Standardization of Sodium Fluoride for Aluminium Titrations

Applications |  | MetrohmInstrumentation
Titration
Industries
Energy & Chemicals
Manufacturer
Metrohm

Summary

Significance of the Topic


Accurate standardization of sodium fluoride is essential for reliable aluminium determinations in industrial and research laboratories. Ensuring precise titrant concentration supports quality control in water analysis, metal manufacturing, and material certification.

Objectives and Study Overview


The study aims to establish a robust thermometric titration procedure to standardize 1 mol/L NaF using an exothermic reaction with aluminium. By generating a calibration curve correlating millimoles of Al with NaF consumption, the true molarity of the fluoride titrant is determined.

Methodology and Instrumentation


  • Primary Al standard (0.4 mol/L): Dissolve high‐purity Al in NaOH, dilute to volume.
  • Secondary Al standard (0.04 mol/L): Dilute primary solution and adjust acidity with HCl to dissolve Al(OH)₃ precipitate.
  • Buffer/reagent solutions: Acetate buffer with Na⁺/K⁺ ionic adjustment at pH 4.5.
  • Titration parameters: 2 mL/min titrant delivery, one exothermic endpoint, data smoothing factor 65, stirring speed setting 7.
  • Sample preparation: A series of at least four solutions mixing known volumes of Al standard, deionized water, and acetate buffer; example uses seven calibration points.

Instrumentation Used


  • Thermometric titrator equipped for exothermic endpoint detection.
  • Dosino dosing unit for precise Al standard addition.
  • 802 magnetic stirrer for consistent mixing.

Main Results and Discussion


A seven‐point calibration yielded a linear relationship between mmoles of Al and NaF volume, with slope 5.88646 and intercept 0.03825 mL (R² = 0.99999). Given the stoichiometry of 6 F⁻ per Al³⁺, the effective NaF concentration calculates to 6 / 5.88646 = 1.0193 mol/L. The negligible titration blank confirms minimal background heat effects.

Benefits and Practical Applications


  • High accuracy and reproducibility for routine aluminium analysis.
  • Simple reagent and sample preparations compatible with standard lab equipment.
  • Applicable across diverse matrices including water, soils, and metallurgical samples.

Future Trends and Opportunities


Advances may include integration with automated sampling systems, miniaturized thermometric sensors, and software‐driven endpoint recognition. Coupling with flow‐injection analysis or microfluidic platforms could further enhance throughput and reduce reagent consumption.

Conclusion


The described thermometric titration protocol delivers a reliable and precise method to standardize sodium fluoride for aluminium determinations. Its straightforward implementation and excellent linearity make it well suited for quality control and research laboratories.

References


No external literature citations were provided in the original application note.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of Aluminium by Fluoride Titration
Thermo. Titr. Application Note No. H-053 Title: Determination of Aluminium by Fluoride Titration Scope: Determination of aluminium in acidic, basic and neutral solutions; including alum, etching solutions and aluminate solutions. Principle: Titration with standard NaF solution in the presence of…
Key words
alum, alumtitration, titrationsolution, solutionaluminate, aluminatebayer, bayernaf, nafliquor, liquoraluminium, aluminiumdsf, dsfendpoints, endpointscurve, curvewaiting, waitingsmoothing, smoothingaliquots, aliquotsspent
Standardization of Titrant for Direct Titration of Sodium
Thermo. Titr. Application Note No. H-062 Title: Standardization of Titrant for Direct Titration of Sodium Scope: Standardization of titrant for direct determination of sodium. Principle: The direct titration of sodium is accomplished with a standardized solution of aluminium containing a…
Key words
titrant, titrantstandardization, standardizationsodium, sodiumstandardized, standardizedtitration, titrationdirect, directcurve, curvemmole, mmolebulb, bulbdsf, dsfmolarity, molaritystoichiometric, stoichiometricendpoints, endpointssolution, solutionexothermic
Automated analysis of etch acid mixtures using the 859 Titrotherm and the 814 USB Sample Processor
Application Bulletin 344/1e Automated Analysis of etch acid mixtures with 859 Titrotherm and 814 USB Sample Processor Page 1/7 Automated analysis of etch acid mixtures using the 859 Titrotherm and the 814 USB Sample Processor Of interest to: solar cell…
Key words
molarity, molaritytitrant, titranttitration, titrationnaoh, naohetch, etchhcl, hclintercept, interceptslope, slopeassignment, assignmenttitrating, titratingcomputed, computednaf, nafformula, formulammol, mmolblank
Determination of Ferric and Cupric Ions in Copper Refining Solutions
Thermo. Titr. Application Note No. H-070 Title: Determination of Ferric and Cupric Ions in Copper Refining Solutions Scope: Determination of Fe 3+ and Cu2+ in copper refining solutions by thermometric titration. It was found that the conventional approach of masking…
Key words
dsf, dsftitration, titrationsmoothing, smoothingimediately, imediatelycupric, cuprichead, headtitre, titreferric, ferricfactor, factorglacial, glacialrefining, refiningtitrant, titrantpermit, permitclicking, clickingcopper
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike