708-DS Dissolution Apparatus Site Preparation Checklist
Manuals | 2017 | Agilent TechnologiesInstrumentation
Correct site preparation for analytical instruments is a critical foundation for reliable operation, consistent performance, and extended lifetime. In dissolution testing and other pharmaceutical analyses, stable environmental and power conditions, proper bench space, and accessible utilities help prevent installation delays, unplanned downtime, and maintenance issues.
This specification and checklist aim to guide laboratory managers and engineers in preparing a site for the Agilent 708-DS Dissolution Apparatus within the 1200 Series SL System. The document outlines spatial, power, and environmental prerequisites along with customer responsibilities to ensure a smooth installation and immediate operational readiness.
The approach consists of an itemized requirements checklist covering:
The document delivers precise numerical specifications:
Adhering to the guidelines enables:
Advancements may include integrated environmental monitoring sensors, remote diagnostics to verify site conditions pre-installation, modular bench designs supporting quick instrument swaps, and energy-efficient power supplies to reduce laboratory footprint.
Comprehensive site preparation transforms a complex instrument delivery into a predictable, trouble-free process. Observing the specified space, power, and environmental criteria ensures the Agilent 708-DS Dissolution Apparatus performs at peak efficiency from day one.
Agilent Technologies. "Hardware Site Preparation Specification: 708-DS Dissolution Apparatus, 1200 Series SL System." Issued July 14, 2017; Revision 03.
Dissolution
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Correct site preparation for analytical instruments is a critical foundation for reliable operation, consistent performance, and extended lifetime. In dissolution testing and other pharmaceutical analyses, stable environmental and power conditions, proper bench space, and accessible utilities help prevent installation delays, unplanned downtime, and maintenance issues.
Objectives and Study Overview
This specification and checklist aim to guide laboratory managers and engineers in preparing a site for the Agilent 708-DS Dissolution Apparatus within the 1200 Series SL System. The document outlines spatial, power, and environmental prerequisites along with customer responsibilities to ensure a smooth installation and immediate operational readiness.
Methodology
The approach consists of an itemized requirements checklist covering:
- Spatial dimensions: bench clearance of at least 30 cm above and 10 cm on each side and rear of the instrument footprint.
- Weight support: bench capacity to bear the 54.4 kg (120 lb) instrument and accessories without overhang.
- Electrical requirements: two dedicated outlets matching either 115 V/60 Hz or 230 V/50 Hz specifications and accounting for any computer systems.
- Environmental stability: ambient temperature maintained between 5 °C and 40 °C, with relative humidity not exceeding 80% non-condensing and minimal vibration.
- Consumables and supplies: ordering required solvents, media, and maintenance items in advance.
Used Instrumentation
- Agilent 708-DS Dissolution Apparatus
- Wide Input Range Heater
- Circulator
Main Results and Discussion
The document delivers precise numerical specifications:
- Operating footprint: 62.2 cm (W) × 59.0 cm (D) × 68.0 cm (H), extending to 99.1 cm when deployed.
- Power consumption: up to 230 W at 115 V/60 Hz (2 A) or 230 W at 230 V/50 Hz (1 A), with heater and circulator drawing up to 1 150 W.
- Thermal and humidity ranges support reliable dissolution testing without risk of condensation or overheating.
Benefits and Practical Applications
Adhering to the guidelines enables:
- Rapid instrument commissioning with minimal configuration issues.
- Consistent dissolution testing results by maintaining stable thermal and humidity conditions.
- Reduced risk of bench damage or electrical hazards.
- Enhanced safety and operator training during installation services.
Future Trends and Opportunities
Advancements may include integrated environmental monitoring sensors, remote diagnostics to verify site conditions pre-installation, modular bench designs supporting quick instrument swaps, and energy-efficient power supplies to reduce laboratory footprint.
Conclusion
Comprehensive site preparation transforms a complex instrument delivery into a predictable, trouble-free process. Observing the specified space, power, and environmental criteria ensures the Agilent 708-DS Dissolution Apparatus performs at peak efficiency from day one.
Reference
Agilent Technologies. "Hardware Site Preparation Specification: 708-DS Dissolution Apparatus, 1200 Series SL System." Issued July 14, 2017; Revision 03.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
708-DS Dissolution Tester Hardware Site Preparation Specification
2019|Agilent Technologies|Manuals
708-DS Dissolution Tester Hardware Site Preparation $JL OHQW6HU LHV 6/6\VWHP Specification Purpose of Procedure Thank you for purchasing an Agilent solution. To get you started and to assure a successful and timely installation of your Agilent solution, please refer…
Key words
circulator, circulatorlhv, lhvdissolution, dissolutionheater, heateroperating, operatingsite, siteconsumption, consumptioninstrument, instrumentweight, weightpower, powerdescription, descriptioncustomer, customerhardware, hardwareheight, heightagilent
705 DS Dissolution Apparatus Site Preparation Checklist
2010|Agilent Technologies|Manuals
705 DS Dissolution Apparatus Site Preparation Checklist Hardware Site Preparation Specification Purpose of Procedure Thank you for purchasing an Agilent solution. To get you started and to assure a successful and timely installation of your Agilent solution, please refer to…
Key words
site, sitechecklist, checklistapprover, approverpreparation, preparationspecification, specificationrevision, revisionconsumption, consumptionhardware, hardwareinstrument, instrumenttechnologies, technologiesagilent, agilentoperating, operatingweight, weightdate, datepower
Agilent 709-DS Dissolution Apparatus Site Preparation Checklist
2011|Agilent Technologies|Manuals
Agilent 709-DS Dissolution Apparatus Site Preparation Agilent 1200 Series SLChecklist System Hardware Site Preparation Specification Purpose of Procedure Thank you for purchasing an Agilent solution. To get you started and to assure a successful and timely installation of your Agilent…
Key words
site, sitepower, powercontroller, controllerheater, heateragilent, agilentconsumption, consumptioninstrument, instrumentpreparation, preparationoperating, operatingweight, weightdescription, descriptionspecification, specificationcustomer, customerhardware, hardwareheight
400 DS Dissolution Apparatus Site Preparation Checklist
2015|Agilent Technologies|Manuals
400 DS Dissolution Apparatus Site Preparation Checklist Hardware Site Preparation Specification Purpose of Procedure Thank you for purchasing an Agilent solution. To get you started and to assure a successful and timely installation of your Agilent solution, please refer to…
Key words
site, sitechecklist, checklistapprover, approverpreparation, preparationspecification, specificationrevision, revisionconsumption, consumptionhardware, hardwareinstrument, instrumenttechnologies, technologiesagilent, agilentoperating, operatingweight, weightdate, datepower