPH-derivatized aldehydes using Shimadzu’s Shim-pack Scepter PFPP-120 column. The goals were to achieve baseline resolution, high sensitivity at the μg/mL level, and reproducible retention times under a gradient elution program compatible with routine environmental analysis.
Methodology and Instrumentation
The derivatized standards (1.0 μg/mL each) were analyzed on a Shim-pack Scepter PFPP-120 column (150 mm × 4.6 mm, 3 μm) installed on a Prominence-I Plus system. The mobile phase consisted of:
- Solvent A: Water
- Solvent B: Methanol/Acetonitrile (8/2, v/v)
A linear gradient from 20% B at 0 min to 55% at 5 min, ramped to 60% at 25 min, held to 35 min, and returned to 20% by 40 min at 1.0 mL/min. Column temperature was maintained at 35 °C, injection volume 20 μL, and detection at 360 nm using a deuterium lamp.
Main Results and Discussion
The PFPP stationary phase exhibited strong π-π interactions and shape selectivity, enabling clear separation of isomeric aldehydes (iso-butyraldehyde vs. n-butyraldehyde, iso-valeraldehyde vs. n-valeraldehyde). All six hydrazones eluted within 35 min with retention time RSDs under 1%. UV detection at 360 nm provided limits of detection in the low μg/mL range. The gradient profile minimized baseline drift and ensured reproducible peak shapes even with complex sample matrices.
Benefits and Practical Applications
- High selectivity for isomeric aldehydes in a single run.
- Reliable quantification for environmental and occupational monitoring.
- Compatibility with standard DNPH sampling protocols.
- Efficient gradient saves time and solvent consumption.
Future Trends and Possibilities
Advancements may include coupling PFPP columns with tandem mass spectrometry to lower detection limits to ng/mL levels and to broaden the analyte scope to ketones and other carbonyl species. Miniaturized UHPLC formats could further reduce analysis time and solvent use, while on-line sampling integrations may enable real-time air monitoring in industrial and urban settings.
Conclusion
The Shim-pack Scepter PFPP-120 column combined with DNPH derivatization and UV detection offers a robust, selective and reproducible method for quantifying six common aldehydes. This approach meets the demands of routine environmental analysis, providing clear resolution of structural isomers and reliable sensitivity for quality control and regulatory compliance.
Reference
Application News L533 (JP, ENG), Shimadzu Corporation, First Edition: Dec. 2021, ERAS-1000-0129
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Simultaneous Analysis of Six DNPH-Derivatized Aldehydes
2021|Shimadzu|Applications
ERAS-1000-0236 LC Reversed-phase Shim-packTM Series Shim-pack ScepterTM PFPP-120 Simultaneous Analysis of Six DNPHDerivatized Aldehydes 236 Keywords: precolumn derivatization 1. Acetaldehyde 2. Propionaldehyde 3. n-Butyraldehyde 4. iso-Butyraldehyde 5. iso-Valeraldehyde 6. n-Valeraldehyde Column Mobile phase Time program Flow rate Column temp. Detection…
Key words
dnphderivatized, dnphderivatizedaldehydes, aldehydespacktm, packtmshim, shimphase, phasesix, sixreversed, reversedsimultaneous, simultaneousseries, seriesprogram, programmobile, mobilerate, ratedetection, detectionflow, flowcolumn
Simultaneous Analysis of DNPH-Derivatized Aldehydes Using Prominence™-i Plus and Shim-pack Scepter™ PFPP-120
2019|Shimadzu|Applications
Application News No. High Performance Liquid Chromatography Simultaneous Analysis of DNPH-Derivatized Aldehydes Using Prominence™-i Plus and Shim-pack Scepter™ PFPP-120 L533 On September 21, 2018, the method for measurement of Specified Offensive Odor Substances in Article 5 of the Ordinance of…
Key words
valeraldehyde, valeraldehydebutyraldehyde, butyraldehydeprominence, prominenceconcentration, concentrationiso, isooffensive, offensivepropionaldehyde, propionaldehydeodor, odoracetaldehyde, acetaldehydearea, areaatmosphere, atmospheresubstance, substancepeak, peakhigh, highspecified
Separation Characteristics of Shim-pack™ Series Phenyl/PFPP Reversed Phase Columns
2025|Shimadzu|Technical notes
C190-E325 Technical Report Separation Characteristics of Shim-pack™ Series Phenyl/PFPP Reversed Phase Columns Tomoaki Shimpo1, Tomoha Somano1 Abstract: Reversed phase chromatography is the most commonly used HPLC separation technique for a wide range of applications. Hydrophobic interactions are the dominant separation…
Key words
pfpp, pfppphenyl, phenylshim, shimpack, packresponsiveness, responsivenesshydrophobic, hydrophobicgist, gistvelox, veloxcapacity, capacitycolumns, columnsretention, retentionporous, porousbiphenyl, biphenylyes, yesbonding
Rapid Analysis of 2,4-DNPH Derivatized Aldehydes and Ketones
2021|Shimadzu|Applications
ERAS-1000-0174A LC Reversed-phase Shim-pack Series Shim-packTM VP-ODS 174A Rapid Analysis of 2,4-DNPH Derivatized Aldehydes and Ketones Keywords: precolumn derivatization, air pollutants 1. Formaldehyde 2. Acetaldehyde 3. Acetone 4. Acrolein 5. Propionaldehyde 6. Crotonaldehyde 7. 2-Butanone 8. Methacrolein 9. n-Butyraldehyde 10.…
Key words
shim, shimketones, ketonesaldehydes, aldehydespacktm, packtmods, odspack, packphase, phasereversed, reversedseries, seriesprogram, programmobile, mobilevolume, volumerate, rateinjection, injectiondetection