Determination of an Anionic Fluorochemical Surfactant (FC-95) in a Steel Bath

Applications | 1998 | Thermo Fisher ScientificInstrumentation
Ion chromatography
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


The quantification of perfluorinated anionic surfactants such as FC-95 in steel etch baths is critical for ensuring proper wetting during chemical milling. Accurate monitoring at low part-per-million levels prevents surface defects, improves final part quality and minimizes cost by optimizing surfactant usage.

Objectives and Overview of the Study


This work aims to establish a sensitive and reproducible ion chromatographic method for determination of FC-95 in a highly corrosive steel bath matrix. The study focuses on preconcentration, matrix removal and quantification over a linear range of 10 to 50 ppm.

Methodology


Sample aliquots of steel bath containing FC-95 are injected into an ion chromatography system equipped with a preconcentration (IonPac NG1 guard) column that traps the surfactant while allowing inorganic bath components to be washed away. Following a 20-minute rinse with 20 mM sodium hydroxide, the surfactant is back-eluted onto an anion-exchange analytical column (OmniPac PAX-500) using an eluent composition of 55% 20 mM sodium hydroxide and 45% acetonitrile at 1 mL/min. Suppressed conductivity detection in external water mode provides sensitive measurement. Calibration standards in the steel bath matrix are prepared at 10, 25 and 50 mg/L FC-95.

Used Instrumentation


  • Dionex DX-500 Ion Chromatography system
  • GP40 Gradient Pump and CD20 Conductivity Detector
  • AutoSuppression ASRS-II (4 mm) in external water mode
  • IonPac NG1 Guard and OmniPac PAX-500 analytical columns
  • Rheodyne injection valves and PeakNet Chromatography Workstation

Main Results and Discussion


System blanks yielded low background conductance (1–4 μS) and minimal noise (∼5 nS). Seven replicate injections of 10 ppm FC-95 in the steel bath showed excellent reproducibility (area RSD 0.59%, retention time RSD 0.73%). The method demonstrated linear response from 10 to 50 ppm (r² = 1.00). Extended column rinsing effectively removed bath matrix interferences that elute in the void volume, ensuring clear detection of the surfactant peak.

Benefits and Practical Applications


  • Enables sensitive, low-ppm monitoring of FC-95 in aggressive etch solutions
  • Improves quality control in chemical milling processes
  • Offers cost savings by optimizing surfactant concentration
  • Compatible with strongly oxidizing and reducing media

Future Trends and Applications


The approach may be extended to other fluorosurfactants and corrosive matrices. Integration with mass spectrometry could provide structural confirmation. Development of online, real-time monitoring systems for process control and environmental surveillance is anticipated.

Conclusion


The described ion chromatographic method delivers reliable, reproducible, and linear quantification of FC-95 in steel etching baths at low ppm levels, facilitating robust quality assurance in metal finishing operations.

References


  1. Langworthy E. Metals Handbook, Aerochem, Inc. AS for Metals, 9th Ed., Vol. 16.
  2. Dionex Corporation. Application Note 119: Determination of an Anionic Fluorochemical Surfactant in a Semiconductor Etch Bath.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Determination of an Anionic Fluorochemical Surfactant in a Semiconductor Etch Bath
Application Note 119 Determination of an Anionic Fluorochemical Surfactant in a Semiconductor Etch Bath INTRODUCTION Perfluorinated surfactants are used as wetting agents in semiconductor acid etching solutions. Acid etching engraves fine patterns in silicon dioxide. Poor wetting of the wafer…
Key words
etch, etchsurfactant, surfactantload, loadfluorochemical, fluorochemicalbath, bathdqp, dqpeluent, eluentconcentrator, concentratorinject, injectrinsing, rinsingsemiconductor, semiconductoranionic, anionicloop, loopautosuppression, autosuppressiondeionized
Determination of Calcium and Magnesium in Brine
Determination of Calcium and Magnesium in Brine
1998|Thermo Fisher Scientific|Applications
Application Note 120 Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of alkaline earth metals. Membrane manufacturers recommend that saturated brines be purified…
Key words
magnesium, magnesiummetpac, metpacbrine, brinecalcium, calciumload, loaddqp, dqptmc, tmcconcentrator, concentratorloop, loopinject, injectcolumn, columneluent, eluenthcl, hclrecycle, recyclesample
The Determination of Trace Anions in Concentrated Phosphoric Acid
Technical Note 44 The Determination of Trace Anions in Concentrated Phosphoric Acid INTRODUCTION EQUIPMENT The determination of trace anions in phosphoric acid is hampered by a large excess of phosphate ion. Chloride determination at 0.1 mg/L (ppm) in 85% (w/w)…
Key words
phosphoric, phosphoricconcentrated, concentrateddeionized, deionizedanions, anionsacid, acidsample, sampleinject, injectice, icewater, waterwaste, wasteautosuppression, autosuppressionnitrate, nitratetrace, tracevalve, valveexclusion
Beverages Applications Notebook - Bottled Water
Beverages Applications Notebook - Bottled Water
2012|Thermo Fisher Scientific|Guides
Beverages Applications Notebook Bottled Water Table of Contents Index of Analytes......................................................................................................................................................................... 3 Introduction to Beverages........................................................................................................................................................... 4 UltiMate 3000 UHPLC+ Systems............................................................................................................................................... 5 IC and RFIC Systems.................................................................................................................................................................. 6 MS Instruments........................................................................................................................................................................... 7 Chromeleon 7 Chromatography Data System Software............................................................................................................ 8 Process Analytical Systems and…
Key words
bromate, bromatebottled, bottledmineral, mineralmdl, mdlhydroxide, hydroxidewater, waterionpac, ionpacdrinking, drinkingchlorate, chlorateeluent, eluentbromide, bromidechlorite, chloritephenols, phenolsamount, amountdetermination
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike