Glycoprotein Monosaccharide Analysis Using HPAE-PAD with Eluent Generation
Applications | 2016 | Thermo Fisher ScientificInstrumentation
Glycoprotein monosaccharide profiling is essential for monitoring the quality, efficacy, and safety of biotherapeutic proteins. Small variations in glycosylation can alter pharmacokinetics and immunogenicity, making robust analytical tools indispensable in process development and quality control.
This work demonstrates a validated HPAE-PAD method for monosaccharide composition analysis of glycoproteins using the Thermo Scientific Dionex ICS-5000 system. Key goals include defining optimal instrument configuration, assessing method reproducibility, establishing linearity and detection limits, and illustrating practical workflows for hydrolyzed protein samples.
Glycoproteins are hydrolyzed under acidic conditions (TFA or HCl) to release neutral and amino sugars. Separations employ a Dionex CarboPac PA20 analytical column preceded by an AminoTrap guard, with an electrolytically generated 10 mM KOH eluent. Pulsed amperometric detection uses a disposable gold-on-PTFE electrode and an Ag/AgCl reference.
A standard mixture of six monosaccharides (Fuc, GalN, GlcN, Gal, Glc, Man) achieved baseline separation within 13 min, with a total run time of 24 min including column regeneration. Fetuin hydrolysates showed expected enrichment of neutral or amino sugars depending on TFA versus HCl treatment. Human IgG, with low glycan content, was analyzed reproducibly despite higher peptide background.
Over seven days and more than 300 injections, retention times drifted by less than 3% and peak area RSDs remained below 5%. A 2-hour wash with 100 mM KOH fully restored retention. Detection limits of 0.08–0.17 µM (0.8–1.7 pmol on column) were achieved, and response was linear up to 170–420 µM, varying by analyte.
The method delivers high throughput monosaccharide analysis without derivatization, offering rapid turnaround for process monitoring and product characterization. The use of electrolytic eluent generation ensures reproducible mobile phase composition, while gold-on-PTFE disposable electrodes extend lifetime, tolerate higher hydroxide concentrations, and improve inter-system consistency compared to earlier disposable designs.
Emerging directions include integration with mass spectrometry for structural glycan analysis, automated on-line hydrolysis platforms, miniaturized flow systems for lower sample requirements, and expanded electrode materials to enhance sensitivity and selectivity. Advances in data processing and machine learning may further streamline glycomic profiling workflows.
The described HPAE-PAD approach on the Dionex ICS-5000 platform offers robust, sensitive, and reproducible monosaccharide composition analysis for glycoproteins. Its high precision, extended electrode lifetime, and automated eluent generation make it well suited for routine use in biopharmaceutical development and quality control.
Ion chromatography
IndustriesClinical Research
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Glycoprotein monosaccharide profiling is essential for monitoring the quality, efficacy, and safety of biotherapeutic proteins. Small variations in glycosylation can alter pharmacokinetics and immunogenicity, making robust analytical tools indispensable in process development and quality control.
Objectives and Study Overview
This work demonstrates a validated HPAE-PAD method for monosaccharide composition analysis of glycoproteins using the Thermo Scientific Dionex ICS-5000 system. Key goals include defining optimal instrument configuration, assessing method reproducibility, establishing linearity and detection limits, and illustrating practical workflows for hydrolyzed protein samples.
Methodology and Instrumentation
Glycoproteins are hydrolyzed under acidic conditions (TFA or HCl) to release neutral and amino sugars. Separations employ a Dionex CarboPac PA20 analytical column preceded by an AminoTrap guard, with an electrolytically generated 10 mM KOH eluent. Pulsed amperometric detection uses a disposable gold-on-PTFE electrode and an Ag/AgCl reference.
- Chromatography system: Dionex ICS-5000 with SP/DP pump, EG Eluent Generator Module, CR-ATC trap, DC compartment, ED detector
- Columns: CarboPac PA20 (3×150 mm) and AminoTrap (3×30 mm)
- Eluent: On-line generated KOH (10 mM) with nitrogen blanketing to minimize carbonate
- Detection waveform: Four-potential PAD cycle optimized for carbohydrates
- Injection: Partial-loop mode (4 µL cut volume for 10 µL injection) to conserve sample
Key Findings and Discussion
A standard mixture of six monosaccharides (Fuc, GalN, GlcN, Gal, Glc, Man) achieved baseline separation within 13 min, with a total run time of 24 min including column regeneration. Fetuin hydrolysates showed expected enrichment of neutral or amino sugars depending on TFA versus HCl treatment. Human IgG, with low glycan content, was analyzed reproducibly despite higher peptide background.
Over seven days and more than 300 injections, retention times drifted by less than 3% and peak area RSDs remained below 5%. A 2-hour wash with 100 mM KOH fully restored retention. Detection limits of 0.08–0.17 µM (0.8–1.7 pmol on column) were achieved, and response was linear up to 170–420 µM, varying by analyte.
Benefits and Practical Applications
The method delivers high throughput monosaccharide analysis without derivatization, offering rapid turnaround for process monitoring and product characterization. The use of electrolytic eluent generation ensures reproducible mobile phase composition, while gold-on-PTFE disposable electrodes extend lifetime, tolerate higher hydroxide concentrations, and improve inter-system consistency compared to earlier disposable designs.
Future Trends and Potential Applications
Emerging directions include integration with mass spectrometry for structural glycan analysis, automated on-line hydrolysis platforms, miniaturized flow systems for lower sample requirements, and expanded electrode materials to enhance sensitivity and selectivity. Advances in data processing and machine learning may further streamline glycomic profiling workflows.
Conclusion
The described HPAE-PAD approach on the Dionex ICS-5000 platform offers robust, sensitive, and reproducible monosaccharide composition analysis for glycoproteins. Its high precision, extended electrode lifetime, and automated eluent generation make it well suited for routine use in biopharmaceutical development and quality control.
Reference
- Varki A. Biological Roles of Oligosaccharides: All the theories are correct. Glycobiology. 1993;3(2):97–130.
- Thermo Scientific Dionex Technical Note 53: Determination of Glycoprotein Monosaccharide Composition by HPAE-PAD Using On-Line Electrolytically Generated Eluents. LPN 1405; 2004.
- Thermo Scientific Dionex Technical Note 71: Eluent Preparation for High-Performance Anion-Exchange Chromatography with Pulsed Amperometric Detection. LPN 1932-02; 2007.
- Pierce Micro BCA Protein Assay Kit 23235 Instructions; Thermo Scientific; 2001.
- Fan JQ, Namiki Y, Matsuoka K, Lee YC. Comparison of Acid Hydrolytic Conditions for Asn-Linked Oligosaccharides. Anal Biochem. 1994;219:375–378.
- Thermo Scientific Dionex Technical Note 110: Carbohydrate Determination by HPAE-PAD with Disposable Au on PTFE Working Electrodes. LPN 2952; 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent
2017|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 72225 Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent Authors Sachin Patil and Jeffrey Rohrer, Thermo Fisher Scientific, Sunnyvale, CA, USA Goal To present an accurate method of determining monosaccharides in glycoproteins using high-performance anion-exchange chromatography with…
Key words
monosaccharide, monosaccharidegaln, galnglcn, glcnfuc, fucglc, glcgal, galtfa, tfaaminotrap, aminotrapmonosaccharides, monosaccharidesoff, offgalactosamine, galactosaminearea, areadionex, dionexhcl, hclpeak
An improved HPAE-PAD method for glycoprotein monosaccharide determination
2017|Thermo Fisher Scientific|Applications
APPLICATION NOTE 72580 An improved HPAE-PAD method for glycoprotein monosaccharide determination Authors Sachin Patil and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Goal To demonstrate an improved HPAE-PAD assay for determination of the monosaccharide composition of a glycoprotein Introduction…
Key words
monosaccharide, monosaccharidegaln, galnglcn, glcndionex, dionexoff, offfuc, fucglc, glcgal, galgalactosamine, galactosaminetfa, tfahcl, hclboratetrap, boratetrapglucosamine, glucosaminepad, padmonosaccharides
Determination of Glycoprotein Monosaccharide Composition by HPAE-PAD Using On-Line Electrolytically Generated Eluents
2002|Thermo Fisher Scientific|Applications
Technical Note 53 Determination of Glycoprotein Monosaccharide Composition by HPAE-PAD Using On-Line Electrolytically Generated Eluents INTRODUCTION EQUIPMENT High-performance anion-exchange chromatography with pulsed amperometric detection (HPAE-PAD) at a gold electrode is a convenient and sensitive technique for carbohydrate analysis. HPAE-PAD provides…
Key words
monosaccharide, monosaccharidegalactosamine, galactosamineglucosamine, glucosamineglycoprotein, glycoproteinmannose, mannosefucose, fucosegalactose, galactosehydrolysis, hydrolysishydrolysates, hydrolysateshpae, hpaepad, padcomposition, compositioneluents, eluentselectrolytically, electrolyticallypmol
Guidelines for Successful Use of Thermo Scientific Dionex AminoTrap Columns
2012|Thermo Fisher Scientific|Applications
Lipika Basumallick, Udayanath Aich, and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Key Words Glycoprotein Monosaccharide Analysis, Gold on PTFE Disposable Electrode, Peak Fronting, HPAE-PAD Introduction The recommended technique to determine the monosaccharide content of a glycoprotein is to…
Key words
aminotrap, aminotrapdionex, dionexoff, offpad, padmonosaccharide, monosaccharideglucose, glucosecolumn, columnlysine, lysinemonosaccharides, monosaccharideshpae, hpaeeluent, eluentkoh, kohfailing, failingfronting, frontingcarbopac