Nucleosil 5SA - 125/4.0

Manuals | 2014 | MetrohmInstrumentation
Consumables, Ion chromatography, LC columns
Industries
Manufacturer
MACHEREY-NAGEL

Summary

Significance of the Topic


The precise separation and determination of divalent cations and aliphatic amines is essential in environmental monitoring, industrial quality control, and pharmaceutical analysis. A robust ion chromatography method ensures reliable data for decision-making in research and routine testing.

Objectives and Overview


This note outlines the operational parameters, eluent composition, column care, and regeneration procedures for a Nucleosil 5SA ion chromatography column. By standardizing preparation and maintenance, it aims to optimize performance and extend column lifetime.

Methodology and Instrumentation


  • Column: Spherical silica gel with sulfonic acid groups, 125×4.0 mm, 5 μm particle size.
  • Guard column: Nucleosil 5SA cartridge/4.0 with compatible holder; dual-holder option available.
  • Eluents:
    • Tartaric/citric acid standard: 4.0 mmol/L tartaric acid, 0.5 mmol/L citric acid, 3.0 mmol/L ethylenediamine, 5 % acetone.
    • Oxalic acid (pH 4.0): 3.5 mmol/L oxalic acid, 2.5 mmol/L ethylenediamine, 5 % acetone.
  • Operating conditions:
    • pH range: 2–8 (optimal 2.5–3.5).
    • Flow rate: up to 2.0 mL/min; pressure ≤ 40 MPa.
    • Temperature: 30–40 °C recommended; max 60 °C.
  • Maintenance:
    • Column storage: short term in eluent; long term in methanol/water (1:4).
    • Regeneration: inject 100 µL 0.1 mol/L Na2H2EDTA (avoid alkaline EDTA), rinse with 30 mL 0.1 mol/L nitric acid, then methanol or other organic solvent.
    • Use pulsation absorber to protect against pressure surges.
    • Microfilter samples (0.45 µm) and match pH of standards and samples.

Main Results and Discussion


While specific chromatograms are not presented, the described conditions deliver stable retention times and peak shapes for target analytes. Regeneration and proper storage prevent baseline drift and ensure reproducible separations. The dual-cartridge guard option facilitates quick cartridge replacement.

Benefits and Practical Applications


  • Reliable quantification of magnesium, calcium, strontium, barium, nickel, zinc, cobalt, and aliphatic amines.
  • Adaptable to various laboratories due to straightforward maintenance and standard eluent recipes.
  • Extended column life through defined regeneration and storage protocols.

Future Trends and Potential Uses


Advances may include integration with mass spectrometric detection for enhanced sensitivity, greener eluent systems, microfluidic high-throughput formats, and automated cartridge switching to minimize downtime.

Conclusion


The Nucleosil 5SA column, combined with tailored eluent compositions and clearly defined operational and maintenance guidelines, provides a robust platform for the separation of divalent cations and aliphatic amines in diverse analytical settings.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Metrosep C 4 - xxx/4.0
Metrosep C 4 - xxx/4.0
2018|Metrohm|Manuals
0.70 mmol/L Dipicolinsäure CH-9100 Herisau [email protected] www.metrohm.com Vorbereitung ■ Die Säule während 1 bis 2 h mit Eluent spülen. ■ 6.1050.4x0 Metrosep C 4 - xxx/4.0 6.1050.410 Metrosep C 4 - 100/4.0 6.1050.420 Metrosep C 4 - 150/4.0 6.1050.430 Metrosep…
Key words
flujo, flujomaximale, maximalepression, pressionvon, vonder, derdébit, débiteluyente, eluyentecolumna, columnacolonne, colonnel'écoulement, l'écoulementcon, confiltro, filtrogegendruck, gegendruckorganische, organischeregeneration
Super-Sep - 100/4.6
Super-Sep - 100/4.6
|Metrohm|Manuals
CH-9100 Herisau [email protected] www.metrohm.com 6.1009.000 Super-Sep - 100/4.6 DE Säulenmaterial Polymethacrylat mit quaternären Ammoniumgruppen, Partikelgrösse 12 µm. Abmessungen 100 × 4.6 mm Vorsäule ■ 6.1009.010 Super-Sep Guard/4.6 (keine Halterung erforderlich) ■ 6.1005.050 PRP-1 guard cartridge/4.0 mit 6.2821.050 Doppelkartuschenhalter. (Ob diese…
Key words
précolonne, précolonneprecolumna, precolumnacolumna, columnad'échantillon, d'échantillonéluants, éluantseluyentes, eluyentesdebe, debepueden, puedenmáximo, máximomaximaler, maximalercolonne, colonnedivalent, divalentles, lesser, serconcentraciones
Metrosep C 4 - xxx/2.0
Metrosep C 4 - xxx/2.0
2018|Metrohm|Manuals
CH-9100 Herisau [email protected] www.metrohm.com Standardeluent 1.70 mmol/L Salpetersäure Wenn die Säule mit organischen Komponenten verunreinigt ist, die Säule wie folgt reinigen: 0.70 mmol/L Dipicolinsäure 1. Die Säule gegen die Flussrichtung der Reihe nach mit folgenden Lösungen spülen (Flussrate 0.20 mL/min,…
Key words
maximale, maximaletour, tourflujo, flujopression, pressionlas, lasder, derd'écoulement, d'écoulementdébit, débiteluyente, eluyentecolumna, columnacolonne, colonnel'écoulement, l'écoulementvon, vonmetales, metalessoluciones
Hamilton PRP-X100 - XX0/4.0
Hamilton PRP-X100 - XX0/4.0
2016|Hamilton Company|Manuals
Hamilton PRP-X100 - XX0/4.0 Metrohm AG CH-9100 Herisau Schweiz [email protected] www.metrohm.com Leaflet 8.107.1026 / 2016-11-30 6.1005.0X0 6.1005.0X0 Hamilton PRP-X100 - XX0/4.0 6.1005.000 Hamilton PRP-X100 - 125/4.0 6.1005.010 Hamilton PRP-X100 - 250/4.0 DE Säulenmaterial Polystyrol/Divinylbenzol-Copolymer mit quartären Ammoniumgruppen. Partikelgrösse 10 µm…
Key words
von, vonflussrate, flussrateeluent, eluentmaximale, maximaleregeneration, regenerationbedarf, bedarfbesoin, besoindiluer, diluerverdünnen, verdünnenl'éluant, l'éluantproben, probenschweiz, schweizgegendruck, gegendruckwechsel, wechselinnerhalb
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike