Installation Instruction for ProfIC Vario 3 Anion
Manuals | | MetrohmInstrumentation
Ion chromatography with inline dialysis offers automated and reliable separation of inorganic anions in complex sample matrices and is critical for laboratories requiring high throughput and data integrity.
This bulletin provides step by step instructions for installing, configuring and validating the ProfIC Vario 3 Anion system combining a Professional IC Vario 3 instrument, Metrohm Inline Dialysis, an 858 Sample Processor and MagIC Net software.
The setup includes the 940 Professional IC Vario for chromatography, the 858 Professional Sample Processor with dialysis cell, a conductivity detector and sequential suppressor modules. Inline dialysis employs a semi permeable membrane to remove matrix components before injection. Software MagIC Net handles instrument configuration, method development, purge and sample series execution.
Calibration across 0.2 to 10 mg/L for fluoride, chloride, nitrite, bromide, nitrate, sulfate and phosphate showed correlation coefficients above 0.9999. Precision studies (n=10) at 2 mg/L yielded relative standard deviations below 1.2 for all analytes. Inline dialysis time and transfer time were optimized to balance signal intensity and analysis speed.
This fully automated approach minimizes manual sample pretreatment, reduces contamination risk and ensures reproducible results for complex samples such as biological fluids and polluted waters. It supports routine QA/QC workflows in environmental and pharmaceutical laboratories.
Future developments may include integration of eluent production modules and liquid handling stations for in line reagent preparation and external needle rinsing, further improving automation. Advanced suppressor regeneration strategies and AI driven method optimization could enhance performance and adaptability.
The ProfIC Vario 3 Anion system with inline dialysis and MagIC Net software provides a robust platform for anion analysis in complex matrices, delivering high precision, automation and ease of use.
Ion chromatography
IndustriesManufacturerMetrohm
Summary
Význam tématu
Ion chromatography with inline dialysis offers automated and reliable separation of inorganic anions in complex sample matrices and is critical for laboratories requiring high throughput and data integrity.
Cíle a přehled studie
This bulletin provides step by step instructions for installing, configuring and validating the ProfIC Vario 3 Anion system combining a Professional IC Vario 3 instrument, Metrohm Inline Dialysis, an 858 Sample Processor and MagIC Net software.
Použitá metodika a instrumentace
The setup includes the 940 Professional IC Vario for chromatography, the 858 Professional Sample Processor with dialysis cell, a conductivity detector and sequential suppressor modules. Inline dialysis employs a semi permeable membrane to remove matrix components before injection. Software MagIC Net handles instrument configuration, method development, purge and sample series execution.
Hlavní výsledky a diskuse
Calibration across 0.2 to 10 mg/L for fluoride, chloride, nitrite, bromide, nitrate, sulfate and phosphate showed correlation coefficients above 0.9999. Precision studies (n=10) at 2 mg/L yielded relative standard deviations below 1.2 for all analytes. Inline dialysis time and transfer time were optimized to balance signal intensity and analysis speed.
Přínosy a praktické využití metody
This fully automated approach minimizes manual sample pretreatment, reduces contamination risk and ensures reproducible results for complex samples such as biological fluids and polluted waters. It supports routine QA/QC workflows in environmental and pharmaceutical laboratories.
Budoucí trendy a možnosti využití
Future developments may include integration of eluent production modules and liquid handling stations for in line reagent preparation and external needle rinsing, further improving automation. Advanced suppressor regeneration strategies and AI driven method optimization could enhance performance and adaptability.
Závěr
The ProfIC Vario 3 Anion system with inline dialysis and MagIC Net software provides a robust platform for anion analysis in complex matrices, delivering high precision, automation and ease of use.
Reference
- Metrohm Application Bulletin 368 Installation Instruction for ProfIC Vario 3 Anion
- Metrohm Manual 940 Professional IC Vario
- Metrohm Manual 858 Professional Sample Processor
- Metrohm Dialysis Cell Manual 8.110.8013
- Metrohm Application Bulletin AB-395
- Metrohm Eluent Production Module 941 Manual
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Installation Instruction for ProfIC Vario 7 Anion
2024|Metrohm|Manuals
Application Bulletin 379 Installation Instruction for ProfIC Vario 7 Anion The ProfIC Vario 7 Anion is a Professional IC Vario system with Inline Dilution and Inline Dialysis for the fully automatic determination of anions with sequential suppression in very high…
Key words
dialysis, dialysisacceptor, acceptorpump, pumpmsm, msmperistaltic, peristalticconnected, connectedrinsing, rinsingcell, celltubing, tubingmagic, magicblack, blacksolution, solutionwaste, wasterotor, rotornet
Installation Instruction for ProfIC Vario 7 Cation
2024|Metrohm|Manuals
Application Bulletin 380 Installation Instruction for ProfIC Vario 7 Cation The ProfIC Vario 7 Cation is a Professional IC Vario system with Inline Dilution and Inline Dialysis for the fully automatic determination of cations in high concentrations in matrices containing…
Key words
dialysis, dialysisacceptor, acceptorperistaltic, peristalticpump, pumpconnected, connectedcell, cellprocessor, processorblack, blackmagic, magicmsm, msmrotor, rotornet, netstation, stationtubing, tubingsample
Installation Instruction for ProfIC Vario 3 Cation
|Metrohm|Manuals
Application Bulletin 369 Installation Instruction for ProfIC Vario 3 Cation The ProfIC Vario 3 Cation is a Professional IC Vario system for the fully automated determination of non-suppressed cations or anions with Metrohm Inline Dialysis. «ProfIC Vario 3» safely masters…
Key words
station, stationdialysis, dialysisperistaltic, peristalticacceptor, acceptorhandling, handlingpump, pumpcell, cellrinsing, rinsingscrews, screwsliquid, liquidprocessor, processorcapillaries, capillariesconnected, connectedrack, racklithium
Installation Instruction for ProfIC Vario 2 Anion
|Metrohm|Manuals
Application Bulletin 365 Installation Instruction for ProfIC Vario 2 Anion ProfIC Vario 2 Anion is a Professional IC Vario system for the fully automatic determination of anions or cations with sequential suppression with Metrohm Inline Ultrafiltration. «ProfIC Vario 2» safely…
Key words
msm, msmstation, stationrinsing, rinsingultrafiltration, ultrafiltrationregeneration, regenerationhandling, handlingconnected, connectedrotor, rotorperistaltic, peristalticcapillary, capillarystoppers, stopperslabeled, labeledexemplary, exemplarysuppressor, suppressormcs