Standardization of Barium Acetate Solution
Applications | | MetrohmInstrumentation
The accurate determination of sulfate content in phosphoric acid is essential for quality control in fertilizer production and related industrial processes. Standardizing the barium acetate titrant ensures reliable and reproducible sulfate measurements, while minimizing matrix effects and improving overall analytical accuracy.
This application note outlines a thermometric titration method for standardizing a 1 mol/L barium acetate solution using a sodium sulfate standard. The procedure closely mimics the actual sulfate determination in phosphoric acid to reduce discrepancies caused by differing sample matrices.
The standardization employs a thermometric titrator with the following parameters:
Reagents:
Procedure summary:
An example dataset using 0.40025 mol/L Na₂SO₄ standard yielded the following titrant volumes vs. sulfate moles:
Linear regression of mL titrant vs. mmol sulfate produced a slope of 0.94316 and intercept of 0.08636 mL (R² = 0.99999). Calculated molarity of the titrant is 1/0.94316 = 1.0603 mol/L, with a procedural blank of 0.0864 mL.
This thermometric titration approach offers:
Advancements may include:
The described method provides a robust and accurate procedure for barium acetate titrant standardization, crucial for reliable sulfate analysis in phosphoric acid. Its reproducibility, minimal sample preparation, and compatibility with existing thermometric titration systems make it valuable for industrial and research laboratories.
Titration
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Summary
Importance of the Topic
The accurate determination of sulfate content in phosphoric acid is essential for quality control in fertilizer production and related industrial processes. Standardizing the barium acetate titrant ensures reliable and reproducible sulfate measurements, while minimizing matrix effects and improving overall analytical accuracy.
Objectives and Study Overview
This application note outlines a thermometric titration method for standardizing a 1 mol/L barium acetate solution using a sodium sulfate standard. The procedure closely mimics the actual sulfate determination in phosphoric acid to reduce discrepancies caused by differing sample matrices.
Methodology and Instrumentation
The standardization employs a thermometric titrator with the following parameters:
- Titrant delivery rate: 6 mL/min
- Endothermic endpoints: 1
- Data smoothing factor: 65
Reagents:
- 1 mol/L barium acetate solution
- Saturated boric acid solution
- 0.4 mol/L sodium sulfate standard (or 0.2 mol/L for manual dispensing)
Procedure summary:
- Load one Dosino with barium acetate and a second with sodium sulfate standard (or dispense manually).
- Prepare titration vessels containing 5 mL boric acid and water to ~50 mL total volume after sulfate addition.
- Pre-dose 0.2 mL barium acetate to form BaSO₄ seed, wait 20 s, then titrate with sodium sulfate standard while stirring.
- After each run, remove vessel, rinse the thermometric sensor and assembly, and brush off BaSO₄ deposits. Fit a soda-lime guard tube on the Dosino to prevent CO₂ contamination.
Instrumentation Used
- Thermometric titrator with temperature sensor and second-derivative endpoint detection
- Dosino dispensers or volumetric pipettes for standard solution
- Stirring module and reaction vessels
- Soda-lime guard tube for CO₂ exclusion
Main Results and Discussion
An example dataset using 0.40025 mol/L Na₂SO₄ standard yielded the following titrant volumes vs. sulfate moles:
- 4.0025 mmol SO₄²⁻ → 3.660 mL Ba(OAc)₂
- 3.6023 mmol SO₄²⁻ → 3.286 mL Ba(OAc)₂
- … down to …
- 1.6010 mmol SO₄²⁻ → 1.396 mL Ba(OAc)₂
Linear regression of mL titrant vs. mmol sulfate produced a slope of 0.94316 and intercept of 0.08636 mL (R² = 0.99999). Calculated molarity of the titrant is 1/0.94316 = 1.0603 mol/L, with a procedural blank of 0.0864 mL.
Benefits and Practical Applications
This thermometric titration approach offers:
- High precision and reproducibility in titrant standardization
- Reduced matrix-related errors by replicating actual sample conditions
- Minimal reagent consumption and rapid endpoint detection
- Compatible with routine QA/QC workflows in industrial laboratories
Future Trends and Potential Applications
Advancements may include:
- Full automation and integration with laboratory information management systems
- Miniaturized sensors and flow-through titration cells for on-line monitoring
- Alternative seeding strategies and enhanced data processing algorithms
- Extension of thermometric titrimetry to other challenging matrices and analytes
Conclusion
The described method provides a robust and accurate procedure for barium acetate titrant standardization, crucial for reliable sulfate analysis in phosphoric acid. Its reproducibility, minimal sample preparation, and compatibility with existing thermometric titration systems make it valuable for industrial and research laboratories.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Titrimetric determination of sulfate
|Metrohm|Applications
Application Bulletin 140/4 e Titrimetric determination of sulfate Branch General analytical chemistry; water, wastewater, air, environmental protection; organic chemistry, chemistry; pharmaceutical industry; food, stimulants, beverages, flavours; fertilizers, base materials, explosives; detergents, surfactants, cosmetics; mineral resources, cement Keywords Sulfate; titration; potentiometric…
Key words
titrant, titrantsulfate, sulfatetiter, titertitration, titrationmol, moltitrimetric, titrimetricequivalence, equivalencedetermination, determinationuntil, untilsample, sampleblank, blankconsumption, consumptionselected, selectedstirring, stirringbarium
Titrimetric determination of sulfate
|Metrohm|Applications
Application Bulletin 140/4 e Titrimetric determination of sulfate Branch General analytical chemistry; water, wastewater, air, environmental protection; organic chemistry, chemistry; pharmaceutical industry; food, stimulants, beverages, flavours; fertilizers, base materials, explosives; detergents, surfactants, cosmetics; mineral resources, cement Keywords Sulfate; titration; potentiometric…
Key words
titrant, titrantsulfate, sulfatetiter, titertitration, titrationmol, moltitrimetric, titrimetricequivalence, equivalencedetermination, determinationuntil, untilsample, sampleblank, blankconsumption, consumptionselected, selectedstirring, stirringbarium
Standardization of Titrant for Direct Titration of Sodium
|Metrohm|Applications
Thermo. Titr. Application Note No. H-062 Title: Standardization of Titrant for Direct Titration of Sodium Scope: Standardization of titrant for direct determination of sodium. Principle: The direct titration of sodium is accomplished with a standardized solution of aluminium containing a…
Key words
titrant, titrantstandardization, standardizationsodium, sodiumstandardized, standardizedtitration, titrationcurve, curvedirect, directmmole, mmolebulb, bulbdsf, dsfmolarity, molaritystoichiometric, stoichiometricsolution, solutionendpoints, endpointsexothermic
Determination of sulfate in granular phosphate fertilizers with 859 Titrotherm
|Metrohm|Applications
AB Application Bulletin 307 e Determination of sulfate in granular fertilizer samples with 859 Titrotherm Page 1/4 Determination of sulfate in granular phosphate fertilizers with 859 Titrotherm Of interest to: Fertilizer industry Summary Sulfate may be rapidly and easily titrated…
Key words
titer, titertitrant, titrantblank, blankintercept, intercepttitration, titrationdetermination, determinationmap, mapfactor, factorslope, slopeendpoint, endpointfilter, filteroff, offsulfate, sulfatetiamotm, tiamotmformula