Gaining Deeper Insights into Thin Film Response
Applications | 2022 | Agilent TechnologiesInstrumentation
Thin film optical coatings are critical components in applications ranging from telecommunications to precision optics. Accurate measurement of their optical constants underpins performance optimization and quality control. Systematic errors in spectrophotometric data, particularly oscillations in total loss spectra caused by differences in measurement geometry, can compromise the reliability of derived film parameters. Advanced accessories that minimize these errors are therefore essential for high-precision thin film characterization.
This application note demonstrates how the Agilent Cary 5000 spectrophotometer, equipped with the universal measurement accessory (UMA), overcomes spectral oscillations in total loss analysis of thin films. The study compares traditional separate measurements of transmittance and reflectance at different angles of incidence (AOI) with simultaneous measurements at the same AOI and sample location. Reproducibility between different UMA units and mounts is also evaluated.
When transmittance and reflectance were measured at different AOI (7° for T and 10° for R), total losses TL(λ)=100%−T−R exhibited oscillations of ~0.4% due to interference effects combined with AOI mismatch. Theoretical modelling confirmed that these oscillations arise primarily from AOI differences. Switching to equal-angle measurements (both T and R at 7°) eliminated these oscillations, demonstrating the impact of precise geometry control. A second experiment using a different UMA unit and sample mount four months later yielded transmittance spectra differing by only ~0.15%, consistent with a thickness nonuniformity of ≈0.1% (Δd≈0.3 nm). This confirms excellent reproducibility across instruments and setups.
Integration of advanced multiangle spectrophotometry with automated sample handling and real-time data analysis will further enhance thin film quality control. Emerging software algorithms for automated thickness and refractive index fitting may leverage such high-accuracy datasets. Broader adoption of accessory technologies like UMA across industrial and research laboratories promises more standardized thin film characterization protocols.
The Agilent Cary 5000 spectrophotometer with the universal measurement accessory effectively eliminates spectral oscillations in total loss analysis by ensuring co-located transmittance and reflectance measurements at identical angles. This capability yields more accurate, reproducible optical constant data for thin films and supports robust quality assurance in optical coating fabrication.
NIR Spectroscopy, UV–VIS spectrophotometry
IndustriesMaterials Testing
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Thin film optical coatings are critical components in applications ranging from telecommunications to precision optics. Accurate measurement of their optical constants underpins performance optimization and quality control. Systematic errors in spectrophotometric data, particularly oscillations in total loss spectra caused by differences in measurement geometry, can compromise the reliability of derived film parameters. Advanced accessories that minimize these errors are therefore essential for high-precision thin film characterization.
Objectives and Study Overview
This application note demonstrates how the Agilent Cary 5000 spectrophotometer, equipped with the universal measurement accessory (UMA), overcomes spectral oscillations in total loss analysis of thin films. The study compares traditional separate measurements of transmittance and reflectance at different angles of incidence (AOI) with simultaneous measurements at the same AOI and sample location. Reproducibility between different UMA units and mounts is also evaluated.
Methodology and Instrumentation
- Sample preparation: Ta2O5 films (~292 nm) deposited on 25 mm diameter Suprasil substrates using magnetron sputtering.
- Measurements: s-polarized transmittance at 7° and 10°, and s-polarized reflectance at 10° for initial tests; equal-angle transmittance and reflectance at 7° for direct comparison.
- Instrumentation: Agilent Cary 5000 UV-Vis-NIR double-beam spectrophotometer with the universal measurement accessory capable of absolute, variable-angle specular reflectance and transmission at a fixed sample spot.
Main Results and Discussion
When transmittance and reflectance were measured at different AOI (7° for T and 10° for R), total losses TL(λ)=100%−T−R exhibited oscillations of ~0.4% due to interference effects combined with AOI mismatch. Theoretical modelling confirmed that these oscillations arise primarily from AOI differences. Switching to equal-angle measurements (both T and R at 7°) eliminated these oscillations, demonstrating the impact of precise geometry control. A second experiment using a different UMA unit and sample mount four months later yielded transmittance spectra differing by only ~0.15%, consistent with a thickness nonuniformity of ≈0.1% (Δd≈0.3 nm). This confirms excellent reproducibility across instruments and setups.
Benefits and Practical Applications
- Elimination of AOI-induced oscillations improves confidence in thin film parameter extraction.
- Simultaneous T and R measurement at the same sample location reduces systematic errors due to film nonuniformity.
- High reproducibility across multiple UMA units supports consistent manufacturing and QC workflows.
Future Trends and Applications
Integration of advanced multiangle spectrophotometry with automated sample handling and real-time data analysis will further enhance thin film quality control. Emerging software algorithms for automated thickness and refractive index fitting may leverage such high-accuracy datasets. Broader adoption of accessory technologies like UMA across industrial and research laboratories promises more standardized thin film characterization protocols.
Conclusion
The Agilent Cary 5000 spectrophotometer with the universal measurement accessory effectively eliminates spectral oscillations in total loss analysis by ensuring co-located transmittance and reflectance measurements at identical angles. This capability yields more accurate, reproducible optical constant data for thin films and supports robust quality assurance in optical coating fabrication.
References
- Amotchkina, T. V. et al. Oscillations in Spectral Behavior of Total Losses (1−R−T) in Thin Dielectric Films. Optics Express, 2 July 2012, 20(14), 16129–44.
- Tikhonravov, A. V. et al. Effect of Systematic Errors in Spectral Photometric Data on the Accuracy of Determination of Optical Parameters of Dielectric Thin Films. Appl. Opt. 2002, 41, 2555–2560.
- Woollam, J. Ellipsometry, Variable Angle Spectroscopic, Wiley Encyclopedia of Electrical and Electronics Engineering, J. Webster, ed. (Wiley, New York, 2000). Supplement 1.
- Tikhonravov, A. V. et al. Optical Parameters of Oxide Films Typically Used in Optical Coating Production. Appl. Opt. 2011, 50, C75–C85.
- Tikhonravov, A. et al. Reliable Determination of Wavelength Dependence of Thin Film Refractive Index. Proc. SPIE 2003, 5188, 331–342.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Optical Characterization of Materials Using Spectroscopy
2023|Agilent Technologies|Guides
Applications of UV-Vis-NIR Optical Characterization of Materials Using Spectroscopy Application Compendium > Return to table of contents Table of contents Introduction 4 Optics 5 Characterizing Sub-Nanometer Narrow Bandpass Filters Evaluation of the Cary Specular Reflectance Accessory for…
Key words
optical, opticalreturn, returnreflectance, reflectancecontents, contentstable, tableangle, angleincidence, incidencemeasurements, measurementswavelength, wavelengthtransmittance, transmittancereflection, reflectionspectrophotometer, spectrophotometermeasurement, measurementcoating, coatingbeam
High Volume Optical Component Testing
2020|Agilent Technologies|Applications
Application Note Materials testing and research High Volume Optical Component Testing Using an Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) with Solids Autosampler Author Introduction Travis Burt Manufacturers of high quality multilayer optical coatings require reliable methods to accurately measure…
Key words
angle, angleincidence, incidencereflectance, reflectancetransmittance, transmittanceoptical, opticalmeasurements, measurementsums, umsmps, mpsuma, umapol, polcoatings, coatingsaoi, aoiabsolute, absolutepatch, patchspecular
A Faster, More Accurate Way of Characterizing Cube Beamsplitters
2022|Agilent Technologies|Applications
Application Note Materials A Faster, More Accurate Way of Characterizing Cube Beamsplitters Using the Agilent Cary 7000 universal measurement spectrophotometer (UMS) Authors Abstract Travis Burt and Chris Colley Agilent Technologies Mulgrave, Victoria, Australia Cube beamsplitters (CBS) are critical optical components…
Key words
polarized, polarizedcoating, coatingbeamsplitter, beamsplittercbs, cbswavelength, wavelengthbeamsplitters, beamsplitterscube, cubeoptical, opticalbeam, beampolarization, polarizationincident, incidentreflected, reflectedangle, anglefilm, filmillumination
A Faster, More Accurate Way of Characterizing Cube Beamsplitters
2022|Agilent Technologies|Applications
Application Note Materials A Faster, More Accurate Way of Characterizing Cube Beamsplitters Using the Agilent Cary 7000 universal measurement spectrophotometer (UMS) Authors Abstract Travis Burt and Chris Colley Agilent Technologies Mulgrave, Victoria, Australia Cube beamsplitters (CBS) are critical optical components…
Key words
polarized, polarizedcoating, coatingbeamsplitter, beamsplittercbs, cbswavelength, wavelengthbeamsplitters, beamsplitterscube, cubeoptical, opticalbeam, beampolarization, polarizationincident, incidentreflected, reflectedangle, anglefilm, filmillumination