EAS: Evaluation of Pump Performance for Long Shallow Gradient Peptide Mapping Analysis

Posters | 2022 | WatersInstrumentation
HPLC
Industries
Pharma & Biopharma, Proteomics
Manufacturer
Waters

Summary

Significance of the Topic


Peptide mapping plays a pivotal role in biopharmaceutical characterization and impurity testing. Long, shallow organic gradients and low flow rates are required to separate diverse peptide mixtures derived from protein digests, placing stringent demands on liquid chromatography pump performance. Reliable retention times and stable baselines are essential for accurate quantitation and consistency across runs.

Objectives and Study Overview


This study aimed to evaluate pump performance in long, shallow gradient peptide mapping by comparing two configurations of the Arc Premier System (binary and quaternary solvent managers) to two conventional quaternary bio LC systems. An enolase digestion standard served as a representative complex sample. Key performance metrics included retention time reproducibility and signal-to-noise (S/N) ratios across eleven target peptide peaks.

Methodology and Instrumentation


The peptide mapping method employed a 96-minute gradient increasing by 0.5% organic solvent per minute. Chromatographic conditions included:
  • Mobile phase A: 0.1% trifluoroacetic acid in water
  • Mobile phase B: 0.1% trifluoroacetic acid in acetonitrile
  • Flow rate: 0.500 mL/min
  • Column: XSelect CSH C18, 2.5 µm, 4.6 × 150 mm, at 60 °C
  • Detection: UV at 214 nm, 10 Hz
  • Sample: Waters MassPREP Enolase digestion standard (25 µL injection, 10 °C sample tray)

Main Results and Discussion


Baseline stability and gradient delivery were assessed via chromatographic ripple and noise measurements. Retention time standard deviations and average S/N ratios were calculated across four systems:
  • Arc Premier Binary System (BSM)
  • Arc Premier Quaternary System (QSM)
  • Quaternary Bio LC System 1
  • Quaternary Bio LC System 2

Key findings:
  • Baseline ripple was minimal on both Arc Premier configurations, whereas quaternary systems exhibited noticeable baseline fluctuations.
  • Average retention time standard deviations for peaks 1–5 were 0.011 min (QSM) and 0.014 min (BSM), compared to 0.032–0.048 min on competing systems.
  • Arc Premier systems achieved the highest average S/N ratios across all eleven peaks, indicating lower baseline noise and improved sensitivity.

Benefits and Practical Applications


The superior retention time precision and enhanced signal-to-noise delivered by the Arc Premier systems enable more reliable quantitation in peptide mapping workflows. Consistent gradient performance reduces variability between runs and supports robust method transfer in quality control and research laboratories.

Future Trends and Potential Applications


Advances in pump and solvent management technologies will further improve performance in challenging biochromatography methods. Integrating optimized gradient delivery with high-resolution mass spectrometry and automated sample preparation can expand applications in proteomics, biotherapeutic characterization, and stability studies.

Conclusion


All tested systems met manufacturer specifications for peptide mapping under long, shallow gradient conditions. The Arc Premier System in both binary and quaternary configurations demonstrated the most consistent retention times, the flattest baselines, and the highest signal-to-noise ratios, illustrating its suitability for demanding biopharmaceutical analyses.

References


  1. Simeone J, Hong P. Peptide Mapping using Binary Biocompatible LC Systems: Evaluation of Retention Time Precision and Mixing Effects on Waters and Competitive LCs. Waters Application Note, 2021 Mar, 720007078.
  2. Delaney K, Birdsall RE, Yu YQ. Improving Peptide Mapping Studies and Reducing Assay Failures Through Reproducible Performance Using the ACQUITY Premier UPLC System (BSM). Waters Application Note, 2022 Apr, 720007593.
  3. Simeone J, Hong P, McConville PR. Performance of the ACQUITY UPLC I-Class PLUS System for Methods which Employ Long, Shallow Gradients. Waters Application Note, 2018 May, 720006290.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Evaluation of Arc™ Premier Pump Performance for Long Shallow Gradient Peptide Mapping Analysis
Application Note Evaluation of Arc™ Premier Pump Performance for Long Shallow Gradient Peptide Mapping Analysis Andrew Steere, Norris Wong, Paula Hong Waters Corporation Abstract The complex nature of protein digestions requires demanding conditions for liquid chromatography (LC) systems to separate…
Key words
arc, arcpremier, premiershallow, shallowmapping, mappingpeptide, peptideevaluation, evaluationlong, longgradient, gradientpump, pumpperformance, performanceanalysis, analysissystem, systemflatter, flatterbsm, bsmdemanding
The Arc Premier System (BSM-R) as a Flexible LC Platform for Peptide Mapping Assays
Application Note The Arc Premier System (BSM-R) as a Flexible LC Platform for Peptide Mapping Assays Robert E. Birdsall, Ying Qing Yu Waters Corporation This is an Application Brief and does not contain a detailed Experimental section. Abstract Peptide mapping…
Key words
premier, premierarc, arcbsm, bsmmapping, mappingassays, assaysplatform, platformflexible, flexiblepeptide, peptidedeployed, deployedacross, acrosssystem, systembinary, binarymanager, managermanufacturing, manufacturingdelivering
Migrating Peptide Mapping Methods from the Arc™ Premier System to the Alliance™ iS Bio System with Gradient SmartStart Technology and Intelligent Method Translator App (iMTA)
Application Note Migrating Peptide Mapping Methods from the Arc™ Premier System to the Alliance™ iS Bio System with Gradient SmartStart Technology and Intelligent Method Translator App (iMTA) Duanduan Han, Robert E. Birdsall, Karen Nyholm Waters Corporation Abstract The Alliance iS…
Key words
imta, imtasmartstart, smartstartalliance, alliancetranslator, translatormigrating, migratingapp, apparc, arcbio, biopremier, premierintelligent, intelligentsystem, systemmapping, mappinggradient, gradientpeptide, peptidetechnology
Demonstrating the Applicability of the ACQUITY™ Premier Binary System for Long Shallow Gradient Peptide Mapping Analysis
Application Note Demonstrating the Applicability of the ACQUITY™ Premier Binary System for Long Shallow Gradient Peptide Mapping Analysis Corey E. Reed, Jennifer Simeone, Paula Hong Waters Corporation Abstract Reversed-Phase Liquid Chromatography (RPLC) is a key analytical tool in peptide mapping…
Key words
shallow, shallowbinary, binarypremier, premierpeptide, peptideacquity, acquitymapping, mappingapplicability, applicabilitydemonstrating, demonstratinglong, longgradient, gradientsystem, systemanalysis, analysiscompetitor, competitordeliver, deliverbiocompatible
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike