Analysis of Chloroacetanilide and Other Acetamide Herbicide Degradates in Drinking Water by LC-MS/MS
Applications | 2009 | Thermo Fisher ScientificInstrumentation
Polar degradates of chloroacetanilide herbicides, such as ethanesulfonic acid (ESA) and oxanillic acid (OA), can contaminate drinking water near agricultural areas. Sensitive and rapid monitoring of these compounds is essential for regulatory compliance and public health.
This application note demonstrates a robust LC-MS/MS method for quantifying nine chloroacetanilide herbicide degradates in drinking water. The study targets low detection limits, reproducible calibration across five concentration levels, and minimized analysis time.
The method uses a Thermo Scientific Accela LC system coupled to a TSQ Quantum Discovery MAX mass spectrometer. Chromatography is performed on a Hypersil GOLD 50 x 2.1 mm, 3 µm column at 65°C with a 5 mM ammonium acetate/methanol gradient at 0.25 mL/min. Key parameters include:
All nine analytes eluted within 22 minutes, allowing total cycle times under 40 minutes. At the lowest calibration level (3–5.7 ng/mL), peaks were clearly detected. Acetochlor ESA and alachlor ESA coeluted closely, requiring peak height quantitation. Dimethachlor ESA exhibited peak splitting; combined peaks were used for quantification. Calibration curves exhibited excellent linearity (R2 0.9972–0.9999). Precision over 25 injections showed RSDs of 2.0% (surrogate) and 2.3% (internal standard).
Implementing automated platforms such as Thermo Scientific EQuan can further enhance sensitivity and workflow efficiency. Adopting sub-2 µm columns and faster gradients will reduce run times. Expanding the method to additional polar degradates and emerging pollutants will address evolving environmental monitoring needs.
This LC-MS/MS method offers a fast, sensitive, and reproducible solution for determining chloroacetanilide herbicide degradates in drinking water. The streamlined workflow and strong performance metrics make it well suited for routine environmental and regulatory laboratories.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Polar degradates of chloroacetanilide herbicides, such as ethanesulfonic acid (ESA) and oxanillic acid (OA), can contaminate drinking water near agricultural areas. Sensitive and rapid monitoring of these compounds is essential for regulatory compliance and public health.
Objectives and Study Overview
This application note demonstrates a robust LC-MS/MS method for quantifying nine chloroacetanilide herbicide degradates in drinking water. The study targets low detection limits, reproducible calibration across five concentration levels, and minimized analysis time.
Methodology and Instrumentation
The method uses a Thermo Scientific Accela LC system coupled to a TSQ Quantum Discovery MAX mass spectrometer. Chromatography is performed on a Hypersil GOLD 50 x 2.1 mm, 3 µm column at 65°C with a 5 mM ammonium acetate/methanol gradient at 0.25 mL/min. Key parameters include:
- Injection volume 25 µL (full loop) using Accela autosampler
- Negative heated electrospray ionization at 3500 V
- Sheath gas 30 units, auxiliary gas 15 units at 300°C, capillary temperature 300°C
- MRM transitions optimized for each analyte and internal standards
- Surrogate (Dimethachlor ESA, 90 ng/mL) and internal standard (Butachlor ESA, 116 ng/mL)
Main Results and Discussion
All nine analytes eluted within 22 minutes, allowing total cycle times under 40 minutes. At the lowest calibration level (3–5.7 ng/mL), peaks were clearly detected. Acetochlor ESA and alachlor ESA coeluted closely, requiring peak height quantitation. Dimethachlor ESA exhibited peak splitting; combined peaks were used for quantification. Calibration curves exhibited excellent linearity (R2 0.9972–0.9999). Precision over 25 injections showed RSDs of 2.0% (surrogate) and 2.3% (internal standard).
Benefits and Practical Applications
- Elimination of offline extraction reduces sample preparation time and potential analyte loss
- High throughput with sub-40 minute cycle times supports routine water testing
- Sensitivity meets regulatory requirements for drinking water contaminants
- Robust reproducibility and linear calibration ensure reliable quantitation
Future Trends and Potential Applications
Implementing automated platforms such as Thermo Scientific EQuan can further enhance sensitivity and workflow efficiency. Adopting sub-2 µm columns and faster gradients will reduce run times. Expanding the method to additional polar degradates and emerging pollutants will address evolving environmental monitoring needs.
Conclusion
This LC-MS/MS method offers a fast, sensitive, and reproducible solution for determining chloroacetanilide herbicide degradates in drinking water. The streamlined workflow and strong performance metrics make it well suited for routine environmental and regulatory laboratories.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of Herbicide Degradates in Drinking Water on the Agilent 500 Ion Trap LC/MS (EPA 535)
2011|Agilent Technologies|Applications
Determination of Herbicide Degradates in Drinking Water on the Agilent 500 Ion Trap LC/MS (EPA 535) Application Note Environmental Author Abstract Ed George The performance of the Agilent 500 Ion Trap LC/MS in EPA Method 535 is evaluated Agilent Technologies,…
Key words
degradates, degradateswater, waterresonant, resonantesa, esaherbicide, herbicidechloroacetanalide, chloroacetanalideesi, esiapi, apiucmr, ucmrdrinking, drinkinggeorge, georgeacetamide, acetamidedrying, dryingrevisions, revisionsnegative
PESTICIDE STANDARDS For Food & Environmental Analysis
2016|Merck|Brochures and specifications
PESTICIDE STANDARDS For Food & Environmental Analysis • Certified Reference Materials (CRMs) TraceCERT® • Certified Reference Materials (CRMs) Matrix Standards • Isotope Labeled Standards • Pesticide Metabolite Standards • Proficiency Testing (PT) 2 Pesticide Standards | From Sigma-Aldrich The Sigma-Aldrich®…
Key words
methyl, methylpackage, packagepesticides, pesticidescrms, crmssulfone, sulfonestandards, standardsdescription, descriptionsize, sizepesticide, pesticidesulfoxide, sulfoxidetracecert, tracecertchlordane, chlordanetoxaphene, toxaphenesoil, soilsolution
Method Validation for Determining 513 Pesticide Residues in Cucumber Using LCMS-8060NX and 308 Residues by GCMS- TQ8040NX
2023|Shimadzu|Applications
Application Note No. 93 Method Validation for Determining 513 Pesticide Residues in Cucumber Using LCMS-8060NX and 308 Residues by GCMSTQ8040NX Rakan Saad Abdullah Alajmi 1, Hatim Zaid Ali Alhazmi 1, Sarah Awwadh Sultan Alotaibi 1 Jayachandran Namasivayam 2, Anant V.…
Key words
methyl, methylisomer, isomersulfone, sulfoneethyl, ethylsulfoxide, sulfoxidersdr, rsdrphorate, phoratemin, minfipronil, fipronildesethyl, desethylaminocarb, aminocarbdodemorph, dodemorphdisulfoton, disulfotonfenthion, fenthionspirotetramat
Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection
2018|Agilent Technologies|Applications
Application Note Environmental Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection ×104 23 Semipolar metabolites 56 52 48 44 Intensity (cps) 40 36 32 28 Three polar metabolites 24 20 DPC 16 DMS…
Key words
spe, speonline, onlinedirect, directpolar, polarsemipolar, semipolarcartridge, cartridgeinjection, injectiondpc, dpcenrichment, enrichmentloop, loopmdpc, mdpcpal, palcps, cpsanalytes, analytesanalytical