Metrosep A Supp 5 - xxx/4.0

Manuals | 2017 | MetrohmInstrumentation
Consumables, Ion chromatography, LC columns
Industries
Manufacturer
Metrohm

Summary

Importance of the Topic


Ion chromatography is a key technique for the separation and quantification of inorganic and small organic anions. High-performance anion‐exchange columns like the Metrosep A Supp 5 series enable reliable, reproducible analysis in environmental monitoring, food and beverage quality control, pharmaceutical testing, and industrial process control.

Objectives and Overview of the Product


This document summarizes the design, operating parameters, maintenance protocols, and application scope of the Metrosep A Supp 5 columns. Variants with lengths of 50, 100, 150, and 250 mm (4.0 mm i.d.) are provided to match analytical throughput and resolution requirements.

Methodology and Instrumentation


The Metrosep A Supp 5 columns feature a polyvinyl alcohol stationary phase functionalized with quaternary ammonium groups (5 µm particle size). They operate effectively over pH 3–12 and at temperatures between 20 and 60 °C. Recommended flow rate is 0.7 mL/min (maximum 0.8 mL/min) under pressures up to 15 MPa (150 bar). The standard eluent is a mixture of 3.2 mmol/L sodium carbonate and 1.0 mmol/L sodium bicarbonate. A guard column (Metrosep A Supp 5 Guard or S-Guard/4.0) protects the analytical column and extends its lifetime. Sample solutions must be filtered through a 0.45 µm membrane before injection. To avoid pressure spikes when introducing organic modifiers (acetone, acetonitrile, methanol up to 100 %), the flow rate should be adjusted gradually over 30 minutes.

Key Features and Discussion


  • Column dimensions: 50×4.0, 100×4.0, 150×4.0, 250×4.0 mm (length × i.d.).
  • Wide pH and temperature tolerance for diverse sample matrices.
  • Chemical and sequential suppression for baseline stability and low detection limits.
  • Comprehensive regeneration protocols to remove hydrophilic ion or organic contaminants and recover chromatographic performance.

Benefits and Practical Applications


The robust design and flexible operating range make this column series suitable for:
  • Environmental water analysis (inorganic anions, haloacetic acids precursors).
  • Food and beverage testing (organic acid profiling).
  • Pharmaceutical quality control (API counter‐ions).
  • Industrial process streams (corrosion monitoring, ultrapure water).

Future Trends and Possibilities


Advancements may include integration with hyphenated detection (IC-MS), development of high‐capacity polymer phases for trace analysis, microcolumn formats for fast screening, and software‐driven method automation.

Conclusion


The Metrosep A Supp 5 series offers versatile, high‐resolution anion separation with proven stability and reproducibility. Adherence to recommended eluent composition, flow settings, and regeneration procedures ensures long column life and reliable performance across a broad range of analytical applications.

Reference

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Metrosep A Supp 7 - XXX/4.0
Metrosep A Supp 7 - XXX/4.0
2017|Metrohm|Manuals
Metrosep A Supp 7 - XXX/4.0 Metrohm AG CH-9100 Herisau Schweiz [email protected] www.metrohm.com 8.107.8010ML / 2017-04-01 6.1006.6X0 6.1006.6X0 Metrosep A Supp 7 - XXX/4.0 6.1006.620: Metrosep A Supp 7 - 150/4.0 6.1006.630: Metrosep A Supp 7 - 250/4.0 DE Säulenmaterial…
Key words
von, vonmit, mitionen, ionenverunreinigung, verunreinigungiones, ionesvalency, valencycontamination, contaminationcontaminación, contaminaciónvalencia, valenciaorganische, organischeorganiques, organiquesflussrate, flussrateorganischen, organischenvalence, valencegamme
Metrosep A Supp 4 - 250/4.0
Metrosep A Supp 4 - 250/4.0
2017|Metrohm|Manuals
Metrosep A Supp 4 - 250/4.0 Metrohm AG CH-9100 Herisau Schweiz [email protected] www.metrohm.com 8.110.1876 / 2017-04-01 Metrosep A Supp 4 - 250/4.0 6.1006.430 6.1006.430 Metrosep A Supp 4 - 250/4.0 DE Säulenmaterial Polyvinylalkohol mit quartären Ammoniumgruppen, Partikeldurchmesser 9 µm Abmessungen…
Key words
regeneración, regeneraciónrégénération, régénérationcolonne, colonneregeneration, regenerationregeneriermethode, regeneriermethodesäulenofen, säulenofenmit, mithorno, hornovon, vonionen, ionenverunreinigung, verunreinigungiones, ionesméthode, méthodecon, conmétodo
Metrosep A Supp 5 - xxx/2.0
Metrosep A Supp 5 - xxx/2.0
2015|Metrohm|Manuals
■ CH-9100 Herisau [email protected] www.metrohm.com 6.1006.2X0 Metrosep A Supp 5 - xxx/2.0 Die Säule während 1 bis 2 h mit Eluent spülen. Vorsäule Metrosep A Supp 5 Guard/2.0 (6.1006.600) DE Säulenmaterial Polyvinylalkohol mit quarternären Ammoniumgruppen, Partikeldurchmesser 5 µm Abmessungen 6.1006.220:…
Key words
con, convalence, valencemit, mitvon, vondes, desmaximale, maximalepression, pressionregeneration, regenerationhidrófilos, hidrófilosionen, ionenverunreinigung, verunreinigungiones, ionescontamination, contaminationaniones, anionescontaminación
Metrosep A Supp 7 – XXX/2.0
Metrosep A Supp 7 – XXX/2.0
2016|Metrohm|Manuals
Metrosep A Supp 7 – XXX/2.0 Metrohm AG CH-9100 Herisau Schweiz [email protected] www.metrohm.com 6.1006.6X0 Leaflet 8.107.8045ML / 2016-06-22 6.1006.6X0 Metrosep A Supp 7 – XXX/2.0 6.1006.640: Metrosep A Supp 7 - 150/2.0 6.1006.650: Metrosep A Supp 7 - 250/2.0 DE…
Key words
valence, valencemit, mitionen, ionenverunreinigung, verunreinigungiones, ionescontaminación, contaminaciónvalencia, valenciagamme, gammemaximale, maximalegama, gamacontamination, contaminationcontaminations, contaminationspar, parions, ionsund