Why Instrument Detection Limit (IDL) is a Better Metric for Determining The Sensitivity of Triple Quadrupole LC/MS Systems
Technical notes | 2014 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
IndustriesManufacturerAgilent Technologies
Key wordsmetric, idl, precision, your, triple, sensitivity, you, repeatability, deviation, noise, parra, lester, confirms, limit, signal, why, quadrupole, analysis, filtering, specification, determining, specified, taylor, benefit, report, reserpine, abundance, realistic, distinguishable, relied, numerical, better, detection, amount, meaningful, standard, statistically, rsd, smoothing, uncertainty, checklist, assessing, aligned, import, provides, evaluating, organizations, customers, daily, variability
Similar PDF
Agilent 6400 Series Triple Quadrupole LC/MS Systems
2015|Agilent Technologies|Brochures and specifications
Agilent 6400 Series Triple Quadrupole LC/MS Systems SUPERIOR SENSITIVITY WITH PROVEN PERFORMANCE AGILENT 6400 SERIES TRIPLE QUADRUPOLE LC/MS SYSTEMS THE AGILENT 6400 SERIES TRIPLE QUADRUPOLE SYSTEMS – PROVEN QUANTITATIVE PERFORMANCE Choose an Agilent 6420, 6460, or 6495 Triple Quadrupole LC/MS…
Key words
chip, chiptriggered, triggeredmrm, mrmforensic, forensicifunnel, ifunneltechnology, technologyagv, agvquantitation, quantitationdroplet, dropletagilent, agilentacquisition, acquisitionlibrary, libraryanalysis, analysisconfirmation, confirmationtoxicology
Instrument Detection Limit at Ultrashort Dwell Times Demonstrated on the Agilent 6495C Triple Quadrupole LC/MS
2020|Agilent Technologies|Technical notes
Technical Overview Instrument Detection Limit at Ultrashort Dwell Times Demonstrated on the Agilent 6495C Triple Quadrupole LC/MS Authors Charles Nichols, Behrooz Zekavat, and Patrick Batoon Agilent Technologies, Inc. Abstract This Technical Overview presents the measurement of the instrument detection limit…
Key words
idl, idldwell, dwellinstrument, instrumenttime, timecounts, countsextremely, extremelychromatographic, chromatographicminutes, minutestimes, timesshort, shortacquisition, acquisitionmrm, mrmrsd, rsdion, ionduty
Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless?
2011||Technical notes
Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless? Technical Overview Authors Abstract Greg Wells, Harry Prest, and The signal-to-noise of a chromatographic peak determined from a single measurement Charles William Russ IV, has served…
Key words
noise, noisesnr, snrsignal, signaldetection, detectionperformance, performancemetric, metricanalyte, analytemerit, meritlimits, limitsguidelines, guidelinesidl, idlinjections, injectionsreplicate, replicatevariance, varianceestablishing
Determination of Endocrine-Disrupting Chemicals in Drinking Water at Sub ng/L Levels Using the Agilent 6495 Triple Quadrupole Mass Spectrometer
2014|Agilent Technologies|Applications
Determination of Endocrine-Disrupting Chemicals in Drinking Water at Sub ng/L Levels Using the Agilent 6495 Triple Quadrupole Mass Spectrometer Application Note Authors Abstract László Tölgyesi This Application Note demonstrates the precise, accurate, and robust Agilent Technologies Sales and determination of…
Key words
equilin, equilinunit, unitandrostenedione, androstenedioneestriol, estrioltestosterone, testosteroneestrone, estronewidest, widestidl, idlcounts, countsacquisition, acquisitionhormones, hormoneswater, watertime, timedrinking, drinkingtap