Determination of HCl (ppm range) in silicone oil
Applications | | MetrohmInstrumentation
Silicone oils are widely used in industrial and consumer applications for their thermal stability and lubricity. Trace amounts of hydrochloric acid in these oils can promote corrosion, degrade polymeric seals, and pose quality issues. Accurate quantification of HCl at low ppm levels is therefore essential for quality control and product safety.
This application note describes a thermometric titration method to quantify HCl in silicone oil at concentrations around 10 ppm. The study aims to establish a simple, reproducible procedure that minimizes sample preparation and delivers reliable results with good precision and accuracy.
Sample aliquots of silicone oil are weighed directly into the titration vessel. A 1:1 mixture of toluene and 2-propanol serves as the solvent. Paraformaldehyde is added to suppress interference. The titrant is 0.01 M KOH in 2-propanol, delivered at 1 mL/min under stirring. The exothermic endpoint is detected thermometrically by monitoring temperature changes and evaluating the second derivative of the thermal curve. Method blanks are determined by plotting titrant volume against sample mass and extrapolating the y-intercept via linear regression.
Two silicone oil samples were analyzed. For Sample 1, HCl content averaged 8.35 ppm (SD 0.14 ppm, RSD 1.68%). For Sample 2, the average was 11.39 ppm (SD 0.68 ppm, RSD 5.94%). Blank determinations yielded high linearity (R2 >0.96), ensuring accuracy. Titrant concentration was verified by standardizing against benzoic acid, yielding 0.0095 mol/L KOH (R2 0.9977). The method demonstrated consistent endpoint detection and reliable quantification across replicates.
Advances in thermometric titration may include inline process monitoring, integration with other analytical techniques like FTIR or mass spectrometry, and improved data automation. The approach could be extended to other trace acids or bases in nonaqueous media and adapted for real-time quality assessment in manufacturing.
The presented thermometric titration method offers a straightforward and reliable approach to determine low-level HCl in silicone oil. With robust blank correction and precise endpoint detection, it fulfills industrial QA/QC requirements for sensitivity, reproducibility, and throughput.
Titration
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Summary
Importance of the Topic
Silicone oils are widely used in industrial and consumer applications for their thermal stability and lubricity. Trace amounts of hydrochloric acid in these oils can promote corrosion, degrade polymeric seals, and pose quality issues. Accurate quantification of HCl at low ppm levels is therefore essential for quality control and product safety.
Aims and Overview of the Study
This application note describes a thermometric titration method to quantify HCl in silicone oil at concentrations around 10 ppm. The study aims to establish a simple, reproducible procedure that minimizes sample preparation and delivers reliable results with good precision and accuracy.
Methodology
Sample aliquots of silicone oil are weighed directly into the titration vessel. A 1:1 mixture of toluene and 2-propanol serves as the solvent. Paraformaldehyde is added to suppress interference. The titrant is 0.01 M KOH in 2-propanol, delivered at 1 mL/min under stirring. The exothermic endpoint is detected thermometrically by monitoring temperature changes and evaluating the second derivative of the thermal curve. Method blanks are determined by plotting titrant volume against sample mass and extrapolating the y-intercept via linear regression.
Used Instrumentation
- Thermometric titrator equipped with temperature sensor and automated titrant dosing
- Burette or dosing unit for precise KOH delivery
- Magnetic stirrer integrated into the titration vessel
- Analytical balance for sample weighing
Main Results and Discussion
Two silicone oil samples were analyzed. For Sample 1, HCl content averaged 8.35 ppm (SD 0.14 ppm, RSD 1.68%). For Sample 2, the average was 11.39 ppm (SD 0.68 ppm, RSD 5.94%). Blank determinations yielded high linearity (R2 >0.96), ensuring accuracy. Titrant concentration was verified by standardizing against benzoic acid, yielding 0.0095 mol/L KOH (R2 0.9977). The method demonstrated consistent endpoint detection and reliable quantification across replicates.
Benefits and Practical Applications of the Method
- Minimal sample preparation: direct weighing and dissolution
- High sensitivity and selectivity for HCl at ppm levels
- Thermometric detection avoids pH electrode issues in organic media
- Rapid analysis suitable for routine quality control
Future Trends and Possibilities for Application
Advances in thermometric titration may include inline process monitoring, integration with other analytical techniques like FTIR or mass spectrometry, and improved data automation. The approach could be extended to other trace acids or bases in nonaqueous media and adapted for real-time quality assessment in manufacturing.
Conclusion
The presented thermometric titration method offers a straightforward and reliable approach to determine low-level HCl in silicone oil. With robust blank correction and precise endpoint detection, it fulfills industrial QA/QC requirements for sensitivity, reproducibility, and throughput.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of TAN in engine oil with 859 Titrotherm
|Metrohm|Applications
Application Bulletin 341 e Determination of TAN in engine oil with 859 Titrotherm Page 1/4 Determination of TAN in engine oil with 859 Titrotherm Of interest to: oil industry Summary Sample The oil is dissolved in a mixture of toluene…
Key words
titer, titerblank, blankengine, enginetitration, titrationtan, tanoil, oilkoh, kohtitrant, titrantintercept, interceptaliquote, aliquotedetermination, determinationslope, sloperegression, regressionfactor, factormol
Determination of free fatty acids (FFA) in edible oils with 859 Titrotherm
|Metrohm|Applications
Application Bulletin 315 e Determination of free fatty acid (FFA) in edible oils with 859 Titrotherm Page 1/4 Determination of free fatty acids (FFA) in edible oils with 859 Titrotherm Of interest to: food industry Catalyst: Standard solution: Summary Edible…
Key words
oil, oiltan, tantiter, titerblank, blankoff, offolive, olivetitration, titrationdetermination, determinationyes, yestitrant, titrantstop, stopfactor, factorrapeseed, rapeseedslope, slopesunflower
Determination of Sulfate and Total Acids in Nitrating Mixture
|Metrohm|Applications
Thermo. Titr. Application Note No. H-038 Title: Determination of Sulfate and Total Acids in Nitrating Mixture Scope: Determination of Sulfate and Total Acids in Nitrating Mixture Principle: In a first titration the sulfate content of the sample solution is determined,…
Key words
sulfate, sulfatetiter, titerdetermination, determinationnaoh, naohacids, acidssize, sizetitration, titrationblank, blanktotal, totalmmol, mmoltitrant, titrantnitrating, nitratingsample, samplecurve, curveendpoint
Determination of Phosphate in Acid Etch Mixture
|Metrohm|Applications
Thermo. Titr. Application Note No. H-037 Title: Determination of Phosphate in Acid Etch Mixture Scope: Determination of phosphate content in an acid etching bath. Principle: The acid mixture is titrated with sodium hydroxide. Sample: Sample Preparation: Aqueous mixture of phosphoric,…
Key words
naoh, naohtiter, titerendpoint, endpointphosphate, phosphatehoac, hoacdetermination, determinationsize, sizemmol, mmolblank, blanksolution, solutionsample, sampleetch, etchmean, meanignored, ignoreddosed