Metrosep Carb 2 – XX0/4.0
Manuals | 2014 | MetrohmInstrumentation
Carbohydrate analysis is a cornerstone in food, beverage, pharmaceutical and biotechnology industries for quality control and research. High‐performance separation of mono‐, di‐ and oligosaccharides supports regulatory compliance, process monitoring and product characterization.
This application note presents operational parameters, eluent composition and regeneration procedures for the Metrosep Carb 2 ion‐exchange column series. The goal is to ensure reproducible carbohydrate separations and maintain column performance over extended use.
The Metrosep Carb 2 column series comprises three lengths (100, 150 and 250 × 4.0 mm i.d.) packed with a polystyrene/divinylbenzene copolymer containing quaternary ammonium groups (5 µm). A guard column (Metrosep Carb 2 Guard or S-Guard) is recommended to extend column life. Key operational parameters include:
Under the specified eluent and flow conditions, the Metrosep Carb 2 columns achieve sharp, baseline‐resolved carbohydrate peaks with stable retention times. Regeneration protocols for organic and inorganic contaminants restore column capacity effectively, preserving chromatographic performance. Guard columns and controlled operating conditions contribute to consistent results and extended column lifetime.
The robust design of the Metrosep Carb 2 series supports routine carbohydrate profiling in QA/QC laboratories and research settings. Flexible eluent formulation accommodates complex sample matrices. Well‐defined regeneration procedures minimize downtime and consumable costs.
Emerging developments include coupling pulsed amperometric detection with mass spectrometry for structural elucidation, miniaturized columns for high‐throughput screening and automated gradient control for comprehensive glycan profiling. Ongoing innovation in stationary phases may further enhance selectivity and sensitivity.
The Metrosep Carb 2 column family offers a robust, reproducible and user‐friendly solution for carbohydrate analysis. Clear operational guidelines and regeneration protocols ensure long‐term performance and analytical reliability across diverse applications.
No additional literature was cited in this document.
Consumables, Ion chromatography, LC columns
IndustriesManufacturerMetrohm
Summary
Significance of the topic
Carbohydrate analysis is a cornerstone in food, beverage, pharmaceutical and biotechnology industries for quality control and research. High‐performance separation of mono‐, di‐ and oligosaccharides supports regulatory compliance, process monitoring and product characterization.
Objectives and study overview
This application note presents operational parameters, eluent composition and regeneration procedures for the Metrosep Carb 2 ion‐exchange column series. The goal is to ensure reproducible carbohydrate separations and maintain column performance over extended use.
Methodology and instrumentation
The Metrosep Carb 2 column series comprises three lengths (100, 150 and 250 × 4.0 mm i.d.) packed with a polystyrene/divinylbenzene copolymer containing quaternary ammonium groups (5 µm). A guard column (Metrosep Carb 2 Guard or S-Guard) is recommended to extend column life. Key operational parameters include:
- Eluent: 100 mmol/L NaOH with 10 mmol/L sodium acetate. Organic modifiers (acetonitrile or methanol) may be added up to 50 % for eluent and up to 100 % for sample matrix.
- Flow rate: Recommended 0.8 mL/min (100 mm), 0.5 mL/min (150 mm), 0.5 mL/min (250 mm). Maximum flow rates are 1.6, 1.2 and 0.8 mL/min respectively, with a maximum operating pressure of 20 MPa (200 bar).
- Temperature: 20–60 °C (standard 30 °C).
- Sample preparation: Microfilter samples (0.45 µm) and use a pulsation dampener to protect the column from injector pressure surges.
Key results and discussion
Under the specified eluent and flow conditions, the Metrosep Carb 2 columns achieve sharp, baseline‐resolved carbohydrate peaks with stable retention times. Regeneration protocols for organic and inorganic contaminants restore column capacity effectively, preserving chromatographic performance. Guard columns and controlled operating conditions contribute to consistent results and extended column lifetime.
Benefits and practical application
The robust design of the Metrosep Carb 2 series supports routine carbohydrate profiling in QA/QC laboratories and research settings. Flexible eluent formulation accommodates complex sample matrices. Well‐defined regeneration procedures minimize downtime and consumable costs.
- Reliable separation of a wide range of carbohydrates.
- Extended column lifetime with guard protection.
- Adaptable method for various sample matrices.
Future trends and possibilities for application
Emerging developments include coupling pulsed amperometric detection with mass spectrometry for structural elucidation, miniaturized columns for high‐throughput screening and automated gradient control for comprehensive glycan profiling. Ongoing innovation in stationary phases may further enhance selectivity and sensitivity.
Conclusion
The Metrosep Carb 2 column family offers a robust, reproducible and user‐friendly solution for carbohydrate analysis. Clear operational guidelines and regeneration protocols ensure long‐term performance and analytical reliability across diverse applications.
References
No additional literature was cited in this document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Metrosep C 4 - xxx/4.0
2018|Metrohm|Manuals
0.70 mmol/L Dipicolinsäure CH-9100 Herisau [email protected] www.metrohm.com Vorbereitung ■ Die Säule während 1 bis 2 h mit Eluent spülen. ■ 6.1050.4x0 Metrosep C 4 - xxx/4.0 6.1050.410 Metrosep C 4 - 100/4.0 6.1050.420 Metrosep C 4 - 150/4.0 6.1050.430 Metrosep…
Key words
flujo, flujomaximale, maximalepression, pressionvon, vonder, derdébit, débiteluyente, eluyentecolumna, columnacolonne, colonnel'écoulement, l'écoulementcon, confiltro, filtrogegendruck, gegendruckorganische, organischeregeneration
Metrosep C Supp 1 - XXX/4.0
2015|Metrohm|Manuals
CH-9100 Herisau [email protected] www.metrohm.com 6.1052.4X0 Metrosep C Supp 1 - XXX/4.0 6.1052.410 Metrosep C Supp 1 - 100/4.0 6.1052.420 Metrosep C Supp 1 - 150/4.0 6.1052.430 Metrosep C Supp 1 - 250/4.0 Eluent ■ Standardeluent: 5 mmol/L Salpetersäure 50 µg/L…
Key words
colonne, colonnecolumna, columnales, lesmaximale, maximalelängen, längenlas, laslongitudes, longitudeslongueurs, longueursrecommandons, recommandonsverunreinigungen, verunreinigungencontaminación, contaminacióntoutes, toutestodas, todasregeneration, regenerationremarque
Metrosep A Supp 10 - xxx/4.0
2014|Metrohm|Manuals
Vorbereitung ■ Die Säule während 1 bis 2 h mit Eluent spülen. CH-9100 Herisau [email protected] www.metrohm.com 6.1020.0x0 Metrosep A Supp 10 - xxx/4.0 ■ Zur Vermeidung von hohem Gegendruck empfehlen wir, die Säule beim Einbau bei niedriger Flussrate (0.10 mL/min)…
Key words
von, vonmaximale, maximalepression, pressiongegendruck, gegendruckorganische, organischehohem, hohemflussrate, flussrateregeneration, regenerationestándar, estándarvermeidung, vermeidungflujo, flujogamme, gammecolonne, colonnegama, gamaque
Metrosep C 4 - xxx/2.0
2018|Metrohm|Manuals
CH-9100 Herisau [email protected] www.metrohm.com Standardeluent 1.70 mmol/L Salpetersäure Wenn die Säule mit organischen Komponenten verunreinigt ist, die Säule wie folgt reinigen: 0.70 mmol/L Dipicolinsäure 1. Die Säule gegen die Flussrichtung der Reihe nach mit folgenden Lösungen spülen (Flussrate 0.20 mL/min,…
Key words
maximale, maximaletour, tourflujo, flujopression, pressionlas, lasder, derd'écoulement, d'écoulementdébit, débiteluyente, eluyentecolumna, columnacolonne, colonnel'écoulement, l'écoulementvon, vonmetales, metalessoluciones